Soft bone primary stability of 3 different miniscrews for orthodontic anchorage: a resonance frequency investigation.

Mario Veltri1, Beatrice Balleri, Cecilia Goracci, Roberto Giorgetti, Piero Balleri, Marco Ferrari.   

Abstract

INTRODUCTION: Several miniscrews are available for skeletal anchorage. They have different geometries with self-drilling or self-tapping features. The aim of this study was to evaluate the soft bone primary stability of 3 different orthodontic screws by using the resonance frequency analysis.
METHODS: Aarhus mini-implant (Aarhus Mini-implant, Charlottenlund, Denmark) (A), Mini Spider Screws (HDC, Sarcedo, Italy) (S), and Micerium Anchorage System (Micerium, Avegno, Italy) (MAS) were investigated. To be compatible with the device used for resonance frequency analysis, the screws were modified (an abutment [Astra Tech, Mölndal, Sweden] was soldered on top). Four screws per system were tested. Each screw was placed in 5 excised rabbit femoral condyles, providing experimental models of soft bone. Placement was drill-free for the A screw, whereas the MAS and S screws required a pilot hole through the cortical layer. After each placement procedure, resonance frequency was assessed as a parameter of primary stability. Differences among the systems were analyzed by using analysis of variance for repeated measures, with the level of significance at P < 0.05.
RESULTS: The recorded resonance frequencies (in Hz) were (mean +/- SD): MAS, 6236.1 +/- 192.1; S, 6270.1 +/- 99.7; and A, 6193.1 +/- 142.4. Differences among the groups were not statistically significant (P > 0.05).
CONCLUSIONS: The resonance frequency analysis is applicable to comparatively assess the primary stability of orthodontic miniscrews. The 3 systems had similar outcomes in an experimental model of soft bone.

Entities:  

Mesh:

Year:  2009        PMID: 19409347     DOI: 10.1016/j.ajodo.2007.05.025

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  3 in total

1.  A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device.

Authors:  Yara K Hosein; S Jeffrey Dixon; Amin S Rizkalla; Ali Tassi
Journal:  Angle Orthod       Date:  2018-11-01       Impact factor: 2.079

Review 2.  Effectiveness of Orthodontic Mini-Screw Implants in Adult Deep Bite Patients during Incisor Intrusion: A Systematic Review.

Authors:  Ahmed Ibrahim Atalla; Mai Hamdy AboulFotouh; Fady Hussein Fahim; Manal Yehia Foda
Journal:  Contemp Clin Dent       Date:  2019 Apr-Jun

3.  Effects of a newly designed apparatus on orthodontic skeletal anchorage.

Authors:  Murat Tozlu; Didem Nalbantgil; Fulya Ozdemir
Journal:  Eur J Dent       Date:  2013-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.