Online mobile C-arm calibration using inertial sensors: a preliminary study in order to achieve CBCT.

Imane Lemammer1,2, Olivier Michel3, Hacheme Ayasso3, Steeve Zozor3, Guillaume Bernard4.   

Abstract

PURPOSE: Cone beam computed tomography (CBCT) became increasingly popular over the last years. It allows more accurate diagnosis and treatment planning with a lower effective radiation dose. However, volume reconstruction algorithms require a very precise knowledge of the imaging geometry. Due to mechanical instabilities, mobile C-arms are incompatible with existing tomography algorithms. Therefore, C-arm online calibration is essential in order to achieve an accurate volume reconstruction.
METHODS: We present an online calibration method for mobile C-arms. It is based on tracking the detector and the X-ray source of the C-arm using three-axis gyroscopes and accelerometers. It aims to be precise and noninvasive. The performance of the calibration algorithm is evaluated in regard to the precision of the sensors and to whether or not dynamic models are considered. In addition, we present an algorithm which propagate the errors from the positions and orientations estimates to the 2D projections on the detector plane. Thus, we can evaluate the impact of the estimation errors on the acquired images.
RESULTS: The experiments are conducted on an experimental C-arm. The reached accuracy is [Formula: see text] for orientation and [Formula: see text] for position. These errors propagate as an error of [Formula: see text] for the 2D projections on the detector plane.
CONCLUSIONS: The proposed calibration algorithm achieves an accuracy comparable to the precision of existing calibration methods. The required angle accuracy by CBCT algorithms is reached. However, improvements are needed to achieve the required position precision. The in-plane translations of the X-ray source and the detector are the most crucial parameters to estimate in order to conduct CBCT on mobile C-arms.

Keywords:  CBCT; Error propagation; IMU; Mobile C-arm; Online calibration; Pose estimation

Mesh:

Year:  2019        PMID: 31506881     DOI: 10.1007/s11548-019-02061-6

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  11 in total

1.  C-arm rotation encoding with accelerometers.

Authors:  Victor Grzeda; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-04-10       Impact factor: 2.924

2.  On the question of 3D seed reconstruction in prostate brachytherapy: the determination of x-ray source and film locations.

Authors:  Mutian Zhang; Marco Zaider; Michael Worman; Gilad Cohen
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

3.  Automatic detection of fiducial markers in fluoroscopy images for on-line calibration.

Authors:  Laurence Smith; Mike Pleasance; Rosalyn Seeton; Neculai Archip; Robert Rohling
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

4.  C-arm angle measurement with accelerometer for brachytherapy: an accuracy study.

Authors:  Thomas Wolff; Andras Lasso; Markus Eblenkamp; Erich Wintermantel; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-07-03       Impact factor: 2.924

Review 5.  C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology.

Authors:  Robert C Orth; Michael J Wallace; Michael D Kuo
Journal:  J Vasc Interv Radiol       Date:  2008-04-23       Impact factor: 3.464

6.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT.

Authors:  M J Daly; J H Siewerdsen; Y B Cho; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

7.  Environment effects at phantom-based X-ray pose measurements.

Authors:  Sabine Thurauf; Mario Korner; Florian Vogt; Oliver Hornung; M Ali Nasseri; Alois Knoll
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2017-07

8.  Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems.

Authors:  Khanlian Chung; Lothar R Schad; Frank G Zöllner
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-05-08       Impact factor: 2.924

9.  Intraoperative cone-beam CT for correction of periaxial malrotation of the femoral shaft: a surface-matching approach.

Authors:  Amal Khoury; Cari M Whyne; Michael Daly; Douglas Moseley; Greg Bootsma; Tomas Skrinskas; Jeffrey Siewerdsen; David Jaffray
Journal:  Med Phys       Date:  2007-04       Impact factor: 4.071

10.  On-line smoothing for an integrated navigation system with low-cost MEMS inertial sensors.

Authors:  Kai-Wei Chiang; Thanh Trung Duong; Jhen-Kai Liao; Ying-Chih Lai; Chin-Chia Chang; Jia-Ming Cai; Shih-Ching Huang
Journal:  Sensors (Basel)       Date:  2012-12-13       Impact factor: 3.576

View more
  3 in total

1.  Rigid and Non-Rigid Motion Compensation in Weight-Bearing CBCT of the Knee Using Simulated Inertial Measurements.

Authors:  Jennifer Maier; Marlies Nitschke; Jang-Hwan Choi; Garry Gold; Rebecca Fahrig; Bjoern M Eskofier; Andreas Maier
Journal:  IEEE Trans Biomed Eng       Date:  2022-04-21       Impact factor: 4.756

2.  3D-XGuide: open-source X-ray navigation guidance system.

Authors:  Ina Vernikouskaya; Dagmar Bertsche; Wolfgang Rottbauer; Volker Rasche
Journal:  Int J Comput Assist Radiol Surg       Date:  2020-10-15       Impact factor: 2.924

3.  Evaluation and Verification of Fast Computational Simulations of Stent-Graft Deployment in Endovascular Aneurysmal Repair.

Authors:  Aymeric Pionteck; Baptiste Pierrat; Sébastien Gorges; Jean-Noël Albertini; Stéphane Avril
Journal:  Front Med Technol       Date:  2021-07-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.