Application of Six Sigma for evaluating the analytical quality of tumor marker assays.

Qian Liu1, Mei Fu1, Fumeng Yang1, Wei Liang1, Chuanxi Yang2, Wenjun Zhu1, Liming Ma1, Changxin Zhao1.   

Abstract

CONTEXT: The results of detection assays for the same specimen are usually quite different in different laboratories or when tested with different detection systems.
OBJECTIVE: This study was designed to investigate the value of applying sigma metrics derived from different standards for allowable total error (TEa) in evaluating the analytical quality of tumor marker assays.
METHODS: Assays were evaluated for these six tumor markers: total prostate-specific antigen (tPSA), carcinoembryonic antigen (CEA), alpha-fetoprotein (AFP), carbohydrate antigen 199 (CA199), carbohydrate antigen 125 (CA125), and carbohydrate antigen 153 (CA153). Sigma values were calculated for two concentrations of quality control products to assess differences in quality of tumor marker assays. Improvement measures were recommended according to the quality goal index, and appropriate quality control rules were selected according to the sigma value.
RESULTS: The sigma value was highest using the higher biological variation-derived "appropriate" TEa standard: it was sigma ≥6 or higher in 16.7% of tumor markers. Sigma was below 6 for all tumor markers using the other three TEa. CEA, AFP, CA199, CA125, and CA153 required improved precision. The marker tPSA required improve precision and accuracy. According to sigma values by using China's external quality assessment standards, CEA, AFP, CA125, and CA153 require 13s /22s /R4s /41s multirules for internal quality control, CA199 requires use of 13s /22s /R4s /41s /8x multirules, and tPSA requires maximum quality control rules.
CONCLUSION: Six Sigma is useful for evaluating performance of tumor markers assays and has important application value in the quality control of these assays.
© 2018 The Authors Journal of Clinical Laboratory Analysis Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  allowable total error; quality control rules; sigma metrics; tumor marker

Mesh:

Substances:

Year:  2018        PMID: 30280434      PMCID: PMC6585744          DOI: 10.1002/jcla.22682

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


Alpha‐fetoprotein Carbohydrate antigen 125 Carbohydrate antigen 153 Carbohydrate antigen 199 Carcinoembryonic antigen Clinical Laboratory Improvement Amendments of 1988 Coefficient of variation Defects per million External quality assessment National Center for Clinical Laboratories Quality goal index Royal College of Pathologists of Australasia Total allowable error Total prostate‐specific antigen

INTRODUCTION

Detection of tumor markers is widely conducted in medical laboratories in China. The results of detection assays for the same specimen are usually quite different in different laboratories or when tested with different detection systems.1 Therefore, it is necessary to improve the quality control of tumor marker detection to ensure accurate and credible results. Sigma refers to “standard deviation” in mathematical statistics and is often used to express the defects per million (dpm) when measuring the performance of production processes. Six Sigma represents a defect rate of detection of 3.4 dpm.2 Six Sigma theory was first applied in the field of medicine by Nevakainen et al.3 As understanding of this theory spread, Six Sigma became an important quality management tool. Therefore, an increasing number of researchers have used Six Sigma to study quality improvement in laboratories.4, 5, 6 The sigma metric can evaluate the measurement performance of different assay processes as the quality targets and their sigma values differ. The concept of “precision medical programs,” first proposed in the United States, has received widespread attention in recent years. The emergence of precision medicine has brought opportunities and challenges to the field of laboratory medicine and led to higher requirements for accuracy and reliability of results in clinical laboratories.7, 8 Tumor markers have become important clinically in tumor diagnosis, treatment, and monitoring, yet different detection systems often lead to different results. Therefore, efforts are necessary to strengthen the quality control of tumor marker detection to reduce experimental error and provide accurate and reliable data for clinical applications.9 Sigma metric analysis of current laboratory assay values can allow the selection of quality control rules that will effectively reduce the probability of false results, thereby ensuring the output of correct results. When the performance index of an assay reaches a sigma value of 6, simple rules of 13s or 13.5s can be used for quality control.10 A recent study by Hens et al used three different TEa standards [Ricos biological variation, the Clinical Laboratory Improvement Amendments of 1988 (CLIA 88), and RiliBÄK] to calculate sigma values of clinical chemistry assays. The authors showed that the main factors influencing sigma values were instrument performance and TEa standards. Accurate assay results would require control rules that are either simple or complex, based on the sigma.11 A few reports have evaluated the performance of tumor marker assays using sigma metrics. Accordingly, the aim of this study is to determine different sigma values based on four different standards for total allowable error (TEa) to analyze performance differences in the detection of tumor markers, and to explore their application in the quality control of tumor marker detection.

