Evaluation of miniscrew stability using an automatic embedding auxiliary skeletal anchorage device.

Yasuhiko Oga, Hiroshi Tomonari, Sangho Kwon, Takaharu Kuninori, Takakazu Yagi, Shouichi Miyawaki.   

Abstract

OBJECTIVE: To clarify the in vivo effect of an automatic embedding device on miniscrew stability.
MATERIALS AND METHODS: 42 miniscrews were implanted into rabbit femurs. The miniscrews with the novel auxiliary device formed the auxiliary group (n = 11 at 4 weeks; n = 11 at 8 weeks) and the miniscrews without the auxiliary device formed the nonauxiliary control group (n = 9 at 4 weeks; n = 11 at 8 weeks). Cortical bone thickness, distance from the cortical bone surface to the miniscrew head, and implantation depth of the spike were measured using micro-computed tomography. The mechanical retention force was evaluated by measuring the displacement of the miniscrew head after it was loaded perpendicular to its long axis. In the lateral displacement test, effects of the auxiliary (with vs without auxiliary), and time (4 vs 8 weeks) were assessed using the Brunner-Langer nonparametric analysis of longitudinal data in factorial experiments.
RESULTS: The mean implantation depth of the spike in the auxiliary group at 4 and 8 weeks was 0.28 mm (median: 0.33; SD: 0.12) and 0.37 mm (median: 0.33; SD: 0.19), respectively. The retention force was approximately 2.0 to 2.8 and 1.6 to 1.8 times greater in the auxiliary group than in the nonauxiliary group at 4 and 8 weeks, respectively.
CONCLUSIONS: The auxiliary device improved the mechanical retention force without the need to increase miniscrew length or diameter. This may enable the safe use of miniscrews in difficult areas.

Entities:  

Keywords:  Animal model; Miniscrew; Skeletal anchorage

Mesh:

Year:  2018        PMID: 30203984      PMCID: PMC8137126          DOI: 10.2319/121117-857.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  29 in total

1.  Effects of preservation on the mechanical strength and chemical composition of cortical bone: an experimental study in sheep femora.

Authors:  Jose Moreno; Francisco Forriol
Journal:  Biomaterials       Date:  2002-06       Impact factor: 12.479

Review 2.  Early wound healing around endosseous implants: a review of the literature.

Authors:  Sangeetha Raghavendra; Marjorie C Wood; Thomas D Taylor
Journal:  Int J Oral Maxillofac Implants       Date:  2005 May-Jun       Impact factor: 2.804

3.  Clinical use of miniscrew implants as orthodontic anchorage: success rates and postoperative discomfort.

Authors:  Shingo Kuroda; Yasuyo Sugawara; Toru Deguchi; Hee-Moon Kyung; Teruko Takano-Yamamoto
Journal:  Am J Orthod Dentofacial Orthop       Date:  2007-01       Impact factor: 2.650

4.  Root proximity is a major factor for screw failure in orthodontic anchorage.

Authors:  Shingo Kuroda; Kazuyo Yamada; Toru Deguchi; Takashi Hashimoto; Hee-Moon Kyung; Teruko Takano-Yamamoto
Journal:  Am J Orthod Dentofacial Orthop       Date:  2007-04       Impact factor: 2.650

5.  Effect of screw diameter on orthodontic skeletal anchorage.

Authors:  Chad Morarend; Fang Qian; Steve D Marshall; Karin A Southard; Nicole M Grosland; Teresa A Morgan; Michelle McManus; Thomas E Southard
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-08       Impact factor: 2.650

6.  Effects of the taper shape, dual-thread, and length on the mechanical properties of mini-implants.

Authors:  Young-Kyun Kim; Yoon-Ji Kim; Pil-Young Yun; Jong-Wan Kim
Journal:  Angle Orthod       Date:  2009-09       Impact factor: 2.079

7.  Two different implant designs and impact of related drilling protocols on primary stability in different bone densities: an in vitro comparison study.

Authors:  Lars Sennerby; Luca Pagliani; Anders Petersson; Damiano Verrocchi; Stefano Volpe; Peter Andersson
Journal:  Int J Oral Maxillofac Implants       Date:  2015 May-Jun       Impact factor: 2.804

8.  Pullout strength of miniscrews placed in anterior mandibles of adult and adolescent dogs: a microcomputed tomographic analysis.

Authors:  Zhiqiang Wang; Zhihe Zhao; Jing Xue; Jinlin Song; Feng Deng; Pu Yang
Journal:  Am J Orthod Dentofacial Orthop       Date:  2010-01       Impact factor: 2.650

9.  Stability changes of miniscrew implants over time.

Authors:  Derid S Ure; Donald R Oliver; Ki Beom Kim; Ana Cláudia Melo; Peter H Buschang
Journal:  Angle Orthod       Date:  2011-05-25       Impact factor: 2.079

Review 10.  Failure rates and associated risk factors of orthodontic miniscrew implants: a meta-analysis.

Authors:  Spyridon N Papageorgiou; Ioannis P Zogakis; Moschos A Papadopoulos
Journal:  Am J Orthod Dentofacial Orthop       Date:  2012-11       Impact factor: 2.650

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.