Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems.

Khanlian Chung1, Lothar R Schad1, Frank G Zöllner2.   

Abstract

PURPOSE: Cone beam computed tomography (CBCT) systems offer physicians crucial 3D and 2D imaging capabilities during interventions. However, certain medical applications only require very specific information from the CBCTs (e.g., determination of the position of high-contrast objects). In diagnostics, tomosynthesis techniques can be used in these cases to minimize dose exposure. Therefore, integrating such techniques on CBCT systems could also be beneficial for interventions. In this paper, we investigate the performance of our implementation of circular tomosynthesis on a CBCT device.
METHODS: The tomosynthesis scan trajectory is realized with step-and-shoot on a clinical C-arm device. The online calibration algorithm uses conventionally acquired 3D CBCT of the scanned object as prior knowledge to correct the imaging geometries. The online calibration algorithm was compared to an offline calibration to test its performance. A ball bearing phantom was used to evaluate the reconstructions with respect to geometric distortions. The evaluation was done for three different scenarios to test the robustness of our tomosynthesis implementation against object deviations (e.g., pen) and different object positioning.
RESULTS: The circular tomosynthesis was tested on a ball bearing and an anthropomorphic phantom. The results show that the calibration is robust against isocenter shifts and object deviations in the CBCT. All reconstructions used 100 projections and displayed limited angle artifacts. The accuracy of the positions and shapes of high-contrast objects were, however, determined precisely. (The maximal center position deviation is 0.31 mm.)
CONCLUSION: For medical procedures that primarily determine the precise position of high-contrast objects, circular tomosynthesis could offer an approach to reduce dose exposure.

Keywords:  Arbitrary scan trajectories; C-arm systems; Circular tomosynthesis; Cone beam computed tomography; Image-guided intervention; Online calibration

Mesh:

Year:  2018        PMID: 29740752     DOI: 10.1007/s11548-018-1782-y

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  16 in total

Review 1.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

2.  Prior image constrained compressed sensing: implementation and performance evaluation.

Authors:  Pascal Thériault Lauzier; Jie Tang; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

3.  A novel extension of the parallel-beam projection-slice theorem to divergent fan-beam and cone-beam projections.

Authors:  Guang-Hong Chen; Shuai Leng; Charles A Mistretta
Journal:  Med Phys       Date:  2005-03       Impact factor: 4.071

4.  A geometric calibration method for cone beam CT systems.

Authors:  Kai Yang; Alexander L C Kwan; DeWitt F Miller; John M Boone
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

5.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

6.  Digital breast tomosynthesis: observer performance study.

Authors:  David Gur; Gordon S Abrams; Denise M Chough; Marie A Ganott; Christiane M Hakim; Ronald L Perrin; Grace Y Rathfon; Jules H Sumkin; Margarita L Zuley; Andriy I Bandos
Journal:  AJR Am J Roentgenol       Date:  2009-08       Impact factor: 3.959

7.  An optimization-based method for geometrical calibration in cone-beam CT without dedicated phantoms.

Authors:  D Panetta; N Belcari; A Del Guerra; S Moehrs
Journal:  Phys Med Biol       Date:  2008-06-26       Impact factor: 3.609

8.  Fast and flexible X-ray tomography using the ASTRA toolbox.

Authors:  Wim van Aarle; Willem Jan Palenstijn; Jeroen Cant; Eline Janssens; Folkert Bleichrodt; Andrei Dabravolski; Jan De Beenhouwer; K Joost Batenburg; Jan Sijbers
Journal:  Opt Express       Date:  2016-10-31       Impact factor: 3.894

9.  Online geometric calibration of cone-beam computed tomography for arbitrary imaging objects.

Authors:  Yuanzheng Meng; Hui Gong; Xiaoquan Yang
Journal:  IEEE Trans Med Imaging       Date:  2012-10-12       Impact factor: 10.048

10.  A geometric calibration method for inverse geometry computed tomography using P-matrices.

Authors:  Jordan M Slagowski; David A P Dunkerley; Charles R Hatt; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-22
View more
  2 in total

1.  Online mobile C-arm calibration using inertial sensors: a preliminary study in order to achieve CBCT.

Authors:  Imane Lemammer; Olivier Michel; Hacheme Ayasso; Steeve Zozor; Guillaume Bernard
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-09-10       Impact factor: 2.924

2.  Feature-based CBCT self-calibration for arbitrary trajectories.

Authors:  Christian Tönnes; Tom Russ; Lothar R Schad; Frank G Zöllner
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-05-20       Impact factor: 3.421

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.