Capillary and Venous Blood Glucose Accuracy in Blood Glucose Meters Versus Reference Standards: The Impact of Study Design on Accuracy Evaluations.

Kirsty Macleod1, Laurence B Katz2, Hilary Cameron1.   

Abstract

BACKGROUND: Anecdotal blood glucose assessments conducted by health care professionals (HCPs) in the field have highlighted differences in results when methodology used is not according to best practices for measuring blood glucose. This study assessed the impact on accuracy of blood glucose measurements when methodology deviates from the recommended study design and recommended reference instrument.
METHODS: Adults with type 1 or type 2 diabetes provided capillary and venous blood samples for accuracy assessments using OneTouch® Verio® (Verio) and OneTouch® Ultra 2® (Ultra) blood glucose meters (BGM) and two different reference instruments.
RESULTS: Increases in mean bias were observed when comparing capillary to venous samples tested on the BGMs and the recommended reference instrument. Mean bias was even greater when a hospital blood glucose analyzer was used to measure venous plasma glucose. Increases in mean bias observed for Ultra BGM when testing venous blood on the meter compared to the recommended reference instrument was likely due to the interfering effects of low oxygen levels in the venous blood sample. Conversely, Verio meters, which are insensitive to low oxygen levels, showed little difference from baseline when testing venous blood on the meter compared to results from the same venous sample measured on a reference instrument.
CONCLUSIONS: Deviations from the best practice study design of comparing capillary blood glucose results tested on the blood glucose meter with the manufacturer's stated reference instrument will affect accuracy of blood glucose measurements.

Entities:  

Keywords:  Ultra; Verio; YSI; blood glucose accuracy; blood glucose monitor

Mesh:

Substances:

Year:  2018        PMID: 30084263      PMCID: PMC6501528          DOI: 10.1177/1932296818790228

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  9 in total

Review 1.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

Review 2.  Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems.

Authors:  Gary H Thorpe
Journal:  Diabetes Technol Ther       Date:  2013-02-15       Impact factor: 6.118

3.  Successful Performance of Laboratory Investigations with Blood Glucose Meters Employing a Dynamic Electrochemistry-Based Correction Algorithm Is Dependent on Careful Sample Handling.

Authors:  Filiz Demircik; David Klonoff; Petra B Musholt; Sanja Ramljak; Andreas Pfützner
Journal:  Diabetes Technol Ther       Date:  2016-10       Impact factor: 6.118

4.  Impact of prandial status on the comparison of capillary glucose meter and venous plasma glucose measurements in healthy volunteers.

Authors:  Akhilesh Swaminathan; Helen Lunt; Winston S J Chang; Florence J Logan; Christopher M A Frampton; Christopher M Florkowski
Journal:  Ann Clin Biochem       Date:  2012-11-12       Impact factor: 2.057

5.  Do the New FDA Guidance Documents Help Improving Performance of Blood Glucose Monitoring Systems Compared With ISO 15197?

Authors:  Guido Freckmann; Annette Baumstark; Stefan Pleus
Journal:  J Diabetes Sci Technol       Date:  2017-06-06

6.  Assessing glucose monitor performance--a standardized approach.

Authors:  John J Mahoney; John M Ellison
Journal:  Diabetes Technol Ther       Date:  2007-12       Impact factor: 6.118

7.  Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640.

Authors:  P J Twomey
Journal:  J Clin Pathol       Date:  2004-07       Impact factor: 3.411

8.  Sodium fluoride: an ineffective inhibitor of glycolysis.

Authors:  Raymond Gambino
Journal:  Ann Clin Biochem       Date:  2012-11-05       Impact factor: 2.057

Review 9.  Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge.

Authors:  Michael Erbach; Guido Freckmann; Rolf Hinzmann; Bernhard Kulzer; Ralph Ziegler; Lutz Heinemann; Oliver Schnell
Journal:  J Diabetes Sci Technol       Date:  2016-08-22
  9 in total
  4 in total

1.  Why the Details of Glucose Meter Evaluations Matters.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2018-10-17

2.  Analysis of "Capillary and Venous Blood Glucose Accuracy in Blood Glucose Meters Versus Reference Standards: The Impact of Study Design on Accuracy Evaluations".

Authors:  Gary H Thorpe
Journal:  J Diabetes Sci Technol       Date:  2018-09-15

3.  Evidence From a Long-Term, Systematic Post-Market Surveillance Program: Clinical Performance of a Hematocrit-Insensitive Blood Glucose Test Strip.

Authors:  Steven Setford; Stuart Phillips; Mike Grady
Journal:  J Diabetes Sci Technol       Date:  2019-02-07

Review 4.  Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.

Authors:  Guido Freckmann; Stefan Pleus; Mike Grady; Steven Setford; Brian Levy
Journal:  J Diabetes Sci Technol       Date:  2018-11-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.