Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation.

B G Carter1, J B Carlin, J Tibballs, H Mead, M Hochmann, A Osborne.   

Abstract

OBJECTIVE: To evaluate the performance of two pulse oximeters in the measurement of arterial hemoglobin saturation in hypoxemic children.
DESIGN: Prospective, repeated-measures observational study.
SETTING: A 16-bed pediatric intensive care unit in a children's tertiary hospital. PATIENTS: Sixty-six patients with arterial saturation of <90%.
INTERVENTIONS: Three arterial blood samples were taken from each subject during a 48-hr period. Pulse oximeter measurements of arterial saturation were compared with arterial saturation determined by cooximetry.
MEASUREMENTS AND MAIN RESULTS: Arterial saturation was measured using one or both pulse oximeters (SpO2) and compared with the arterial hemoglobin saturation determined by cooximetry (SaO2). Sixty-two subjects were studied, using the Ohmeda pulse oximeter giving 185 data points (78 with saturations <75% [defined by the average of pulse oximeter and cooximeter]); 53 subjects were studied, using the Hewlett-Packard pulse oximeter yielding 155 data points (60 with saturations <75%). SpO2 ranged from 24% to 94%. Bias and precision of the Ohmeda pulse oximeter were -2.8% and 4.8% >75% and -0.8% and 8.0% <75%. Bias and precision of the Hewlett-Packard pulse oximeter were -0.5% and 5.1% >75% and 0.4% and 4.6% <75%. Intrapatient regression coefficient (r) for the differences between pulse oximeter and cooximeter was 0.58 for the Ohmeda and 0.59 for the Hewlett-Packard. Regression coefficients for predicting change in cooximeter value given a change in the Ohmeda pulse oximeter were 0.59 and 0.71 <75% and >75%, respectively. Similar coefficients for the Hewlett-Packard pulse oximeter were 0.50 and 0.70, respectively.
CONCLUSION: The performance of the Ohmeda pulse oximeter deteriorated below an SpO2 of 75%. The Hewlett-Packard pulse oximeter performed consistently above and below an SpO2 of 75%. The ability of both pulse oximeters to reliably predict change in SaO2 based on change in pulse oximetry was limited. We recommend measurement of PaO2 or SaO2 for important clinical decisions.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9635666     DOI: 10.1097/00003246-199806000-00040

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  9 in total

1.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

2.  Evaluation of a pulse oximeter sensor tester.

Authors:  Shuba Dugani; Iljaz Hodzovic; Seema Sindhakar; Aida Nadra; Clare Dunstan; Anthony Richard Wilkes; John Mecklenburgh
Journal:  J Clin Monit Comput       Date:  2011-05-18       Impact factor: 2.502

3.  Comparative Evaluation of Accuracy of Pulse Oximeters and Factors Affecting Their Performance in a Tertiary Intensive Care Unit.

Authors:  Anupam Kumar Singh; Malvinder Singh Sahi; Bablesh Mahawar; Sajjan Rajpurohit
Journal:  J Clin Diagn Res       Date:  2017-06-01

4.  Assessment of a Non Invasive Brain Oximeter in Volunteers Undergoing Acute Hypoxia.

Authors:  Barry Dixon; David B MacLeod
Journal:  Med Devices (Auckl)       Date:  2020-06-30

5.  Oxyhemoglobin concentrations do not support hemoglobinopathy in COVID-19.

Authors:  Patrick Maher; Hamna Zafar; Kusum Mathews
Journal:  Respir Med       Date:  2021-08-30       Impact factor: 4.582

6.  Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing.

Authors:  Mona Ascha; Anirban Bhattacharyya; Jose A Ramos; Adriano R Tonelli
Journal:  Med Sci Sports Exerc       Date:  2018-10       Impact factor: 5.411

7.  Pulse oximeter accuracy and precision at five different sensor locations in infants and children with cyanotic heart disease.

Authors:  Jyotirmoy Das; Amit Aggarwal; Naresh Kumar Aggarwal
Journal:  Indian J Anaesth       Date:  2010-11

8.  Assessment of a Non-Invasive Brain Oximeter in a Sheep Model of Acute Brain Injury.

Authors:  Barry Dixon; Renee Turner; Chris Christou
Journal:  Med Devices (Auckl)       Date:  2019-12-03

9.  Monitoring of limb perfusion after vascular surgery in critical limb ischemia using near-infrared spectroscopy: A prospective observational study.

Authors:  Tanveer Singh Kundra; Ashwini Thimmarayappa; Sunder Singh Subash; Parminder Kaur
Journal:  Ann Card Anaesth       Date:  2020 Oct-Dec
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.