Effects of the taper shape, dual-thread, and length on the mechanical properties of mini-implants.

Young-Kyun Kim1, Yoon-Ji Kim, Pil-Young Yun, Jong-Wan Kim.   

Abstract

OBJECTIVE: To analyze the mechanical effects of the length and the various shapes such as cylindrical shape, taper shape, and dual-thread shape on the insertion and removal torque of mini-implants.
MATERIALS AND METHODS: Mini-implants (diameter 1.6 mm and length 6 mm and 8 mm) consisting of cylindrical, taper, and dual-thread groups were inserted and removed in Sawbones while measuring the torque and time. Mechanical analysis was done of maximum insertion torque (MIT), maximum removal torque (MRT), torque ratio (TR; MRT/MIT), insertion angular momentum (IAM), removal angular momentum (RAM), and time of MIT. Measurements were statistically evaluated to analyze any differences of shapes and lengths.
RESULTS: The cylindrical shape had the lowest MIT and MRT in each length. Although taper shape showed the highest MIT in each length, dual-thread shape showed significantly higher MRT, TR, and RAM in each length (P < .05). Dual-thread groups showed a gentle increase of insertion torque and a gentle decrease of removal torque in contrast to the other shape groups. However, it had higher IAM and time of MIT. The long length group showed significantly higher measurements except for TR.
CONCLUSIONS: Dual-thread shape provided better mechanical stability with high removal torque on the broad range than other shapes. However, dual-thread shape may need improvement for reducing the long insertion time to decrease the stress to the surrounding tissue.

Mesh:

Substances:

Year:  2009        PMID: 19705930     DOI: 10.2319/071808-374.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  18 in total

1.  Morphological and structural characteristics of orthodontic mini-implants.

Authors:  Saeed AlSamak; Elias Bitsanis; Margarita Makou; George Eliades
Journal:  J Orofac Orthop       Date:  2012-01-12       Impact factor: 1.938

2.  Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro Study.

Authors:  Aniruddh V Yashwant; S Dilip; R Krishnaraj; K Ravi
Journal:  J Clin Diagn Res       Date:  2017-05-01

3.  Evaluation of miniscrew stability using an automatic embedding auxiliary skeletal anchorage device.

Authors:  Yasuhiko Oga; Hiroshi Tomonari; Sangho Kwon; Takaharu Kuninori; Takakazu Yagi; Shouichi Miyawaki
Journal:  Angle Orthod       Date:  2018-09-01       Impact factor: 2.079

4.  Stability comparison between commercially available mini-implants and a novel design: part 1.

Authors:  Christine Hong; Haofu Lee; Richard Webster; Jinny Kwak; Benjamin M Wu; Won Moon
Journal:  Angle Orthod       Date:  2011-02-09       Impact factor: 2.079

5.  Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2.

Authors:  Christine Hong; Peter Truong; Ha Na Song; Benjamin M Wu; Won Moon
Journal:  Angle Orthod       Date:  2011-06-29       Impact factor: 2.079

6.  Significance of bone morphology and quality on the primary stability of orthodontic mini-implants: in vitro comparison between human bone substitute and artificial bone.

Authors:  Sachin Chhatwani; Ouafaa Kouji-Diehl; Kristian Kniha; Ali Modabber; Frank Hölzle; Jozsef Szalma; Gholamreza Danesh; Stephan Christian Möhlhenrich
Journal:  J Orofac Orthop       Date:  2022-03-18       Impact factor: 1.938

7.  Effect of rotation bump on removal torque of orthodontic mini-implants.

Authors:  Odontuya Gansukh; Jong-Wha Jeong; Jong-Wan Kim; Young-Kyun Kim; Jong-Ho Lee; Tae-Woo Kim
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2013-12-23

8.  Effects of predrilling depth and implant shape on the mechanical properties of orthodontic mini-implants during the insertion procedure.

Authors:  Keun-Chul Cho; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2011-11-03       Impact factor: 2.079

9.  Root and bone response to the proximity of a mini-implant under orthodontic loading.

Authors:  Yang-Ku Lee; Jong-Wan Kim; Seung-Hak Baek; Tae-Woo Kim; Young-Il Chang
Journal:  Angle Orthod       Date:  2010-05       Impact factor: 2.079

10.  The effect of the thread depth on the mechanical properties of the dental implant.

Authors:  Sun-Young Lee; Sung-Jun Kim; Hyun-Wook An; Hyun-Seung Kim; Dong-Guk Ha; Kyung-Ho Ryo; Kwang-Bum Park
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.