| Literature DB >> 3049661 |
Abstract
A new enzyme immunoassay (Rubenostika; Organon Teknika, Turnhout, Belgium), a new latex agglutination test (Rubalex; Orion Diagnostica, Espoo, Finland), and three other accepted methods for the determination of rubella immunity were compared with a standard hemagglutination inhibition assay. Of 224 serum samples tested, 54 (24%) were nonreactive and 24 (11%) were low titered. All procedures were very specific (94 to 100%). Rubenostika was the least sensitive method (88%), and Rubalex was the most sensitive (98%).Entities:
Mesh:
Substances:
Year: 1988 PMID: 3049661 PMCID: PMC266665 DOI: 10.1128/jcm.26.8.1582-1583.1988