Comparison of impact of target delineation of computed tomography- and magnetic resonance imaging-guided brachytherapy on dose distribution in cervical cancer.

Lalida Tuntipumiamorn1, Suphalerk Lohasammakul2, Pittaya Dankulchai1, Pitchayut Nakkrasae1.   

Abstract

PURPOSE: The dose distributions obtained from three imaging approaches for target delineation in cervical cancer using high-dose-rate (HDR) brachytherapy were investigated.
MATERIAL AND METHODS: Ten cervical cancer patients receiving four fractions of HDR brachytherapy were enrolled. Based on different imaging approaches, three brachytherapy plans were developed for each patient: with the high-risk clinical target volume (HRCTV) delineated on magnetic resonance (MRI) images for every fraction (approach A; MRI-only); on MRI for the first fraction and computed tomography (CT) images for the subsequent fractions (approach B; MRI1st/CT); and on CT images for all fractions (approach C; CT-only). The volume, height, width at point A, width at maximum level, and dosimetric parameters (D100, D98, D95, and D90 of the HRCTV; and D0.1cc, D1cc, and D2cc of all organs at risk, or organ at risk - OAR: bladder, rectum, sigmoid colon, and bowel) provided by each approach were compared.
RESULTS: The mean HRCTV volume, width, and height obtained from approach C (CT-only) were overestimated compared to those from approaches A (MRI-only) and B (MRI1st/CT). The doses to the HRCTV for approaches A and B were similar. However, the HRCTV doses for approach C were significantly lower than those for approaches A and B for all parameters (D95-D100). As to the OAR, the three approaches showed no differences.
CONCLUSIONS: A combination of MRI and CT is a safe alternative approach for cervical cancer HDR brachytherapy. The technique provides comparable dosimetric outcomes to MRI-based planning, while being more cost-effective.

Entities:  

Keywords:  CT; MRI; brachytherapy; cervix cancer; target delineation

Year:  2018        PMID: 30479618      PMCID: PMC6251453          DOI: 10.5114/jcb.2018.78993

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

High-dose-rate (HDR) brachytherapy (BT) plays an essential role in the treatment of invasive cervical cancer [1,2]. With current technologies, the planning images have advanced from 2-dimensional-based planning to 3-dimensional image-guided brachytherapy (3D IGBT) [3,4,5,6]. The treatment delivery can be guided by either computed tomography (CT) or magnetic resonance imaging (MRI), but MRI has been demonstrated to provide superior soft-tissue contrast compared to CT, better image of extension of primary tumor, and its relationship with organs at risk is evidenced [7,8,9,10,11,12]. GEC-ESTRO after employment of MRI in cervix brachytherapy, with institutional series of hundreds of patients, EMBRACE study, and RetroEMBRACE study, have demonstrated better local controls and survival in these patients, when BT treatment is defined in a volume not in point A, and with the use of MRI [13,14,15]. However, the availability of MRI-based BT is limited at some radiotherapy centers from time, labor, and resource intensive. Transition to MRI-based BT as more accessible for institutions, with safe and efficient accomplishment was described by Harkenrider et al. [16]. Given these limitations, the application of planning involving a combination of MRI and CT has been proposed (MRI1st/CT). With this approach, the MRI is performed only in the first fraction of the treatment for target and organ at risk (OAR) delineation, applicator reconstruction, and dose optimization [17,18]. The contour information is then used to guide the radiation oncologist in determining the tumor target in the subsequent fraction by using CT images. The present research was therefore conducted to compare the impacts of the target delineation accuracies obtained using imaging with CT only, MRI only, and MRI1st/CT on the dose distributions in cervical cancer brachytherapy planning.

Material and methods

Patient selection

The study was approved by the Siriraj Institutional Review Board, Faculty of Medicine, Siriraj Hospital, Mahidol University (project number 760/2560). Between August 2015 and December 2016, 10 cervical cancer patients with stage IB2 to IIIB tumors were enrolled. All patients received 50 Gy external beam radiotherapy to the whole pelvis prior to the brachytherapy. The HDR brachytherapy schedule consisted of 4 treatment fractions with prescribed dose of 7 Gy. Intracavitary radiation therapy, using MRI-compatible tandem/ring applicators with or without interstitial plastic needles (Varian Medical Systems, Palo Alto, CA, USA) was performed for each treatment application.

