Relationship of beam angulation and radiation exposure in the cardiac catheterization laboratory.

Shikhar Agarwal1, Akhil Parashar1, Navkaranbir Singh Bajaj2, Imran Khan3, Imran Ahmad1, Fredrick A Heupler1, Matthew Bunte1, Dhruv K Modi1, E Murat Tuzcu1, Samir R Kapadia4.   

Abstract

OBJECTIVES: The aim of this study was to analyze the relationship between beam angulation and air kerma in a modern cardiac catheterization laboratory.
BACKGROUND: Recent reports have identified the merits of reducing radiation scatter, an important determinant of radiation dose in the catheterization laboratory. Radiation scatter is poorly characterized in the context of catheterization laboratories using modern digital equipment. Understanding the principles of dosimetry may reduce the radiation exposure to patients, providers, and medical staff.
METHODS: Prospectively captured radiation data were extracted from a database of 1,975 diagnostic catheterizations (DCs) and 755 percutaneous coronary interventions (PCIs), which included 138,342 fluoroscopic and 35,440 acquisition (cine) sequences. Fluoroscopy and acquisition modes were categorized into tertiles based on the total air kerma measured at a standard reference point. Radiation maps were modeled according to the relative proportion of exposure in each projection.
RESULTS: Median air kerma during DCs and PCIs was 677 and 2,188 mGy, respectively. Fluoroscopy contributed to 66.3% of total dose during PCIs compared with 39.7% during DCs (p < 0.001). Fluoroscopy was more sensitive to changes in angulation with a rapid increase in total air kerma on small increases in beam angulation. Complex spatial maps were created to study the impact of angulation and other covariates on total air kerma. Besides beam angulation, body surface area was the strongest predictor of the total air kerma.
CONCLUSIONS: This study uniquely describes radiation dosimetry using contemporary equipment in a real-world setting. Extreme angulations were associated with high air kerma values. Fluoroscopy compared with acquisition was more sensitive to changes in angulation, with relatively larger increases in total air kerma with small increases in steepness of the angulation.
Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  air kerma; beam angulation; kerma rate; projections; radiation

Mesh:

Year:  2014        PMID: 24746652     DOI: 10.1016/j.jcin.2013.12.203

Source DB:  PubMed          Journal:  JACC Cardiovasc Interv        ISSN: 1936-8798            Impact factor:   11.195


  16 in total

1.  Physicians' exposure to radiation during electrophysiology procedures.

Authors:  Laurent Faroux; Charline Daval; François Lesaffre; Thierry Blanpain; Jean-Pierre Chabert; Angeline Martin; Mathias Guinot; Nicolas Luconi; Madeline Espinosa; Pierre Nazeyrollas; Christophe Tourneux; Damien Metz
Journal:  J Interv Card Electrophysiol       Date:  2019-06-08       Impact factor: 1.900

2.  The lateral plane delivers higher dose than the frontal plane in biplane cardiac catheterization systems.

Authors:  Osamah Aldoss; Sonali Patel; Kyle Harris; Abhay Divekar
Journal:  Pediatr Cardiol       Date:  2015-01-14       Impact factor: 1.655

3.  Predictors of Excess Patient Radiation Exposure During Chronic Total Occlusion Coronary Intervention: Insights From a Contemporary Multicentre Registry.

Authors:  Georgios E Christakopoulos; Georgios Christopoulos; Dimitri Karmpaliotis; Khaldoon Alaswad; Robert W Yeh; Farouc A Jaffer; Michael R Wyman; William L Lombardi; Muhammad Nauman J Tarar; J Aaron Grantham; David E Kandzari; Nicholas Lembo; Jeffrey W Moses; Ajay J Kirtane; Manish Parikh; Philip Green; Matthew Finn; Santiago Garcia; Anthony H Doing; Raja Hatem; Craig A Thompson; Subhash Banerjee; Emmanouil S Brilakis
Journal:  Can J Cardiol       Date:  2016-11-11       Impact factor: 5.223

4.  Reducing radiation exposure during invasive coronary angiography and percutaneous coronary interventions implementing a simple four-step protocol.

Authors:  Moritz Seiffert; Francisco Ojeda; Kai Müllerleile; Elvin Zengin; Christoph Sinning; Christoph Waldeyer; Edith Lubos; Ulrich Schäfer; Karsten Sydow; Stefan Blankenberg; Dirk Westermann
Journal:  Clin Res Cardiol       Date:  2015-01-22       Impact factor: 5.460

5.  Reduction of Fluoroscopy Time and Radiation Dosage During Catheter Ablation for Atrial Fibrillation.

Authors:  Kenichiro Yamagata; Bashar Aldhoon; Josef Kautzner
Journal:  Arrhythm Electrophysiol Rev       Date:  2016-08

6.  The Effectiveness and Safety between Monoplane and Biplane Imaging During Coronary Angiographies.

Authors:  Chon-Seng Hong; Zhih-Cherng Chen; Kuo-Ting Tang; Wei-Ting Chang
Journal:  Acta Cardiol Sin       Date:  2020-03       Impact factor: 2.672

7.  Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures.

Authors:  Kelly S Wilson-Stewart; Davide Fontanarosa; Eva Malacova; Jamie V Trapp
Journal:  Phys Eng Sci Med       Date:  2022-05-09

8.  Operator Radiation and the Efficacy of Ceiling-Suspended Lead Screen Shielding during Coronary Angiography: An Anthropomorphic Phantom Study Using Real-Time Dosimeters.

Authors:  Qianjun Jia; Ziman Chen; Xianxian Jiang; Zhenjun Zhao; Meiping Huang; Jiahua Li; Jian Zhuang; Xiaoqing Liu; Tianyu Hu; Wensheng Liang
Journal:  Sci Rep       Date:  2017-02-07       Impact factor: 4.379

9.  Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography.

Authors:  Ali Tarighatnia; Amirhossien Mohammadalian; Morteza Ghojazade; Leili Pourafkari; Alireza Farajollahi
Journal:  Anatol J Cardiol       Date:  2017-08-02       Impact factor: 1.596

Review 10.  Reduction of Radiation Risk to Interventional Cardiologists and Patients during Angiography and Coronary Angioplasty.

Authors:  Mohsen Mohammadi; Leili Danaee; Effat Alizadeh
Journal:  J Tehran Heart Cent       Date:  2017-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.