Literature DB >> 29657646

Iterative Reconstruction as a Method for Optimisation of Computed Tomography Procedures.

Maria Staniszewska1, Dariusz Chrusciak2.   

Abstract

BACKGROUND: Computed tomography (CT) is still commonly regarded as a method that causes a high radiation exposure. For that reason, producers intensively try to find new solutions for dose reduction while maintaining a high diagnostic value of images. One of the recent strategies focuses on CT image reconstruction. Iterative reconstruction (IR) is an alternative for filtered back projection (FBP) that is commonly used today.The aim of the article is to demonstrate and compare the effects of two IR algorithms on dose value and image details. MATERIAL/
METHODS: Investigations were performed on two 128 multi-detector (MDCT) CT scanners: - iCT (Philips Healthcare with iDose4); - Definitions AS+ (Siemens Medical Solutions with SAFIRE system).The measurements involved: - image quality indicators for the CATPHAN 600 phantom; - dosimetric indicators of exposure (DLP i CTDIvol).
RESULTS: The signal-to-noise ratios (SNR) in the images reconstructed with IR and FBP were analysed, and the SNR(IR)/SNR(FBP) ratios were calculated and correlated with CTDIvol values. The effects of IR and FBP algorithms on low-contrast resolution were also compared in relation to CTDIvol values. The smallest diameter of supra-slice objects in the Catphan phantom were taken into consideration.Both iterative algorithms definitely improved the visibility of low-contrast objects in comparison to a standard algorithm (FBP) with similar exposure parameters.These algorithms allow an 80% reduction of the CTDIvol value while maintaining an acceptable visibility of low-contrast objects. However, the results obtained with each of the studied iterative algorithms differ.

Entities:  

Keywords:  Image Processing, Computer-Assisted; Radiation Dosage; Tomography, X-Ray Computed

Year:  2017        PMID: 29657646      PMCID: PMC5894051          DOI: 10.12659/PJR.903557

Source DB:  PubMed          Journal:  Pol J Radiol        ISSN: 1733-134X


Background

Computed tomography (CT) is still commonly regarded as a method that causes a high radiation exposure. For that reason, producers intensively try to find new solutions for dose reduction while maintaining a high diagnostic value of images. Earlier approaches relied on optimization of CT exposure parameters (also special paediatric protocols) and automatic tube-current modulation [1,2]. A newer dose reduction strategy focuses on optimization of CT image reconstruction. To date, most of CT scanners have used filtered back projection (FBP) as the algorithm for image reconstruction. FBP assumes that the acquired projection data are free of noise. Consequently, after mathematical filtering of projection data (e.g. smoothing or edge-enhancement), they are projected back into image space to reconstruct the imaged volume. The FBP algorithm is fast and produces acceptable images in most situations. However, when radiation dose is lowered, FBP- reconstructed images are highly noisy [3]. Iterative reconstruction (IR) is an alternative image reconstruction method that allows imaging at lower doses while maintaining image quality comparable to routine-dose FBP. The concept of IR is well known, and the first IR algorithms were elaborated already in the 1970s to good effect. Due to the limited computational power of early CT workstations, the image creation process was very slow and the subsequently elaborated algorithm was quicker; it was the filtered back projection (FBP) algorithm, which has been successfully applied to date, even for quick multi-row CT scanners. A relatively high level of image noise is a disadvantage of the FBP algorithm. This results from the necessity of raw image data interpolation and is the cause of higher radiation doses to which patients are exposed in order to achieve good quality diagnostic images. The improved computational effectiveness of CT reconstruction workstations permitted the introduction of IR. Consequently, all major CT manufacturers have recently implemented new IR algorithms for CT imaging as the best solution for dose reduction that does not compromise image quality. Full iterative reconstruction consists of forward and backward reconstruction and thus operates in the raw data domain and image data domain. Simpler algorithms iterate successively in the raw data domain and then in the image data domain. The result of IR in the raw data domain is transmitted to the image domain by back projection [3]. There are differences in the approach to IR algorithms of different vendors. Each vendor has its own uniquely named IR technique, and the main CT vendors compete to achieve the lowest possible radiation exposure [4,5]. This article presents the effects of two IR algorithms of the same level of computational complexity (installed on the same CT workstations) on the image quality - dose exposure relationship in comparison to FBP algorithms. A special aim of the article is to demonstrate the benefit of iterative algorithms for optimization of exam results in comparison to the FBP algorithm when routinely used exposure parameters are chosen.