Image acquisition

Patients with an applicator underwent T2-weighted MR imaging, with a 3 mm slice thickness using a 1.5-Tesla MRI scanner (Philips Archiva, Philips Medical Systems B.V., Eindhoven, The Netherlands) and CT imaging with a 1.25 mm slice thickness (GE Brightspeed, GE Medical Systems, Milwaukee, WI, USA) for brachytherapy planning. All patients followed the bladder filling protocol with 50-100 ml saline before the MRI and CT scans. The images were exported to brachytherapy planning system (BrachyVision version 13.6, Varian Medical Systems, Palo Alto, CA, USA). All target volumes and OAR delineations were performed on both image modalities by a same radiation oncologist.

Treatment planning

The treatment plans were generated for three different imaging schemes (Figure 1), in MRI-only approach in MRI1st/CT approach, and in CT-only approach. The MRI-only scheme was used as a standard approach. In MRI-only, the high-risk clinical target volume (HRCTV) and all OAR (bladder, rectum, sigmoid colon, and small bowel) on each fraction were contoured on the MRI, as recommended by GEC-ESTRO. In MRI1st/CT, images from the first fraction (MRI) were registered to the subsequent CT images (second to fourth fractions) using image registration software in the treatment planning system. Then, the HRCTV on the first fraction’s MRI was transferred to the corresponding subsequent CT with adjustment to the CT images. In CT-only, the HRCTV and all OAR were drawn on the CT images. To avoid bias in the contouring for CT-only, the HRCTV was delineated on the CT images by a single radiation oncologist before being contoured on the MRI.
Fig. 1

Schematic workflow of the three imaging approaches

Schematic workflow of the three imaging approaches After the HRCTV and all OAR delineations were completed, treatment plans were generated. On the MRI- and CT-based plans, the applicator reconstruction and dose optimization were done via a manual technique, using the TG-43 dose calculation algorithm to compute the desired dose distribution. The dwell times and location of the radioactive source were optimized for prescribed doses 7 Gy to D90 of the HRCTV, with the total target dose from external beam RT and BT ≥ 84 Gy, in 2 Gy equivalent (EQD2) and similar acceptable doses to the OAR in the same fraction for each planning approach to be ≤ 90 EQD2 for D2cc bladder, ≤ 70 for D2cc rectum and sigmoid as suggested by GEC-ESTRO Working Group II [19].

Data collection and statistical analyses

The volume, height, width at point A, width at the maximum level as well as the dosimetric parameters (D100, D98, D95, and D90 of the HRCTV; and D0.1cc, D1cc, and D2cc of all OAR) were collected from the dose-volume histogram and analyzed. The data were reported as averages (mean and SD). Repeated measure ANOVA was used to compare the differences between each fraction of parameters. Differences with a p value < 0.05 were considered to be statistically significant. All calculations were performed with SPSS Statistics for Windows, version 16.0 (SPSS Inc, Chicago, Ill, USA).

Results

HRCTV volume and dimension

The HRCTV volume from MRI of the 10 patients in this study was in the range of 11.9-30.2 cc (mean = 18.8, SD = 6.95). From Table 1, statistically significant ratios of the HRCTV CT to the MRI were 1.48 for the volume, 1.20 for the width at point A, and 1.21 for the height. Using MRI1st/CT, the ratios of the overestimated volume, width, and height (fraction 2-4) were reduced to 1.11, 1.12, and 1.07, respectively. As to fraction-by-fraction comparisons, no differences were detected between MRI and MRI1st/CT in terms of the total volume, the width at the maximum level, the width at point A, and the height (Table 2). However, the HRCTV volume, including the width at point A and the height obtained from CT, was significantly larger than that from MRI. Also, similar results of volume and height were shown for comparisons of the MRI1st/CT and CT approaches.
Table 1

Overall results of the HRCTV volume and dimensions for the ten patients

HRCTVFrMRI (A)MRI1st/CT (B)CT (C)Ratio
Mean ± SDMean ± SDMean ± SDA vs. Bp-valueA vs. Cp-value
Volume1-418.8 ±6.727.7 ±10.11.480.00*
2-418.6 ±6.820.8 ±6.826.7 ±9.31.110.02*1.470.00*
Width (maximum)1-44.2 ±0.44.4 ±0.61.060.04
2-44.1 ±0.44.1 ±0.54.3 ±0.40.980.781.050.09
Width (point A)12.7 ±0.73.2 ±0.71.200.01*
2-42.5 ±0.72.8 ±0.73.1 ±0.71.120.081.180.05*
Height12.7 ±0.73.3 ±0.41.210.01*
2-42.8 ±0.72.9 ±0.63.3 ±0.41.070.481.220.02*

HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation

statistically significant (p < 0.05)

Table 2

HRCTV volume and dimensions for fractions 2-4 for the three imaging approaches

HRCTVFrMRI (A)MRI1st/CT (B)CT (C)Approach A vs. BApproach A vs. CApproach B vs. C
Mean ± SDMean ± SDMean ± SDMean diffSEp-valueMean diffSEp-valueMean diffSEp-value
Volume219.0 ±7.320.5 ±7.027.7 ±9.6–1.530.950.42–8.701.330.00*–7.171.350.00*
318.8 ±6.920.7 ±6.926.7 ±8.8–1.860.580.31–7.931.320.00*–6.061.310.00*
418.4 ±6.721.8 ±6.926.3 ±10.1–3.411.040.03*–7.920.740.00*–4.511.710.03*
Width (maximum)24.1 ±0.44.1 ±0.64.3 ±0.6–0.030.111.00–0.250.120.07–0.230.100.06
34.2 ±0.54.1 ±0.54.3 ±0.30.110.100.83–0.110.210.60–0.200.190.33
44.1 ±0.44.2 ±0.54.4 ±0.6–0.060.111.00–0.300.140.06–0.280.090.02*
Width (point A)22.7 ±0.83.0 ±0.83.2 ±0.6–0.270.220.74–0.550.210.03*–0.270.170.15
32.6 ±0.62.8 ±0.73.2 ±0.8–0.250.170.54–0.760.240.01*–0.490.150.01*
42.6 ±0.62.8 ±0.53.1 ±0.9–0.260.140.32–0.560.170.01*–0.290.160.10
Height22.7 ±0.72.8 ±0.63.2 ±0.4–0.130.110.84–0.530.180.02*–0.400.200.08
32.9 ±0.72.9 ±0.63.3 ±0.50.010.091.00–0.460.160.02*–0.470.140.01*
42.7 ±0.72.8 ±0.63.3 ±0.4–0.100.151.00–0.620.230.03*–0.520.150.01*

HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error

statistically significant (p < 0.05)

Overall results of the HRCTV volume and dimensions for the ten patients HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation statistically significant (p < 0.05) HRCTV volume and dimensions for fractions 2-4 for the three imaging approaches HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error statistically significant (p < 0.05)

Dosimetric parameters

The dosimetric parameters for the HRCTV and all OAR are presented in Tables 3 and 4. There were no differences in the doses to the HRCTV for approaches A and B. In contrast, the doses to the HRCTV for approach C were significantly lower than those for MRI-only for all parameters (D95-D100). Similar results were shown for the comparisons of approaches B and C. As to the OAR doses, there were no differences between the three approaches.
Table 3

Summary of dosimetric results for HRCTV for the three imaging approaches

HRCTVFrMRI (A)MRI1st/CT (B)CT (C)Approach A vs. BApproach A vs. CApproach B vs. C
Mean ± SDMean ± SDMean ± SDMean diffSEp-valueMean diffSEp-valueMean diffSEp-value
D90 26.8 ±0.76.9 ±1.56.0 ±1.3–0.100.391.000.830.360.140.920.330.06
36.9 ±0.56.8 ±1.16.0 ±1.10.120.241.000.940.280.02*0.830.250.03*
46.9 ±0.37.0 ±0.76.3 ±0.7–0.060.191.000.690.220.04*0.750.190.01*
D95 26.2 ±0.86.3 ±1.55.3 ±1.2–0.070.371.000.890.350.100.960.340.06
36.3 ±0.66.2 ±1.15.3 ±1.10.120.221.000.960.290.03*0.840.260.03*
46.3 ±0.46.4 ±0.85.6 ±0.7–0.080.191.000.730.240.05*0.810.210.02*
D98 25.7 ±0.85.7 ±1.44.7 ±1.2–0.060.351.000.950.370.091.010.350.05
35.8 ±0.75.6 ±1.04.7 ±1.10.170.181.001.060.290.02*0.900.280.04*
45.8 ±0.55.9 ±0.85.0 ±0.8–0.070.191.000.760.240.04*0.830.220.01*
D100 24.5 ±0.84.4 ±1.13.4 ±1.00.080.251.001.050.290.02*0.970.290.03*
34.7 ±0.84.5 ±0.93.5 ±0.90.170.221.001.200.220.00*1.030.300.02*
44.6 ±0.54.6 ±0.83.8 ±0.6–0.060.211.000.750.200.02*0.800.210.02*

HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error

statistically significant (p < 0.05); D90, D95, D98, D100 – the minimum dose to 90%, 95%, 98%, 100% of the CTV

Table 4

Summary of dosimetric results for OAR for the three imaging approaches

OARFrMRI-only (A)MRI1st/CT (B)CT-only (C)Approach A vs. BApproach A vs. CApproach B vs. C
Mean ± SDMean ± SDMean ± SDMean diffSEp-valueMean diffSEp-valueMean diffSEp-value
Bladder
 0.1 cc26.9 ±1.07.4 ±2.07.4 ±2.0–0.540.490.89–0.550.490.87–0.010.010.94
36.8 ±0.78.2 ±2.08.2 ±2.0–1.370.600.14–1.370.590.140.000.011.00
47.0 ±0.67.5 ±1.47.5 ±1.4–0.490.551.00–0.480.551.000.010.010.59
 1 cc25.5 ±0.76.1 ±1.66.1 ±1.6–0.560.360.46–0.560.360.500.000.011.00
35.6 ±0.56.3 ±1.06.3 ±1.0–0.640.360.310.650.360.31–0.010.011.00
45.7 ±0.45.8 ±1.15.7 ±1.2–0.140.381.00–0.090.411.000.040.041.00
 2 cc25.0 ±0.75.5 ±1.55.5 ±1.5–0.510.310.42–0.510.310.410.000.000.58
35.1 ±0.45.7 ±0.85.7 ±0.8–0.600.290.22–0.600.290.220.000.001.00
45.1 ±0.45.2 ±0.95.2 ±1.00–0.120.311.00–0.080.341.000.040.041.00
Rectum
 0.1 cc25.0 ±0.85.4 ±1.15.4 ±1.1–0.390.170.14–0.400.170.13–0.010.011.00
35.2 ±0.46.3 ±1.96.3 ±1.9–1.080.630.35–1.090.630.36–0.010.011.00
45.2 ±0.65.9 ±1.05.8 ±1.1–0.720.280.09–0.640.320.230.080.081.00
 1 cc24.1 ±0.74.3 ±0.94.3 ±0.9–0.220.160.58–0.220.160.580.000.001.00
34.3 ±0.44.9 ±1.24.9 ±1.2–0.630.410.47–0.630.410.470.000.001.00
44.1 ±0.54.6 ±0.64.6 ±0.7–0.540.240.16–0.470.280.290.070.060.95
 2 cc23.6 ±0.63.9 ±0.93.9 ±0.9–0.270.170.41–0.270.170.400.000.001.00
33.8 ±0.44.4 ±1.04.4 ±1.0–0.610.340.32–0.620.340.32–0.010.011.00
43.7 ±0.54.1 ±0.54.1 ±0.6–0.420.230.31–0.360.250.580.060.061.00
Sigmoid
 0.1 cc22.9 ±1.24.2 ±2.14.2 ±2.0–1.240.450.06–1.250.440.060.000.011.00
32.7 ±1.23.9 ±2.13.9 ±2.1–1.150.530.17–1.140.530.170.000.001.00
42.9 ±1.04.4 ±2.34.3 ±2.3–1.500.620.13–1.450.630.150.040.041.00
 1 cc22.4 ±0.93.2 ±1.53.2 ±1.5–0.850.340.10–0.820.350.130.030.020.59
32.1 ±0.93.0 ±1.63.0 ±1.6–0.930.430.19–0.930.440.190.000.011.00
42.3 ±0.93.2 ±1.73.1 ±1.8–0.930.430.19–0.820.500.430.110.100.85
 2 cc22.1 ±0.82.8 ±1.32.8 ±1.3–0.670.310.19–0.670.310.180.000.000.50
31.8 ±0.82.7 ±1.42.8 ±1.5–0.840.370.15–0.940.350.07–0.100.100.99
42.1 ±0.82.9 ±1.52.8 ±1.6–0.810.380.21–0.710.440.430.100.080.77
Bowel
 0.1 cc23.0 ±1.54.1 ±1.14.1 ±1.1–1.130.610.30–1.130.610.310.000.011.00
33.3 ±1.24.1 ±2.14.1 ±2.1–0.790.510.49–0.790.520.500.000.011.00
43.9 ±1.25.7 ±2.45.7 ±2.4–0.190.730.13–1.880.730.12–0.010.000.28
 1 cc22.3 ±1.03.1 ±0.83.1 ±0.8–0.840.470.33–0.840.470.330.000.000.68
32.6 ±0.93.0 ±1.53.0 ±1.5–0.430.370.82–0.430.370.820.000.001.00
43.08 ±0.904.4 ±1.74.4 ±1.7–1.300.470.10–1.290.470.100.010.010.59
 2 cc22.11 ±0.962.75 ±0.672.75 ±0.67–0.630.440.56–0.630.440.560.000.001.00
32.31 ±0.842.64 ±1.362.64 ±1.36–0.330.331.00–0.330.331.000.000.001.00
42.75 ±0.813.87 ±1.493.87 ±1.49–1.120.390.09–1.120.390.090.000.001.00

