| Literature DB >> 30413154 |
Na Yeon Jung1, Dong Hui Lim1,2, Sung Soon Hwang1, Joo Hyun1,3, Tae-Young Chung4.
Abstract
BACKGROUND: To compare the clinical outcome of Precizon toric intraocular lens (IOL) (Ophtec Inc.) to that of Tecnis toric IOL (Abbott Medical Optics Inc.).Entities:
Keywords: Astigmatism; Intraocular Lens; Precizon; Tecnis; Toric
Mesh:
Year: 2018 PMID: 30413154 PMCID: PMC6230282 DOI: 10.1186/s12886-018-0955-3
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Schematic images of the toric intraocualr lens. a: Precizon toric intraocular lens. b: Tecnis toric intraocular lens (b)
Demographics and clinical information of patients included in this study
| Precizon | Tecnis | ||
|---|---|---|---|
| Eyes (n) | 20 | 20 | |
| Patients (n) | 20 | 20 | |
| Age (y) | 64.64 ± 19.55 (22.5 to 87.1) | 64.51 ± 8.40 (46.6 to 87.8) | 0.980 |
| Male sex, n (%) | 8 (40) | 11 (55) | 0.527 |
| Right eyes, n (%) | 8 (40) | 10 (50) | 0.751 |
| UCDVA (logMAR) | 0.50 ± 0.17 (0.30 to 0.82) | 0.37 ± 0.13 (0.22 to 0.52) | 0.123 |
| BCDVA (logMAR) | 0.30 ± 0.18 (0 to 0.7) | 0.21 ± 0.13 (0 to 0.4) | 0.266 |
| Manifest Refraction | |||
| Sphere (D) | −0.12 ± 1.32 (−1.75 to 2.75) | 0.64 ± 2.82 (− 6.00 to 3.75) | 0.418 |
| Cylinder (D) | − 1.06 ± 0.94 (− 2.50 to − 0.50) | − 1.83 ± 1.29 (− 4.00 to − 0.50) | 0.129 |
| SE (D) | −0.66 ± 1.47 (− 2.50 to 2.75) | −0.28 ± 2.64 (− 6.50 to 1.88) | 0.680 |
| Corneal astigmatism (D) | 1.32 ± 0.45 (0.53 to 2.15) | 1.47 ± 0.47 (0.72 to 2.09) | 0.465 |
| IOL power (D) | 19.56 ± 2.35 (15.75 to 23.75) | 20.11 ± 3.52 (16.00 to 25.50) | 0.629 |
| IOL Cylinder power (D) | 1.96 ± 0.84 (1.00 to 3.50) | 2.36 ± 0.76 (1.50 to 4.00) | 0.266 |
| Axial length (mm) | 23.86 ± 1.04 (22.01 to 25.57) | 24.22 ± 0.83 (23.15 to 25.22) | 0.387 |
Mean ± SD (range)
Y years, LogMAR Logarithm of the minimum angle of resolution, D diopter, UCDVA uncorrected distance visual acuity, BCDVA best corrected distance visual acuity, IOL intraocular lens, SE spherical equivalent refraction
Preoperative and postoperative clinical data in the Precizon toric intraocular lens group and Tecnis toric intraocular lens group at 3-month postoperatively
| Parameters | Precizon | Tecnis | |||||
|---|---|---|---|---|---|---|---|
| Preop | Postop | Preop | Postop | ||||
| UCDVA (logMAR) | 0.50 ± 0.17 (0.30 to 0.82) | 0.09 ± 0.09 (0 to 0.30) |
| 0.38 ± 0.13 (0.22 to 0.52) | 0.08 ± 0.12 (0 to 0.30) |
| 0.904 |
| BCDVA (logMAR) | 0.30 ± 0.18 (0 to 0.7) | 0.02 ± 0.02 (0 to 0.05) |
| 0.21 ± 0.13 (0 to 0.50) | 0.01 ± 0.02 (0 to 0.05) |
| 0.582 |
| UCIVA (logMAR) | No data | 0.26 ± 0.13 (0.09 to 0.49) | No data | 0.40 ± 0.16 (0.20 to 0.60) | 0.114 | ||
| Manifest Refraction | |||||||
| Sphere (D) | −0.12 ± 1.32 (− 1.75 to 2.75) | 0.25 ± 0.35 (− 0.25 to 1.00) | 0.324 | 0.64 ± 2.82 (− 6.00 to 3.75) | 0.19 ± 0.48 (− 0.25 to 1.25) | 0.260 | 0.753 |
