Evaluation of optical performance of 4 aspheric toric intraocular lenses using an optical bench system: Influence of pupil size, decentration, and rotation.

Min-Ji Kim1, Young-Sik Yoo1, Choun-Ki Joo2, Geunyoung Yoon1.   

Abstract

PURPOSE: To evaluate the effect of pupil size, degree of intraocular lens (IOL) decentration, and rotation of 4 aspheric toric IOLs on the image quality.
SETTING: Department of Ophthalmology, Seoul St. Mary's Hospital, Seoul, South Korea.
DESIGN: Experimental study.
METHODS: Four aspheric toric intraocular lenses (IOLs)-the Precizon (transitional conic toric IOL), AT Torbi 709M (bitoric IOL), SN6AT4 (posterior toric surface IOL), and ZCT225 (anterior toric surface IOL)-were evaluated using the optical bench metrology system. Measurements included changes in spherical aberrations, relative spherical equivalent (SE), and image quality at different pupil diameters and image quality degradation due to decentration and rotation of the IOLs.
RESULTS: Change in relative SE with pupil size in aberration-free toric IOLs (transitional conic toric and bitoric IOLs) was greater than in negatively aspheric toric IOLs (posterior toric surface and anterior toric surface IOLs). In contrast, the aberration-free IOLs showed higher contrast than the negatively aspheric IOLs. When IOLs were decentered by 1.0 mm, the contrast reduction rates at 17.6 cycles per degree for the transitional conic toric IOL, bitoric IOL, posterior toric surface IOL, and anterior toric surface IOL were 5.1%, 3.1%, 12.2%, and 15.8%, respectively. Rotation-induced deterioration of contrast to 0.5 required a much higher rotation for the transitional conic toric IOL than for the other 3 IOLs.
CONCLUSIONS: The transitional conic toric IOL and bitoric IOL provided superior image quality despite pupil size changes and the presence of decentration. The transitional conic toric IOL demonstrated maximum rotation tolerance compared with the other IOLs. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2015        PMID: 26703305     DOI: 10.1016/j.jcrs.2015.10.059

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Influence of across-pupil sutures on the optical quality of intraocular lenses.

Authors:  Chang Rae Rho; Min-Ji Kim; Choun-Ki Joo
Journal:  Int Ophthalmol       Date:  2017-02-04       Impact factor: 2.031

2.  Effect of decentration, tilt and rotation on the optical quality of various toric intraocular lens designs: a numerical and experimental study.

Authors:  Jesús Pérez-Gracia; Jorge Ares; Francisco J Ávila; Laura Remón
Journal:  Biomed Opt Express       Date:  2022-03-08       Impact factor: 3.562

Review 3.  Tilt and decentration with various intraocular lenses: A narrative review.

Authors:  Xiao-Yong Chen; Yu-Chen Wang; Tian-Yao Zhao; Zi-Zhen Wang; Wei Wang
Journal:  World J Clin Cases       Date:  2022-04-26       Impact factor: 1.534

4.  Transitional conic toric intraocular lens for the management of corneal astigmatism in cataract surgery.

Authors:  Francisco Bandeira; Merce Morral; Daniel Elies; Sergio Eguiza; Spyridoula Souki; Felicidad Manero; Jose L Güell
Journal:  Clin Ophthalmol       Date:  2018-06-08

Review 5.  Continuous Transitional Focus (CTF): A New Concept in Ophthalmic Surgery.

Authors:  Ramón Ruiz Mesa; Tiago Monteiro
Journal:  Ophthalmol Ther       Date:  2018-06-18

6.  Comparison of clinical outcomes of toric intraocular lens, Precizon vs Tecnis: a single center randomized controlled trial.

Authors:  Na Yeon Jung; Dong Hui Lim; Sung Soon Hwang; Joo Hyun; Tae-Young Chung
Journal:  BMC Ophthalmol       Date:  2018-11-09       Impact factor: 2.209

7.  How does the world appear to patients with multifocal intraocular lenses?: a mobile model eye experiment.

Authors:  Eun Chul Kim; Kyung-Sun Na; Hyun Seung Kim; Ho Sik Hwang
Journal:  BMC Ophthalmol       Date:  2020-05-06       Impact factor: 2.209

8.  Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes.

Authors:  Jiaqi Meng; Wenwen He; Xianfang Rong; Ao Miao; Yi Lu; Xiangjia Zhu
Journal:  Eye Vis (Lond)       Date:  2020-04-07
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.