OAR – organ at risk; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error; *statistically significant (p < 0.05); D0.1 cc, D1 cc, and D2cc – minimum dose to the most exposed 0.1 cm3, 1 cm3, and 2 cm3

Summary of dosimetric results for HRCTV for the three imaging approaches HRCTV – high-risk clinical target volume; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error statistically significant (p < 0.05); D90, D95, D98, D100 – the minimum dose to 90%, 95%, 98%, 100% of the CTV Summary of dosimetric results for OAR for the three imaging approaches OAR – organ at risk; MRI – magnetic resonance imaging; CT – computed tomography; Fr – fraction; SD – standard deviation; mean diff – mean difference; SE – standard error; *statistically significant (p < 0.05); D0.1 cc, D1 cc, and D2cc – minimum dose to the most exposed 0.1 cm3, 1 cm3, and 2 cm3

Discussion

It has been confirmed that 3D IGBT using MRI- or CT-based planning offers a better assessment of tumor volume than 2D-based planning. Their accuracy of the target volume and all OAR delineations effectively deliver more tumor-specific high doses, with less adverse effects on the surrounding normal tissues [3,4,20,21,22,23,24,25,26]. However, with CT, it is very difficult to identify the cervix and parametrium, and cause the tumor volume delineation to be overestimated, which results in a reduced dose coverage for the tumor volume and increased doses to OAR [7,8,9,10,26,27]. On the other hand, MRI-based planning provides better accuracy for the HRCTV and all OAR delineations. Superior treatment outcomes with impact in local control and survival from MRI-based planning have been presented by various groups [15,23,24,28,29]. However, the relative lack of availability of this superior modality is a disadvantage. Therefore, the combination of MRI1st/CT technique was introduced by Beriwal et al. [17]. The study showed that this approach is feasible and has excellent local control. A similar dose coverage to the HRCTV (especially in case of a small tumor volume) to that provided by fully MRI-based planning has also been reported by Nesvacil et al. [18]. Regarding the difference in the target volume using CT and MRI for brachytherapy for cervical cancer, Swanick et al. [30] studied the HRCTV volume using two image modalities, and showed that the mean HRCTV volume obtained using CT (44.1 cm3) was larger than that obtained using MRI (35.1 cm3), with statistical significance (p < 0.0001). Due to increased discrepancy in the HRCTV volume obtained with MRI, they concluded that MRI-based brachytherapy planning should be considered for the use in patients with a higher body mass index and a tumor size of 5 cm or greater, with parametrial invasion. In the present study, the HRCTV volume obtained from fully CT-based planning was overestimated compared to the volumes acquired using full MRI-based and MRI1st/CT approaches. As a consequence, the HRCTV doses obtained from CT-based planning were lower than those provided by the other two approaches in almost every fraction, and the range of dose differences were larger for the higher doses to the target (e.g., for approaches A and C: D90cc 0.69-0.83 Gy and D100cc 0.75-1.20 Gy, respectively). These outcomes are very similar to those in Nesvacil study [18], and it can be concluded that the overestimations on the CT images can be overcome by using MRI in the 1st fraction to guide contouring on the subsequent fractions of the CT images. This outcome was also confirmed by the results of better the HRCTV volumes and dimensions, and improved doses for approach B than approach C, even though all were obtained from CT images. A study by Choong et al. [31] analyzed 76 cervical cancer patients who had been divided into 49 cases with CT- and MRI-guided brachytherapy (the same as approach B in the present study) and 27 cases with MRI-guided only brachytherapy (the same as MRI-only in this study). The results showed no statistically significant differences in the clinical outcomes or dosimetric analyses of the two approaches. Wang et al. [32] reviewed 13 clinical studies comprising a total of 465 patients to evaluate the