| Cylinder (D) | −1.06 ± 0.94 (− 2.50 to − 0.50) | −0.31 ± 0.29 (− 0.75 to 0) |
| −1.83 ± 1.29 (− 4.00 to − 0.50) | −0.41 ± 0.33 (− 0.75 to 0) |
| 0.491 |
| SE (D) | − 0.65 ± 1.47 (− 2.50 to 2.75) | 0.06 ± 0.38 (− 0.50 to 0.75) | 0.184 | −0.28 ± 2.64 (− 6.50 to 1.88) | −0.04 ± 0.50 (− 0.63 to 1.00) | 0.726 | 0.600 |
| Rotation (°) | No data | 1.50 ± 0.84 (0.18 to 3.02) | No data | 2.56 ± 0.68(1.50 to 3.50) |
| ||
Mean ± SD (range)
SD standard deviation, UCDVA uncorrected distance visual acuity, LogMAR Logarithm of the minimum angle of resolution, BCDVA best corrected distance visual acuity, UCIVA uncorrected intermediate visual acuity, D diopter, SE spherical equivalent refraction
*P values between the two groups, P < 0.05
Vector Analysis of astigmatism at 3-Month Postoperatively
| Parameters | Mean ± SD (range) | ||
|---|---|---|---|
| Precizon | Tecnis | ||
| TIA (D) | 1.41 ± 0.49 (0.54 to 2.30) | 1.41 ± 0.43 (0.75 to 2.18) | 0.982 |
| SIA (D) | 1.35 ± 0.52 (0.53 to 2.15) | 1.57 ± 0.79 (0.70 to 2.84) | 0.468 |
| DV (D) | 0.31 ± 0.23 (0.01 to 0.76) | 0.42 ± 0.26 (0.01 to 0.82) | 0.343 |
| Correction index (SIA/TIA) | 0.97 ± 0.25 (0.62 to 1.45) | 1.08 ± 0.27 (0.67 to 1.49) | 0.377 |
| Magnitude of error (arithematic SIA/TIA) | −0.06 ± 0.34 (− 0.58 to 0.46) | 0.16 ± 0.45 (− 0.45 to 0.81) | 0.233 |
| Angle of error (degree) | 0.19 ± 4.80 (− 6.7 to 14.4) | −2.78 ± 6.71 (− 20.4 to 1.6) | 0.274 |
| Absolute angle of error (degree) | 2.42 ± 4.10 (0.00 to 14.4) | 3.18 ± 6.52 (0.10 to 20.40) | 0.761 |
| Index of success (DV/TIA) | 0.23 ± 0.20 (0.02 to 0.75) | 0.30 ± 0.20 (0.01 to 0.68) | 0.470 |
SD standard deviation, TIA target induced astigmatism, SIA Surgically induced astigmatism, DV difference vector
Ocular Aberrometry Analysis at 3-Month Postoperatively
| Parameters | Mean ± SD (range) | ||
|---|---|---|---|
| Precizon | Tecnis | ||
| Total Aberrations RMS (μm) | 1.39 ± 0.86 (0.50 to 3.08) | 0.78 ± 0.53 (0.24 to 1.67) | 0.138 |
| HOA RMS(μm) | 0.37 ± 0.15(0.20 to 0.76) | 0.22 ± 0.11 (0.10 to 0.38) | 0.200 |
| Vertical coma (μm) | 0.07 ± 0.23 (− 0.22 to 0.47) | −0.05 ± 0.18 (− 0.30 to 0.23) | 0.290 |
| Horizontal coma (μm) | 0.01 ± 0.73 (− 1.94 to 0.60) | −0.10 ± 0.73 (− 1.40 to 0.59) | 0.757 |
| Vertical trefoil (μm) | −0.01 ± 0.33 (− 0.47 to 0.50) | 0.10 ± 0.24 (− 0.16 to 0.49) | 0.495 |
| Horizontal trefoil (μm) | 0.37 ± 0.54 (− 0.63 to 1.02) | 0.10 ± 0.67 (− 0.99 to 0.79) | 0.412 |
| Spherical aberration (μm) | 0.33 ± 0.16 (0.18 to 0.68) | 0.06 ± 0.16 (− 0.19 to 0.30) |
|
SD standard deviation, RMS root mean square, HOA higher order aberrations Malacara notation was converted to Optical Society of America standard notation, P < 0.05
Fig. 2Slit lamp images of the toric intraocular lens at 3 months postoperatively. a: Precizon toric intraocular lens. b: Tecnis toric intraocular lens. The axis markings can be clearly seen