Reduction of operator radiation dose by an extended lower body shield.

John P Gonzales1, Christopher Moran2, James E Silberzweig2.   

Abstract

PURPOSE: To quantify the reduction in operator exposure to scatter radiation by using an extension component in addition to a commonly used lower body radiation shield attached to an interventional radiology procedure table.
MATERIALS AND METHODS: An anthropomorphic pelvis phantom was exposed to fluoroscopy at varying C-arm angles to simulate a standard interventional procedure. A MAVIG UT60 lower body shield (MAVIG, Munich, Germany) (48 cm × 78 cm, 0.5 mm lead equivalent), with an attachable extension component (48 cm × 36 cm), was suspended from the edge of the table adjacent to the pelvic phantom. Using a handheld Geiger counter, scatter radiation exposure rates were measured at the level of an operator's eye, chest, waist, and knee, with various C-arm angles both with and without the attachable extension component. Mean exposure rates for each experimental setup were calculated and compared.
RESULTS: Overall, scatter radiation exposures were lower with the addition of the extension component, with the largest reductions (> 80%) measured at the operator's waist and knee levels, for all C-arm angles. The highest reduction in scatter radiation exposure was measured at knee level, at 0 degrees left posterior oblique projection, where the use of the lower body shield extension component reduced the exposure rate from 4.80 mR/h to 0.44 mR/h (90.8% reduction, P < .001). Reductions in scatter radiation were less at eye and chest levels.
CONCLUSIONS: The use of the additional extension component to the lower body radiation shield can result in large (> 80%) reductions in operator scatter radiation exposure, particularly to the lower body.
© 2014 SIR Published by SIR All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 24332242     DOI: 10.1016/j.jvir.2013.11.006

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures.

Authors:  Lucie Sukupova; Ondrej Hlavacek; Daniel Vedlich
Journal:  Radiol Res Pract       Date:  2018-01-30

2.  Reduction of radiation exposure to operating physician and assistant using a real-time auditory feedback dosimeter during femoral artery puncturing: a study on swine model.

Authors:  Muhammad Umair Ahmad Khan; Byung-Ju Yi
Journal:  Eur Radiol Exp       Date:  2019-09-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.