Ryan M Jones1, Lulu Deng1, Kogee Leung1, Dallan McMahon1,2, Meaghan A O'Reilly1,2, Kullervo Hynynen1,2,3.
Abstract
Focused ultrasound (FUS)-mediated blood-brain barrier (BBB) opening recently entered clinical testing for targeted drug delivery to the brain. Sources of variability exist in the current procedures, motivating the development of real-time monitoring and control techniques to improve treatment safety and efficacy. Here we used three-dimensional (3D) transcranial microbubble imaging to calibrate FUS exposure levels for volumetric BBB opening.Entities:
Keywords: blood-brain barrier; focused ultrasound; microbubble contrast agents; three-dimensional acoustic imaging; transmit/receive phased array
Mesh:
Substances:
Year: 2018 PMID: 29896293 PMCID: PMC5996357 DOI: 10.7150/thno.24911
Source DB: PubMed Journal: Theranostics ISSN: 1838-7640 Impact factor: 11.556
Transducer element dimensions of multi-frequency dual-mode array. Transducer element dimensions and center frequencies.
| Frequency | Outer Diameter (mm) | Inner Diameter (mm) | Height (mm) |
|---|---|---|---|
| 10 | 7 | 6 | |
| 5 | 3.5 | 3 | |
| 2.5 | 1.75 | 1.5 |
MRI parameters. T1w, T2w, and T2*w MRI sequence types and parameters employed in this study.
| T1w | T2w | T2*w | |
|---|---|---|---|
| turbo spin echo | turbo spin echo | 3D gradient echo | |
| 8.6 | 82 | 15 | |
| 500 | 3100 | 27 | |
| 4 | 8 | 1 | |
| 3 | 2 | 2 | |
| 10 × 10 | 10 × 10 | 10 × 10 × 2.4 | |
| 256 × 256 | 256 × 256 | 256 × 256 × 16 | |
| 1.5 | 1.5 | 1.5 | |
| ±62.7 | ±15.4 | ±33.3 | |
| 150 | 150 | 13 |
Summary of acoustic imaging data for pulses with subharmonic activity. The peak negative pressure subharmonic threshold (psub) values are in-situ SPTP estimates. The image metric values (positional error, -3 dB main lobe beamwidths) are based on offline, retrospective reconstructions (FOV = 10 mm × 10 mm × 10 mm, voxel size = 0.25 mm × 0.25 mm × 0.25 mm). Positional error is defined as the distance between the location of SPTA source field intensity and the intended target. The main lobe beamwidths (short/long axis sizes) are calculated based on 3D ellipsoidal fits of -3 dB source field intensity isosurfaces.
| Mean ± SD | Minimum | Maximum | |
|---|---|---|---|
| 0.68 ± 0.15 | 0.40 | 1.00 | |
| 47 ± 10 | 30 | 65 | |
| 10 ± 4 | 4 | 18 | |
| 89 ± 4 | 78 | 97 | |
| 0.7 ± 0.5 | 0 | 3.0 | |
| 3.0 ± 0.5 | 2.2 | 3.9 | |
| 5.4 ± 0.5 | 4.6 | 6.6 |
Summary of MRI and histological findings from treatment group. Individual focal spot analysis of MRI and histological data from the treatment group. 5 animals each received 4 non-overlapping 36-point grid exposures at the 50% target level (i.e., N = 720 focal spots total, N = 20 calibration points).
| Time Post-FUS | CE-T1w Hyper | T2w Hyper | T2*w Hypo | RBC Extravs | |
|---|---|---|---|---|---|
| 0 days | 336 (46.6%) | 12 (1.6%) | 0 | N/A | |
| 1 week | 0 | 0 | 0 | 1 (0.1%) | |
| 0 days | 20 (100%) | 0 | 0 | N/A | |
| 1 week | 0 | 0 | 0 | 0 |
Hyper: hyperintense; Hypo: hypointense; Extravs: extravasations.