MATERIALS AND METHODS

Total allowable error standards

This study evaluated four different standards for TEa: (a) external quality assessment (EQA) standards of China; (b) quality requirements of the Royal College of Pathologists of Australasia (RCPA); (c) “appropriate” quality standards derived from biological variation12; and (d) standards from the 2015 quality guide created by the German medical laboratory quantitative analysis and quality assessment committee (RiliBÄK).

Tumor markers

The following markers were assessed: total prostate‐specific antigen (tPSA), carcinoembryonic antigen (CEA), alpha‐fetoprotein (AFP), carbohydrate antigen 199 (CA199), carbohydrate antigen 125 (CA125), and carbohydrate antigen 153 (CA153).

Instruments and reagents

The tumor markers were measured by chemiluminescence using the AIA2000 automatic chemiluminescence analyzer and corresponding reagents (TOSOH Co., Ltd., Tokyo, Japan). Calibrator was the original matched reagent. Internal quality control products were obtained from the Randox (lot 1:1575EC; lot 2:1578EC; Crumline, UK). All EQA samples were provided by the National Center for Clinical Laboratories (lots 201711, 201712, 201713, 201714, 201715).

Assay performance evaluation

Evaluation of precision

The coefficient of variation (CV) was used to represent precision. We collected laboratory quality control data from the tumor marker program at our laboratory over 6 months from January 2017 to June 2017 and determined CV values for two concentration levels of quality control products (level 1 and level 2; Table 1).
Table 1

Coefficient of variation, bias, and total allowable error based on four standards for six tumor marker assays

AssayCV (%)BiasTEa (%)
Level 1Level 2EQA standards of ChinaRCPA standardsBiological variabilityRiliBÄK
tPSA6.496.358.42252033.625
CEA5.535.361.24252024.724
AFP5.345.251.31252021.824
CA1996.126.363.5825153924[Link]
CA1254.224.513.55252035.424[Link]
CA1535.125.052.73251520.824

AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; EQA, external quality assessment; RCPA, Royal College of Pathologists of Australasia; TEa, total allowable error; tPSA, total prostate‐specific antigen.

Value for the CA153 standard.

Coefficient of variation, bias, and total allowable error based on four standards for six tumor marker assays AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; EQA, external quality assessment; RCPA, Royal College of Pathologists of Australasia; TEa, total allowable error; tPSA, total prostate‐specific antigen. Value for the CA153 standard.

Evaluation of bias

Percentage difference was used to evaluate bias. Based on first‐time data from the first tumor marker assays performed in 2017, the laboratory obtained a 5‐percent difference. The absolute value of the mean percentage difference was used to evaluate the bias of our laboratory (Table 1).

Calculation of sigma value

Sigma values were calculated using this formula: sigma = [TEa (%) − bias (%)]/CV (%). Sigma values were determined for four different TEa standards and two concentration levels of quality control products.

Calculation of quality goal index

The quality goal index (QGI) was calculated according to this formula: QGI = bias (%)/1.5 × CV (%). We calculated the QGI for each tumor marker at both concentrations of quality control products, noted differences in QGI corresponding to different sigma values, and determined priority measures for quality improvement. When QGI was <0.8, the CV was relatively large, suggesting that the priority should be to improve precision. When QGI was higher than 1.2, the bias was relatively large, suggesting that the priority should be to improve accuracy. A QGI between 0.8 and 1.2 suggested that both precision and accuracy should be improved.

RESULTS

Distribution of sigma metrics for four total allowable error standards

Using the EQA standards of China, the distribution of sigma values for the six tumor marker assays was as follows (Table 2): more than 6, none; 4 to 6, 66.7% (4/6); and less than 4, 33.3% (2/6). Using the RCPA standards, all sigma values were <4. Using the “appropriate” standards, the distribution of sigma values was as follows: more than 6, 16.7% (1/6); 4 to 6, 33.3% (2/6); and less than 4, 50% (3/6). Using the German RiliBÄK standards, the distribution of sigma values was the same as for the EQA standards of China (Table 2).
Table 2

Sigma metrics of six tumor marker assays using four different standards for total allowable error target values

AssayEQA standards of ChinaStandards of RCPABiological variabilityRiliBÄK
Level 1Level 2Level 1Level 2Level 1Level 2Level 1Level 2
tPSA2.552.611.781.823.883.972.552.61
CEA4.34.433.393.54.244.384.124.25
AFP4.444.513.53.563.843.94.254.32
CA1993.53.371.871.85.795.573.343.21
CA1255.084.763.93.657.557.064.854.53
CA1534.354.412.42.433.533.584.154.21

AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; EQA, external quality assessment; RCPA, Royal College of Pathologists of Australasia; tPSA, total prostate specific antigen.

Sigma metrics of six tumor marker assays using four different standards for total allowable error target values AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; EQA, external quality assessment; RCPA, Royal College of Pathologists of Australasia; tPSA, total prostate specific antigen.