accuracy of two image modalities for the brachytherapy of cervical cancer. Ten studies compared CT to MRI, while the other three compared MRI1st/CT to MRI-based imaging; however, none of these studies compared all three imaging approaches on one patient (unlike the current study). Wang et al. found that, compared to MRI, the HRCTV width had been overestimated, while the height might have been underestimated when using CT. The dosimetric analyses for the HRCTV were also lower for the CT-only approach compared with at least one fraction of the MRI-based brachytherapy. These outcomes were similar to the results of the current study. However, the height of the HRCTV obtained from our CT approach was contoured at two-thirds of the uterine height, as suggested by Hegazy et al. [33] and was found to be significantly higher compared to the results obtained with the MRI-only and the MRI1st/CT approaches. As to fraction-by-fraction comparisons, there was a significant overestimation of the volume, and the dimension was obtained from the CT modality. With regards to doses delivered to OAR, most dosimetric parameters in the dose-volume histogram, including D0.1cc, D1cc, and D2cc showed no differences for the three imaging approaches used in the present study. In comparison, the Wang study [33] demonstrated that there were no statistically significant differences between the two modalities for the dosimetric analysis of OAR conducted in 4 studies. Furthermore, the MRI-guided brachytherapy provided an improved dose to the bladder in 4 studies, to the rectum in 4 other studies, and to the sigmoid in the remaining study. However, the clinical toxicity incidence showed no statistically significant difference between the two modalities. The results of the present study indicate that the use of MRI-guided brachytherapy for target localization in MRI1st/CT approach in cervical cancer brachytherapy provides an improved accuracy and precision of the tumor volume as well as a better dose coverage to the HRCTV, when it is compared with CT exclusive-based planning. At least one MRI is required to assess the tumor extension due to the overestimation of the HRCTV volume obtained using CT images (which was similar to the findings of other studies). However, there are some limitations of this study: it included small number of patients, and the interobserver variation was not assessed.

Conclusions

MRI1st/CT approach is a safe alternative for brachytherapy in cervical cancer patients, especially in those centers with limited MRI availability. Moreover, it is more cost-effective than the MRI-based approach.
  31 in total

1.  Early invasive cervical cancer: tumor delineation by magnetic resonance imaging, computed tomography, and clinical examination, verified by pathologic results, in the ACRIN 6651/GOG 183 Intergroup Study.

Authors:  Donald G Mitchell; Bradley Snyder; Fergus Coakley; Caroline Reinhold; Gillian Thomas; Marco Amendola; Lawrence H Schwartz; Paula Woodward; Harpreet Pannu; Hedvig Hricak
Journal:  J Clin Oncol       Date:  2006-12-20       Impact factor: 44.544

2.  MRI-guided adaptive radiotherapy in locally advanced cervical cancer from a Nordic perspective.

Authors:  Jacob Christian Lindegaard; Lars Ulrik Fokdal; Søren Kynde Nielsen; Jens Juul-Christensen; Kari Tanderup
Journal:  Acta Oncol       Date:  2013-08-21       Impact factor: 4.089

3.  Clinical impact of computed tomography-based image-guided brachytherapy for cervix cancer using the tandem-ring applicator - the Addenbrooke's experience.

Authors:  L T Tan; C E Coles; C Hart; E Tait
Journal:  Clin Oncol (R Coll Radiol)       Date:  2008-12-19       Impact factor: 4.126

Review 4.  Comparison of computed tomography and magnetic resonance imaging in cervical cancer brachytherapy: A systematic review.

Authors:  Fang Wang; Qiuying Tang; Guoxiao Lv; Feng Zhao; Xue Jiang; Xinli Zhu; Xinke Li; Senxiang Yan
Journal:  Brachytherapy       Date:  2016-12-10       Impact factor: 2.362

5.  Early invasive cervical cancer: CT and MR imaging in preoperative evaluation - ACRIN/GOG comparative study of diagnostic performance and interobserver variability.

