| Literature DB >> 28445016 |
Hee Jin Huh1, Yun Mi Park2, Seungok Lee3, Chunhwa Ihm4, Soyeon Seo5, Sang Gon Lee6, Joonseok Park7, Hae Il Park8.
Abstract
Entities:
Mesh:
Substances:
Year: 2017 PMID: 28445016 PMCID: PMC5409015 DOI: 10.3343/alm.2017.37.4.339
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
The analytical performance of electrolyte analyzers according to peer group for six specimens (CC-16-01-CC-16-06) in two episodes of proficiency testing by the Korean Association of Quality Assurance for Clinical Laboratory (KAQACL) in 2016
| Analyte | Mean of 6 PT specimens by direct/indirect ISE (mmol/L)* | Peer group | CV (%) | Bias (%) | SDI | CVR | TE (%) | Sigma-metric | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| French† | SEKK‡ | Belgium§ | Spanish∥ | Minimum Biological¶ | CLIA** | Rili-BAEK†† | RCPA‡‡ | Desirable Biological§§ | |||||||||
| Na+ | TEa (%) | 2.5 | 5.0 | 3.4 | 5.0 | 1.1 | 4 mmol/L | 5.0 | 2.0 (3.0 mmol/L) | 0.7 | |||||||
| Direct ISE | 144.6, 125.7, 125.4, 146.8, 127.1, 146.9 | Nova | 1.4 | 0.0 | −0.03 | 0.78 | 2.6 | 1.80 | 3.65 | 2.46 | 3.65 | 0.75 | 2.29 | 3.65 | 1.70 | 0.46 | |
| Roche | 1.6 | −0.1 | −0.05 | 1.00 | 3.1 | 1.52 | 3.10 | 2.09 | 3.10 | 0.63 | 1.81 | 3.10 | 1.34 | 0.39 | |||
| Siemens | 1.5 | 0.4 | 0.24 | 1.00 | 3.5 | 1.38 | 2.90 | 1.91 | 2.90 | 0.47 | 1.67 | 2.90 | 1.19 | 0.24 | |||
| Indirect ISE | 143.3, 126.2, 125.5, 146, 127.6, 145.9 | Beckman Coulter | 1.1 | −0.7 | −0.54 | 0.89 | 1.5 | 1.56 | 3.75 | 2.36 | 3.75 | 0.34 | 1.89 | 3.75 | 1.27 | 0.03 | |
| Denka Seiken | 1.0 | 0.7 | 0.56 | 0.83 | 2.8 | 1.69 | 3.94 | 2.47 | 3.94 | 0.42 | 2.17 | 3.94 | 1.47 | 0.06 | |||
| Roche | 0.9 | 0.5 | 0.37 | 0.66 | 2.4 | 2.13 | 4.90 | 3.13 | 4.90 | 0.63 | 2.70 | 4.90 | 1.86 | 0.26 | |||
| Sekisui | 1.0 | −0.2 | −0.14 | 0.73 | 1.8 | 2.33 | 4.83 | 3.23 | 4.83 | 0.93 | 2.68 | 4.83 | 1.96 | 0.57 | |||
| Siemens | 0.9 | 1.2 | 0.94 | 0.72 | 3.0 | 1.42 | 4.11 | 2.37 | 4.11 | −0.09 | 2.03 | 4.11 | 1.17 | −0.57 | |||
| K+ | TEa (%) | 4.0 | 8.0 | 4.8 | 8.0 | 8.4 | 0.5 mmol/L | 8.0 | 5.0 (0.2 mmol/L) | 5.6 | |||||||
| Direct ISE | 3.97, 6.01, 6.14, 3.93, 5.99, 3.94 | Nova | 2.5 | 1.8 | 0.77 | 1.06 | 6.6 | 0.91 | 2.54 | 1.24 | 2.54 | 2.71 | 3.53 | 2.54 | 1.32 | 1.57 | |
| Roche | 1.9 | 0.9 | 0.39 | 0.85 | 4.3 | 1.81 | 3.99 | 2.26 | 3.99 | 4.21 | 5.68 | 3.99 | 2.39 | 2.70 | |||
| Siemens | 2.7 | −2.8 | −1.26 | 1.13 | 2.3 | 0.48 | 1.94 | 0.80 | 1.94 | 2.09 | 3.04 | 1.94 | 0.89 | 1.09 | |||
| Indirect ISE | 3.97, 6.13, 6.24, 3.94, 6.12, 3.94 | Beckman Coulter | 1.4 | −1.4 | −0.82 | 0.82 | 1.5 | 1.82 | 4.68 | 2.39 | 4.68 | 4.96 | 4.98 | 4.68 | 2.53 | 2.97 | |
| Denka Seiken | 1.3 | 0.3 | 0.22 | 0.86 | 2.9 | 2.70 | 5.78 | 3.32 | 5.78 | 6.09 | 7.11 | 5.78 | 3.47 | 3.94 | |||