Quality corrective actions

Using the EQA standards of China, the QGI was calculated for tumor marker assays with sigma values <6 (ie, all tumor marker assays). The results showed that for tPSA, both precision and accuracy required improvement, whereas for the other markers, only measures to improve precision are required (Table 3).
Table 3

Quality target indices and quality improvement measures of six tumor marker assays according to the Chinese EQA standards

AssaysTEa (%)Sigma valuesQGIPriority measures
Level 1Level 2Level 1Level 2
tPSA252.552.610.860.88Precision and accuracy
CEA254.34.430.150.15Precision
AFP254.444.510.160.17Precision
CA199253.53.370.390.38Precision
CA125255.084.760.560.52Precision
CA153254.354.410.360.36Precision

AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; QGI, quality goal index; TEa, total allowable error; tPSA, total prostate specific antigen.

Quality target indices and quality improvement measures of six tumor marker assays according to the Chinese EQA standards AFP, alpha‐fetoprotein; CA125, carbohydrate antigen 125; CA153, carbohydrate antigen 153; CA199, carbohydrate antigen 199; CEA, carcinoembryonic antigen; CV, coefficient of variation; QGI, quality goal index; TEa, total allowable error; tPSA, total prostate specific antigen.

Measures of internal quality control

Using the EQA standards of China, Westgard sigma rules were used to develop individualized internal quality control measures. The Westgard sigma rules for two levels of controls are shown in Figure 1. The most important aspect is the sigma scale at the bottom of the diagram, which provides guidance for rules that should be applied based on the sigma value of each assay. For CEA, AFP, CA125, and CA153 (with sigma values of 4‐5), 13s/22s/R4s/41s multirules are required, with two batches of two quality control measurements (N = 2, R = 2) or 1 batch of 4 quality control measurements (N = 4, R = 1). The CA199 assay (with sigma values around 3) requires implementation of 13s/22s/R4s/41s/8x multirules, with 2 batches of 4 quality control measurements (N = 4, R = 2) or 4 batches of 2 quality control measurements (N = 2, R = 4). Because the sigma value of tPSA was <3, maximum quality control rules are necessary to ensure quality control (Figure 1).
Figure 1

Westgard sigma rules for two levels of controls. Sigma value = (total allowable error [%] − |bias [%] |)/coefficient of variation (%)

Westgard sigma rules for two levels of controls. Sigma value = (total allowable error [%] − |bias [%] |)/coefficient of variation (%)

DISCUSSION

Our study used four different TEa standards to calculate sigma values of six tumor markers. RCPA standards are the strictest, “appropriate” standards based on biological variability are the least strict, and the EQA standards of China and German RiliBÄK standards exhibit intermediate levels of strictness. Especially for CA199 and CA 125, these two values are assumed values, the same as for CA153 (Table 1). Using the current EQA standards of China, the sigma values of all six tumor markers in our laboratory were <6, indicating that assay performance for these markers requires quality improvement. Using TEa derived from biological variation, Westgard et al13 calculated sigma values of 53 assays, including CA125, CA153, and CA199; their results were consistent with the results of our current study. Westgard et al13 suggested that laboratorians can use Six Sigma tools as aids in choosing high‐quality assays, further contributing to the delivery of high‐quality healthcare for patients. We calculated QGI for all six tumor markers, and the sigma values were less than 6 according to the Chinese EQA standards. Our results showed that all of six assays required quality improvement measures to improve precision, and tPSA also required measures to improve accuracy. Westgard et al (2016) established Westgard multirules, which are compiled in a new quality control tool called “Westgard Sigma Rules.” This tool may provide an individualized program to select quality control rules and determine the appropriate number of quality control measurements.14 Using the EQA standards of China and applying the “Westgard Sigma Rules” based on assay sigma levels, an individualized internal quality control scheme can be developed to reduce the false rejection rate, increase the error detection rate, and provide a continuous, feasible method for improving quality of tumor marker detection. To assess the quality of laboratory testing in the United States, Westgard et al15 used sigma metrics to assess the quality of proficiency testing survey results from several national programs complying with CLIA regulations. Their research showed that sigma values of different analytes are significantly different, and sigma values are mainly distributed from 1.2 to 3.5. These authors therefore suggested that no less than two concentrations of quality control products should be used every day for internal quality control to ensure acceptable test results.15 Xia et al16 also showed that sigma metric is a useful tool to evaluate assay performance and suggested that an assay with a high sigma value could use a simple internal quality control rule, whereas an assay with a low sigma value should be monitored strictly. The conclusions of our study are in agreement with the results of the above scholars, indicating that sigma metrics can play an important role in the field of quality control. Different standards for assessing quality often lead to different sigma values, which are dynamically changing. According to current quality specifications, clinical laboratories should regularly measure detection performance for tumor markers and develop individualized quality control plans for all analytes to provide long‐term, feasible measures for continuous improvement of assay quality. Six Sigma has an important guiding value for evaluating the quality of tumor marker assays.