Authors:  Hedvig Hricak; Constantine Gatsonis; Fergus V Coakley; Bradley Snyder; Caroline Reinhold; Lawrence H Schwartz; Paula J Woodward; Harpreet K Pannu; Marco Amendola; Donald G Mitchell
Journal:  Radiology       Date:  2007-11       Impact factor: 11.105

6.  Clinical impact of MRI assisted dose volume adaptation and dose escalation in brachytherapy of locally advanced cervix cancer.

Authors:  Richard Pötter; Johannes Dimopoulos; Petra Georg; Stefan Lang; Claudia Waldhäusl; Natascha Wachter-Gerstner; Hajo Weitmann; Alexander Reinthaller; Tomas Hendrik Knocke; Stefan Wachter; Christian Kirisits
Journal:  Radiother Oncol       Date:  2007-05       Impact factor: 6.280

7.  Physics contributions and clinical outcome with 3D-MRI-based pulsed-dose-rate intracavitary brachytherapy in cervical cancer patients.

Authors:  Cyrus Chargari; Nicolas Magné; Isabelle Dumas; Taha Messai; Lisa Vicenzi; Norman Gillion; Philippe Morice; Christine Haie-Meder
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-09-05       Impact factor: 7.038

Review 8.  How one institution overcame the challenges to start an MRI-based brachytherapy program for cervical cancer.

Authors:  Matthew M Harkenrider; Steven M Shea; Abbie M Wood; Bonnie Chinsky; Amishi Bajaj; Michael Mysz; Joseph H Yacoub; Ari Goldberg; Margaret Liotta; Ronald Potkul; Murat Surucu; John Roeske; William Small
Journal:  J Contemp Brachytherapy       Date:  2017-03-30

9.  Adaptive image guided brachytherapy for cervical cancer: a combined MRI-/CT-planning technique with MRI only at first fraction.

Authors:  Nicole Nesvacil; Richard Pötter; Alina Sturdza; Neamat Hegazy; Mario Federico; Christian Kirisits
Journal:  Radiother Oncol       Date:  2012-10-12       Impact factor: 6.280

10.  Non isocentric film-based intracavitary brachytherapy planning in cervical cancer: a retrospective dosimetric analysis with CT planning.

Authors:  Kirti Tyagi; Hari Mukundan; Deboleena Mukherjee; Manoj Semwal; Arti Sarin
Journal:  J Contemp Brachytherapy       Date:  2012-09-29
View more
  5 in total

1.  Comparison of computed tomography- and magnetic resonance imaging-based target delineation for cervical cancer brachytherapy.

Authors:  Fang Wang; Luyi Bu; Qun Wu; Xue Jiang; Lingyun Wu; Yu Li; Bin Xi; Zhongjie Lu; Senxiang Yan
Journal:  J Contemp Brachytherapy       Date:  2020-08-21

2.  RefineNet-based automatic delineation of the clinical target volume and organs at risk for three-dimensional brachytherapy for cervical cancer.

Authors:  Xue Jiang; Fang Wang; Ying Chen; Senxiang Yan
Journal:  Ann Transl Med       Date:  2021-12

3.  Clinical Outcomes of MRI-Guided Adaptive Brachytherapy for Each Fraction in Locally Advanced Cervical Cancer: A Single Institution Experience.

Authors:  Yunbo Chi; Ying Pan; Ning Zhang; Dongmei Han; Xin Guo; Zhuang Mao; Guanghui Cheng
Journal:  Front Oncol       Date:  2022-03-17       Impact factor: 6.244

4.  Dosimetric comparison of graphical optimization and inverse planning simulated annealing for brachytherapy of cervical cancer.

Authors:  Bin Tang; Xiangyu Liu; Xianliang Wang; Shengwei Kang; Pei Wang; Jie Li; Lucia Clara Orlandini
Journal:  J Contemp Brachytherapy       Date:  2019-08-29

5.  Inter-observer and intra-observer variability in reporting vaginal dose points for cervical cancer in high-dose-rate brachytherapy.

Authors:  Nuria Carrasco; Jose Chimeno; Mar Adrià-Mora; María José Pérez-Calatayud; Blanca Ibáñez; Vicente Carmona; Francisco Celada; Jose Gimeno; Françoise Lliso; José Pérez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2020-04-17
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.