Multi-channel pre-beamformed data acquisition system for research on advanced ultrasound imaging methods.

Chris C P Cheung, Alfred C H Yu, Nazila Salimi, Billy Y S Yiu, Ivan K H Tsang, Benjamin Kerby, Reza Zahiri Azar, Kris Dickie.   

Abstract

The lack of open access to the pre-beamformed data of an ultrasound scanner has limited the research of novel imaging methods to a few privileged laboratories. To address this need, we have developed a pre-beamformed data acquisition (DAQ) system that can collect data over 128 array elements in parallel from the Ultrasonix series of research-purpose ultrasound scanners. Our DAQ system comprises three system-level blocks: 1) a connector board that interfaces with the array probe and the scanner through a probe connector port; 2) a main board that triggers DAQ and controls data transfer to a computer; and 3) four receiver boards that are each responsible for acquiring 32 channels of digitized raw data and storing them to the on-board memory. This system can acquire pre-beamformed data with 12-bit resolution when using a 40-MHz sampling rate. It houses a 16 GB RAM buffer that is sufficient to store 128 channels of pre-beamformed data for 8000 to 25 000 transmit firings, depending on imaging depth; corresponding to nearly a 2-s period in typical imaging setups. Following the acquisition, the data can be transferred through a USB 2.0 link to a computer for offline processing and analysis. To evaluate the feasibility of using the DAQ system for advanced imaging research, two proof-of-concept investigations have been conducted on beamforming and plane-wave B-flow imaging. Results show that adaptive beamforming algorithms such as the minimum variance approach can generate sharper images of a wire cross-section whose diameter is equal to the imaging wavelength (150 μm in our example). Also, planewave B-flow imaging can provide more consistent visualization of blood speckle movement given the higher temporal resolution of this imaging approach (2500 fps in our example).

Entities:  

Year:  2012        PMID: 24626032     DOI: 10.1109/TUFFC.2012.2184

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  11 in total

1.  Comparison of analytical and numerical approaches for CT-based aberration correction in transcranial passive acoustic imaging.

Authors:  Ryan M Jones; Kullervo Hynynen
Journal:  Phys Med Biol       Date:  2015-11-25       Impact factor: 3.609

2.  Experimental demonstration of passive acoustic imaging in the human skull cavity using CT-based aberration corrections.

Authors:  Ryan M Jones; Meaghan A O'Reilly; Kullervo Hynynen
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

3.  Transcranial passive acoustic mapping with hemispherical sparse arrays using CT-based skull-specific aberration corrections: a simulation study.

Authors:  Ryan M Jones; Meaghan A O'Reilly; Kullervo Hynynen
Journal:  Phys Med Biol       Date:  2013-06-27       Impact factor: 3.609

Review 4.  Photoacoustic-guided surgery from head to toe [Invited].

Authors:  Alycen Wiacek; Muyinatu A Lediju Bell
Journal:  Biomed Opt Express       Date:  2021-03-16       Impact factor: 3.732

5.  SUPRA: open-source software-defined ultrasound processing for real-time applications : A 2D and 3D pipeline from beamforming to B-mode.

Authors:  Rüdiger Göbl; Nassir Navab; Christoph Hennersperger
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-03-28       Impact factor: 2.924

6.  ULA-OP 256: A 256-Channel Open Scanner for Development and Real-Time Implementation of New Ultrasound Methods.

Authors:  Enrico Boni; Luca Bassi; Alessandro Dallai; Francesco Guidi; Valentino Meacci; Alessandro Ramalli; Stefano Ricci; Piero Tortoli
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2016-05-11       Impact factor: 2.725

7.  Ultrasound Open Platforms for Next-Generation Imaging Technique Development.

Authors:  Enrico Boni; Alfred C H Yu; Steven Freear; Jorgen Arendt Jensen; Piero Tortoli
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-07       Impact factor: 2.725

8.  Quest for the Vulnerable Atheroma: Carotid Stenosis and Diametric Strain--A Feasibility Study.

Authors:  Canxing Xu; Chun Yuan; Edward Stutzman; Gador Canton; Keith A Comess; Kirk W Beach
Journal:  Ultrasound Med Biol       Date:  2015-12-23       Impact factor: 2.998

9.  Three-dimensional transcranial microbubble imaging for guiding volumetric ultrasound-mediated blood-brain barrier opening.

Authors:  Ryan M Jones; Lulu Deng; Kogee Leung; Dallan McMahon; Meaghan A O'Reilly; Kullervo Hynynen
Journal:  Theranostics       Date:  2018-04-16       Impact factor: 11.556

10.  Performance characteristics of an interventional multispectral photoacoustic imaging system for guiding minimally invasive procedures.

Authors:  Wenfeng Xia; Daniil I Nikitichev; Jean Martial Mari; Simeon J West; Rosalind Pratt; Anna L David; Sebastien Ourselin; Paul C Beard; Adrien E Desjardins
Journal:  J Biomed Opt       Date:  2015-08       Impact factor: 3.170