Clinical outcomes after implantation of a toric intraocular lens with a transitional conic toric surface.

Bettina C Thomas1,2, Ramin Khoramnia1,2, Gerd U Auffarth1,2, Mike P Holzer1,2.   

Abstract

PURPOSE: To evaluate the postoperative outcomes after implantation of a hydrophilic toric intraocular lens (tIOL) with a transitional conic toric surface in patients with astigmatism undergoing refractive lens exchange (RLE) or cataract surgery.
METHODS: In this monocentric study, clinical outcomes following implantation of the hydrophilic Precizon toric IOL model 565 (Ophtec) were analysed. For intraoperative axis alignment, the Verion Image Guided System was used. Intended versus achieved refraction, IOL position, visual acuity outcomes and power vector analyses were evaluated 3 months postoperatively.
RESULTS: Forty-three eyes of 28 patients (40 eyes cataract, 3 eyes RLE) received a Precizon toric IOL with a cylindrical power between 1.0 and 7.5 D. Target refraction was emmetropia except for four eyes that had a myopic target. The median postoperative IOL rotation at slit lamp was 3° with a range from 0° to 16°. Uncorrected distance visual acuity (UDVA) increased from median 0.5 logMAR preoperatively to median 0.06 logMAR at 3 months postoperatively. Corrected distance visual acuity increased from 0.20 to -0.02 logMAR. Median preoperative subjective cylinder was 1.5 D and 3 months postoperatively 0.25 D.
CONCLUSION: Preoperative diagnostics, IOL selection and correct intraoperative positioning of tIOLs are essential for good outcomes and postoperative spectacle independence following cataract or RLE surgery. Visual acuities and rotational stability of the Precizon toric IOL showed very good outcomes and the subjective postoperative astigmatism could be reduced significantly. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

Entities:  

Keywords:  axis alignment; power vector analysis; toric intraocular lens; transitional conic toric surface

Mesh:

Year:  2017        PMID: 28774937     DOI: 10.1136/bjophthalmol-2017-310386

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  4 in total

1.  Comparison of the clinical outcomes of Eyecryl and Tecnis® toric intraocular lenses: A real-world study.

Authors:  Mithun Thulasidas; Ajita Sasidharan; Balam Pradeep
Journal:  Indian J Ophthalmol       Date:  2022-03       Impact factor: 2.969

Review 2.  Optimizing outcomes with toric intraocular lenses.

Authors:  Manpreet Kaur; Farin Shaikh; Ruchita Falera; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

3.  Comparison of clinical outcomes of toric intraocular lens, Precizon vs Tecnis: a single center randomized controlled trial.

Authors:  Na Yeon Jung; Dong Hui Lim; Sung Soon Hwang; Joo Hyun; Tae-Young Chung
Journal:  BMC Ophthalmol       Date:  2018-11-09       Impact factor: 2.209

4.  Effect of 1.8-mm steep-axis clear corneal incision on the posterior corneal astigmatism in candidates for toric IOL implantation.

Authors:  Xi Li; Xiang Chen; Suhong He; Wen Xu
Journal:  BMC Ophthalmol       Date:  2020-05-06       Impact factor: 2.209