AUTHOR CONTRIBUTIONS

QL and F.Y conceived and designed the experiments. QL, MF, WL, and C.Y performed the experiments. WZ and C.Z analyzed the data. QL, MF, and F.Y wrote the paper.
  13 in total

Review 1.  Quality control review: implementing a scientifically based quality control system.

Authors:  James O Westgard; Sten A Westgard
Journal:  Ann Clin Biochem       Date:  2015-07-06       Impact factor: 2.057

2.  The quality of laboratory testing today: an assessment of sigma metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance.

Authors:  James O Westgard; Sten A Westgard
Journal:  Am J Clin Pathol       Date:  2006-03       Impact factor: 2.493

Review 3.  Six Sigma Quality Management System and Design of Risk-based Statistical Quality Control.

Authors:  James O Westgard; Sten A Westgard
Journal:  Clin Lab Med       Date:  2016-12-22       Impact factor: 1.935

4.  Application of Sigma Metrics Analysis for the Assessment and Modification of Quality Control Program in the Clinical Chemistry Laboratory of a Tertiary Care Hospital.

Authors:  Sahar Iqbal; Tazeen Mustansar
Journal:  Indian J Clin Biochem       Date:  2016-04-07

5.  Sigma metrics used to assess analytical quality of clinical chemistry assays: importance of the allowable total error (TEa) target.

Authors:  Koen Hens; Mario Berth; Dave Armbruster; Sten Westgard
Journal:  Clin Chem Lab Med       Date:  2014-07       Impact factor: 3.694

6.  Assessing precision, bias and sigma-metrics of 53 measurands of the Alinity ci system.

Authors:  Sten Westgard; Victoria Petrides; Sharon Schneider; Marvin Berman; Jörg Herzogenrath; Anthony Orzechowski
Journal:  Clin Biochem       Date:  2017-09-14       Impact factor: 3.281

7.  [Precision medicine: new opportunities and challenges for molecular epidemiology].

Authors:  Jing Song; Yonghua Hu
Journal:  Zhonghua Liu Xing Bing Xue Za Zhi       Date:  2016-04

8.  Quality specifications of routine clinical chemistry methods based on sigma metrics in performance evaluation.

Authors:  Jun Xia; Su-Feng Chen; Fei Xu; Yong-Lie Zhou
Journal:  J Clin Lab Anal       Date:  2017-06-23       Impact factor: 2.352

9.  Sigma-Metrics of Electrolyte Tests From a Recently Launched New-Generation Proficiency Testing Program of the Korean Association of Quality Assurance for Clinical Laboratory.

Authors:  Hee Jin Huh; Yun Mi Park; Seungok Lee; Chunhwa Ihm; Soyeon Seo; Sang Gon Lee; Joonseok Park; Hae Il Park
Journal:  Ann Lab Med       Date:  2017-07       Impact factor: 3.464

10.  Application of Six Sigma for evaluating the analytical quality of tumor marker assays.

Authors:  Qian Liu; Mei Fu; Fumeng Yang; Wei Liang; Chuanxi Yang; Wenjun Zhu; Liming Ma; Changxin Zhao
Journal:  J Clin Lab Anal       Date:  2018-10-02       Impact factor: 2.352

View more
  4 in total

1.  The application of sigma metrics in the laboratory to assess quality control processes in South Africa.

Authors:  Marli van Heerden; Jaya A George; Siyabonga Khoza
Journal:  Afr J Lab Med       Date:  2022-06-22

2.  Application of a six sigma model to evaluate the analytical performance of urinary biochemical analytes and design a risk-based statistical quality control strategy for these assays: A multicenter study.

Authors:  Qian Liu; Guangrong Bian; Xinkuan Chen; Jingjing Han; Ying Chen; Menglin Wang; Fumeng Yang
Journal:  J Clin Lab Anal       Date:  2021-10-15       Impact factor: 2.352

3.  Application of Six Sigma for evaluating the analytical quality of tumor marker assays.

Authors:  Qian Liu; Mei Fu; Fumeng Yang; Wei Liang; Chuanxi Yang; Wenjun Zhu; Liming Ma; Changxin Zhao
Journal:  J Clin Lab Anal       Date:  2018-10-02       Impact factor: 2.352

4.  Application of Sigma metrics in the quality control strategies of immunology and protein analytes.

Authors:  Yanfen Luo; Xingxing Yan; Qian Xiao; Yifei Long; Jieying Pu; Qiwei Li; Yimei Cai; Yushun Chen; Hongyuan Zhang; Cha Chen; Songbang Ou
Journal:  J Clin Lab Anal       Date:  2021-10-04       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.