| Literature DB >> 29666706 |
Lucie Sukupova1, Ondrej Hlavacek2, Daniel Vedlich2.
Abstract
The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters. Measurements were performed in different projections and with different positions of the ceiling-mounted shielding: without shielding; shielding closest to the patient; and shielding closest to the physician performing the procedure. The protection provided by the shielding was assessed for cardiology when the femoral access is used and for radiology when the physician performs the procedure in the abdominal area. The results show that the use of the ceiling-mounted shielding can decrease the dose from the scatter radiation by 95% at the position of the performing physician. In cardiology, the impact is more pronounced when the left oblique projection is used. In radiology, a large decrease was observed for right oblique projections, compared to cardiology. The ceiling-mounted shielding should be placed as close to the physician as possible. The idea of creating the largest radiation shadow by placing the radiation shielding as close to the patient as possible does not provide as effective radiation protection of the operator as it might be thought.Entities:
Year: 2018 PMID: 29666706 PMCID: PMC5831950 DOI: 10.1155/2018/4287973
Source DB: PubMed Journal: Radiol Res Pract ISSN: 2090-195X
Figure 1The geometry of measurement on the angiography system for cardiology.
Figure 2The geometry of measurement on the angiography system for radiology.
Figure 3Different positions of the ceiling-mounted shielding for cardiology.
Figure 4Different positions of the ceiling-mounted shielding for radiology.
Doses (μGy) from scatter radiation per 10 frames in cardiology for different projections (in each cell, dose without the use of shielding, with shielding placed close to the patient, and with shielding placed close to the physician. Digits in bold do not differ significantly at 0.05.).
| RAO | 0° | LAO | |||||
|---|---|---|---|---|---|---|---|
| 90° | 60° | 30° | 30° | 60° | 90° | ||
| CD | |||||||
| 30° | 0.77/0.17/0.08 | 0.69/0.12/0.06 | 2.76/0.28/0.12 | ||||
| 15° | 0.18/ | 0.27/0.10/0.04 | 0.54/0.21/0.03 | 1.73/0.73/0.06 | 2.36/ | ||
| 0° | 0.30/ | 0.19/ | 0.19/0.08/0.03 | 0.59/0.44/0.02 | 1.20/0.74/0.05 | 2.27/0.09/0.04 | 2.40/ |
| CR | |||||||
| 15° | 0.22/ | 0.45/0.28/0.06 | 0.76/0.47/0.09 | 2.59/0.46/0.07 | |||
| 30° | 0.74/0.38/0.13 | 0.83/0.60/0.11 | 4.51/0.97/0.29 | ||||
Figure 5Doses from the scatter radiation in cardiology for RAO 90° to LAO 90°, CD/CR 0°.
Figure 6Doses from the scatter radiation in cardiology for LAO/RAO 0°, CD 30° to CR 30°.
Doses (μGy) from scatter radiation per 10 fr in radiology for different projections (in each cell, dose without the use of shielding, with shielding placed close to the patient, and with shielding placed close to the physician. Hyphens mean that the values could not be measured because the dose rate was too high. When there are only two values in the cell, the measurements with shielding placed close to the patient and with shielding placed close to the physician were the same. All the values differ significantly at 0.05.).
| RAO | 0° | LAO | |||
|---|---|---|---|---|---|
| 90° | 30° | 30° | 90° | ||
| CD | |||||
| 20° | 28.38/1.43 | - | |||
| 0° | 10.72/2.67 | 34.68/1.55 | 48.35/4.64/0.89 | 137.31/15.97/2.32 | - |
| CR | |||||
| 20° | 44.35/2.92/2.01 | 119.69/6.55/4.22 | |||
Dose from scatter radiation normalized to PKA (μGy/μGy ∗ m2) for cardiology (in each cell dose without the use of shielding, with shielding placed close to the patient and with shielding placed close to the physician).
| RAO | 0° | LAO | |||||
|---|---|---|---|---|---|---|---|
| 90° | 60° | 30° | 30° | 60° | 90° | ||
| CD | |||||||
| 30° | 0.010/0.002/0.001 | 0.016/0.003/0.001 | 0.037/0.004/0.002 | ||||
| 15° | 0.005/0.002/0.001 | 0.007/0.003/0.001 | 0.019/0.007/0.001 | 0.040/0.017/0.001 | 0.081/0.003/0.003 | ||
| 0° | 0.011/0.002/0.001 | 0.006/0.001/0.001 | 0.007/0.003/0.001 | 0.022/0.016/0.001 | 0.037/0.023/0.001 | 0.086/0.003/0.001 | 0.086/0.003/0.002 |
| CR | |||||||
| 15° | 0.007/0.002/0.001 | 0.009/0.006/0.001 | 0.026/0.016/0.003 | 0.052/0.009/0.001 | |||
| 30° | 0.007/0.003/0.001 | 0.016/0.012/0.002 | 0.043/0.009/0.003 | ||||
Dose from scatter radiation normalized to PKA (μGy/μGy ∗ m2) for radiology (in each cell dose without the use of shielding, with shielding placed close to the patient, and with shielding placed close to the physician. When there are only two values in the cell, the measurements with shielding placed close to the patient and with shielding placed close to the physician were the same).
| RAO | 0° | LAO | |||
|---|---|---|---|---|---|
| 90° | 30° | 30° | 90° | ||
| CD | |||||
| 20° | 0.021/0.001 | - | |||
| 0° | 0.008/0.002 | 0.026/0.001 | 0.044/0.004/0.001 | 0.107/0.012/0.002 | - |
| CR | |||||
| 20° | 0.034/0.002/0.002 | 0.090/0.005/0.003 | |||