| Roche | 1.8 | 1.1 | 0.70 | 1.00 | 4.5 | 1.67 | 3.96 | 2.13 | 3.96 | 4.19 | 5.92 | 3.96 | 2.25 | 2.59 | |||
| Sekisui | 1.6 | −0.5 | −0.29 | 0.88 | 2.6 | 2.19 | 4.69 | 2.69 | 4.69 | 4.94 | 6.96 | 4.69 | 2.87 | 3.19 | |||
| Siemens | 1.2 | 1.1 | 0.78 | 0.78 | 3.3 | 2.60 | 5.99 | 3.32 | 5.99 | 6.32 | 6.29 | 5.99 | 3.49 | 3.99 | |||
| Cl− | TEa (%) | 4.0 | 7.0 | 5.1 | 9.0 | 2.2 | 5.0 | 8.0 | 3.0 (3.0 mmol/L) | 1.5 | |||||||
| Direct ISE | 102.3, 86, 85.9, 104.9, 88.4, 104.5 | Nova | 2.9 | 1.8 | 0.64 | 1.03 | 7.4 | 0.77 | 1.82 | 1.16 | 2.53 | 0.14 | 1.12 | 2.18 | 0.48 | −0.12 | |
| Roche | 4.4 | 1.0 | 0.32 | 1.40 | 9.5 | 0.67 | 1.36 | 0.92 | 1.84 | 0.26 | 0.90 | 1.61 | 0.48 | 0.10 | |||
| Siemens | 2.7 | 0.2 | 0.06 | 0.94 | 5.0 | 1.30 | 2.44 | 1.72 | 3.19 | 0.65 | 1.68 | 2.82 | 0.98 | 0.41 | |||
| Indirect ISE | 101.5, 83.7, 83.6, 105.4, 86.6, 104.8 | Beckman Coulter | 1.2 | −1.7 | −0.69 | 0.43 | 0.3 | 2.11 | 4.59 | 2.96 | 6.36 | 0.48 | 2.90 | 5.48 | 1.43 | −0.23 | |
| Denka Seiken | 1.3 | −1.1 | −0.46 | 0.55 | 1.8 | 2.08 | 4.25 | 2.86 | 5.80 | 0.77 | 2.79 | 5.03 | 1.41 | 0.29 | |||
| Roche | 1.9 | 0.2 | 0.04 | 0.61 | 4.4 | 1.77 | 3.35 | 2.35 | 4.41 | 0.82 | 2.30 | 3.88 | 1.31 | 0.48 | |||
| Sekisui | 3.1 | 3.4 | 1.20 | 1.29 | 9.6 | 0.18 | 1.13 | 0.53 | 1.78 | −0.40 | 0.50 | 1.46 | −0.07 | −0.62 | |||
| Siemens | 1.4 | −0.6 | −0.21 | 0.48 | 1.7 | 2.16 | 4.14 | 2.87 | 5.62 | 1.05 | 2.80 | 4.88 | 1.77 | 0.57 | |||
*Mean of PT specimen CC-16-01, CC-16-02, CC-16-03, CC-16-04, CC-16-05, and CC-16-06; †French acceptance criteria for external quality assessment (EQA) participants; ‡SEKK Dmax analytical performance specifications; §Belgium EQA performance specifications; ∥Spanish minimum consensus performance specifications; ¶Minimum specifications from biological variation database; **Clinical Laboratory Improvement Amendments (CLIA) requirements for analytical quality; ††Rili-BAEK - German guidelines for quality; ‡‡Royal College of Pathologists of Australasia (RCPA) allowable limits; §§Desirable biological variation database specifications [7].
Abbreviations: SDI, SD index; CVR, CV ratio; TE, total analytical error; TEa, allowable total error; ISE, ion-selective electrode.
Fig. 1Sigma Proficiency Assessment Chart of the Korean Association of Quality Assurance for Clinical Laboratory (KAQACL) proficiency testing (PT) in 2016 for Na+ (A), K+ (B), and Cl− (C). Allowable total error (TEa) values for Na+, K+, and Cl− were 5.0%, 8.0%, and 8.0%, respectively, which were from Rili-BAEK German guidelines for quality. Six points in the same color belong to one of eight peer groups and are from each of six specimens (CC-16-01-CC-16-06). Point size is proportional to the size of the peer group.
Abbreviation: ISE, ion-selective electrode.