Abstract
The purpose of this study was to prospectively evaluate the image quality and the diagnostic ability of low tube voltage and reduced contrast material dose hepatic dynamic computed tomography (CT) reconstructed with adaptive iterative dose reduction 3-dimensional (AIDR 3D).Eighty-nine patients underwent hepatic dynamic CT using one of the 2 protocols: tube voltage of 120 kVp, contrast dose of 600 mgI/kg, and filtered back projection in Protocol A (n = 46), and tube voltage of 100 kVp, contrast dose of 500 mgI/kg, and AIDR 3D in Protocol B (n = 43). The volume CT dose index (CTDIvol) and size-specific dose estimates (SSDEs) were compared between the 2 groups. Objective image noise and tumor to liver contrast-to-noise ratio (CNR) were also compared. Three radiologists independently reviewed image quality. The jackknife alternative free-response receiver-operating characteristic (JAFROC) analysis was performed to compare diagnostic performance.The mean CTDIvol and SSDE of Protocol B (14.3 and 20.2, respectively) were significantly lower than those of Protocol A (22.1 and 31.4, P < .001). There were no significant differences in either objective image noise or CNR. In the qualitative analysis, 2 readers assigned significant lower scores to images of Protocol B for at least one of the 3 phases regarding overall image quality (P < .05). There was no significant difference in the JAFROC1 figure of merit between protocols.Low tube voltage CT with AIDR 3D yielded a reduction in radiation dose and in the amount of contrast material while maintaining diagnostic performance.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30142778 PMCID: PMC6113013 DOI: 10.1097/MD.0000000000011857
Source DB: PubMed Journal: Medicine (Baltimore) ISSN: 0025-7974 Impact factor: 1.817
Acquisition and reconstruction parameters of computed tomography examination.
Mean CTDIvol, DLP, and SSDE of each protocol.
Mean CT number of the liver, image noise, and CNR for each protocol.
Mean visual scores of image noise, image sharpness, and over all image quality.
Figure 1A male in his sixties who underwent hepatic dynamic computed tomography with both Protocols A and B within 3 months. The conspicuity of 2 hepatocellular carcinomas (arrows) in arterial phase images (A: Protocol A; B: Protocol B) and equilibrium phase images (C: Protocol A; D: Protocol B) is almost equivalent for these 2 protocols. Streak artifacts caused by ribs and spine are slightly more prominent in Protocol B (arrowhead).
Figure of merit, sensitivity, and positive predictive value for the detection of hypervascular hepatocellular carcinoma.
Figure 2Initial computed tomography (CT) (A: Protocol B) and follow-up CT (B: Protocol A) images of a female in her seventies with hepatocellular carcinoma. The contrast between the hepatocellular carcinoma and the liver parenchyma is almost equivalent for both protocols. A faint blotchy appearance may be seen in Protocol B.