Literature DB >> 29771822

Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing.

Mona Ascha1, Anirban Bhattacharyya2, Jose A Ramos2, Adriano R Tonelli2.   

Abstract

INTRODUCTION/
PURPOSE: Peripheral capillary oxygen saturation (SpO2) is used as surrogate for arterial blood oxygen saturation. We studied the degree of discrepancy between SpO2 and arterial oxygen (SaO2) and identified parameters that may explain this difference.
METHODS: We included patients who underwent cardiopulmonary exercise testing at Cleveland Clinic. Pulse oximeters with forehead probes measured SpO2 and arterial blood gas samples provided the SaO2 both at rest and peak exercise.
RESULTS: We included 751 patients, 54 ± 16 yr old with 53% of female gender. Bland-Altman analysis revealed a bias of 3.8% with limits of agreement of 0.3% to 7.9% between SpO2 and SaO2 at rest. A total of 174 (23%) patients had SpO2 ≥ 5% of SaO2, and these individuals were older, current smokers with lower forced expiratory volume in the first second and higher partial pressure of carbon dioxide and carboxyhemoglobin. At peak exercise (n = 631), 75 (12%) SpO2 values were lower than the SaO2 determinations reflecting difficulties in the SpO2 measurement in some patients. The bias between SpO2 and SaO2 was 2.6% with limits of agreement between -2.9% and 8.1%. Values of SpO2 ≥ 5% of SaO2 (n = 78, 12%) were associated with the significant resting variables plus lower heart rate, oxygen consumption, and oxygen pulse. In multivariate analyses, carboxyhemoglobin remained significantly associated with the difference between SpO2 and SaO2 both at rest and peak exercise.
CONCLUSIONS: In the present study, pulse oximetry commonly overestimated the SaO2. Increased carboxyhemoglobin levels are independently associated with the difference between SpO2 and SaO2, a finding particularly relevant in smokers.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29771822      PMCID: PMC6138536          DOI: 10.1249/MSS.0000000000001658

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  25 in total

1.  Standardisation of spirometry.

Authors:  M R Miller; J Hankinson; V Brusasco; F Burgos; R Casaburi; A Coates; R Crapo; P Enright; C P M van der Grinten; P Gustafsson; R Jensen; D C Johnson; N MacIntyre; R McKay; D Navajas; O F Pedersen; R Pellegrino; G Viegi; J Wanger
Journal:  Eur Respir J       Date:  2005-08       Impact factor: 16.671

2.  Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry.

Authors:  H M Sami; B S Kleinman; V A Lonchyna
Journal:  J Clin Monit       Date:  1991-10

3.  Low Pulse Oximetry Reading: Time for Action or Reflection?

Authors:  Andrew R Tomlinson; Benjamin D Levine; Tony G Babb
Journal:  Chest       Date:  2017-04       Impact factor: 9.410

4.  Range of accuracy of two wavelength oximetry.

Authors:  K R Chapman; F L Liu; R M Watson; A S Rebuck
Journal:  Chest       Date:  1986-04       Impact factor: 9.410

5.  Accuracy of pulse oximetry in the intensive care unit.

Authors:  A Van de Louw; C Cracco; C Cerf; A Harf; P Duvaldestin; F Lemaire; L Brochard
Journal:  Intensive Care Med       Date:  2001-10       Impact factor: 17.440

6.  Hemoglobin oxygen saturation discrepancy using various methods in patients with sickle cell vaso-occlusive painful crisis.

Authors:  Shahid Ahmed; Anita K Siddiqui; Cristina P Sison; Rabia K Shahid; Joseph Mattana
Journal:  Eur J Haematol       Date:  2005-04       Impact factor: 2.997

Review 7.  Potential errors in pulse oximetry. II. Effects of changes in saturation and signal quality.

Authors:  R K Webb; A C Ralston; W B Runciman
Journal:  Anaesthesia       Date:  1991-03       Impact factor: 6.955

8.  Transesophageal versus surface pulse oximetry in intensive care unit patients.

Authors:  M N Vicenzi; H Gombotz; H Krenn; C Dorn; P Rehak
Journal:  Crit Care Med       Date:  2000-07       Impact factor: 7.598

9.  Accuracy of two pulse oximeters during maximal cycling exercise.

Authors:  R J Wood; C J Gore; A G Hahn; K I Norton; G C Scroop; D P Campbell; D B Watson; D L Emonson
Journal:  Aust J Sci Med Sport       Date:  1997-06

10.  Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation.

Authors:  B G Carter; J B Carlin; J Tibballs; H Mead; M Hochmann; A Osborne
Journal:  Crit Care Med       Date:  1998-06       Impact factor: 7.598

View more
  4 in total

1.  Role of the active cycle of breathing technique combined with phonophoresis for the treatment of patients with chronic obstructive pulmonary disease (COPD): study protocol for a preliminary randomized controlled trial.

Authors:  M D Shen; L R Guo; Y W Li; R T Gao; X Sui; Z Du; L Q Xu; H Y Shi; Y Y Ni; X Zhang; Y Pang; W Zhang; T Z Yu; F Li
Journal:  Trials       Date:  2021-03-23       Impact factor: 2.279

2.  Is pulse oximeter a reliable tool for non-critically ill patients with COVID-19?

Authors:  Aslıhan Gürün Kaya; Miraç Öz; İrem Akdemir Kalkan; Ezgi Gülten; Güle Çınar; Alpay Azap; Akın Kaya
Journal:  Int J Clin Pract       Date:  2021-10-21       Impact factor: 3.149

3.  Oxygen Saturation Behavior by Pulse Oximetry in Female Athletes: Breaking Myths.

Authors:  Pilar Martín-Escudero; Ana María Cabanas; Manuel Fuentes-Ferrer; Mercedes Galindo-Canales
Journal:  Biosensors (Basel)       Date:  2021-10-14

4.  Effects of a taped filter mask on peak power, perceived breathlessness, heart rate, blood lactate and oxygen saturation during a graded exercise test in young healthy adults: a randomized controlled trial.

Authors:  Hoi Lam Ng; Johannes Trefz; Martin Schönfelder; Henning Wackerhage
Journal:  BMC Sports Sci Med Rehabil       Date:  2022-02-07
  4 in total