Materials and Methods

Investigations were performed with two CT scanners: iCT scanner (Philips Healthcare with iDose4); Definitions AS+ (Siemens Medical Solutions with SAFIRE system). Both IR solutions operate successively in the raw data domain and then in the image domain. Both IR systems are multilevel: iDose4 has 7 levels and SAFIRE has 5 levels, corresponding to the number of iterative loop repetitions. The workstations of the both scanners can independently use the FBP algorithm. A starting point for the investigations was to set the exposure parameters to those commonly chosen for CT examinations of the trunk in adult patients: high voltage U=120 kV, anode current I=500 mA, rotation time trot=0.5 s, pitch p≈1. Both scanners have 128 active detector rows of the unit width ∆z=0.625 mm for iCT and ∆z=0.6 mm for Definition AS+, respectively. The measurements involved: image quality indicators for the CATPHAN 600 phantom, dosimetric indicators of exposure (DLP i CTDIvol). For the dosimetric measurements, the Barracuda device (RTI, Sweden) and the HEAD 5-holes phantom (Ǿ=16 cm) were applied. The following indicators of image quality were evaluated: low-contrast resolution (in the CATPHAN phantom the so-called supra-slice and sub-slice objects), high-contrast resolution (i.e. space resolution), signal-to-noise ratio (SNR). SNR was evaluated for a region-of-interest (ROI) of 100 mm2 placed in the area surrounding the objects of a nominal contrast level (CL) of 1%. In this ROI, the Hounsfield numbers and their standard deviations were interpreted as the signal level and the level of noise, respectively. Independently, computed tomography dose index (CTDI) was calculated as the dosimetric indicator of CT exposure. For multi-row CT scanners, CTDI is calculated according to the following formula where N – number of active rows, Δz – width of the single detector, p – pitch, the mean weighted value of dose length product dose length product in the central hole of the phantom, dose length product in the peripheral holes of the phantom. CTDIvol should be understood as the absorbed dose averaged for the whole scanned volume. Thus, this value estimates radiation risk in patients undergoing an exam with given exposure parameters. The measurements were performed for the three following exposure settings: 120 kV and 250 mAs, 120 kV and 150 mAs and 80 kV and 150 mAs. CATPHAN images were reconstructed using successively the following algorithms: for iCT: FBP, iDose4 – level 2, iDose4 – level 4 and iDose4 – level 6, for Definition AS+: FBP, SAFIRE – level 2, SAFIRE – level 3, SAFIRE – level 5. CATPHAN images were recorded with an intentionally thin reconstructive layer of 3 mm, corresponding to the smallest low-contrast sub-slice objects (i.e. 3 mm height) chosen for the analysis.

Results

The results corresponding to the particular settings of exposure parameters are given: in Tables 1A, 2A, and 3A for iCT and
Table 1A

iCT (Philips) Determinants of image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=249 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP7.73 mm5 mm0.06336.0
IDose-29.03 mm5 mm
IDose-411.03 mm5 mm
IDose-613.82 mm3 mm

Diameter (in mm) of the smallest differentiable object is given

value after reconstruction using mathematical „bone” filter.

Table 2A

iCT (Philips) Determinants od image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=149 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP8.34 mm7 mm0.06321.5
IDose-29.83 mm7 mm
IDose-411.63 mm7 mm
IDose-614.73 mm5 mm

Diameter (in mm) of the smallest differentiable object is given;

value after reconstruction using mathematical „bone” filter.

Table 3A

iCT (Philips) Determinants od image quality and CT dose index for exposure parameters U=80 kV, Ixtrot=149 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP2.24 mm7 mm0.0637.3
IDose-22.73 mm7 mm
IDose-43.13 mm7 mm
IDose-63.93 mm5 mm

Diameter (in mm) of the smallest differentiable object is given

value after reconstruction using mathematical „bone” filter.

in Tables 1B, 2B, and 3B for Definition AS+.
Table 1B

Definition AS+ (Siemens) Determinants of image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=250 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP14.73 mm7 mm0.07139.3
SAFIRE-213.83 mm7 mm
SAFIRE-317.83 mm7 mm
SAFIRE-418.83 mm7 mm

Diameter (in mm) of the smallest differentiable object is given

value after reconstruction using mathematical „bone” filter.

Table 2B

Definition AS+ (Siemens) Determinants od image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=150 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP10.44 mm7 mm0.07123.7
SAFIRE-212.84 mm7 mm
SAFIRE-313.63 mm7 mm
SAFIRE-416.63 mm7 mm

Diameter (in mm) of the smallest differentiable object is given

value after reconstruction using mathematical „bone” filter.

Table 3B

Definition AS+ (Siemens) Determinants od image quality and CT dose index for exposure parameters U=80 kV, Ixtrot=150 mAs.

Algorithm of image reconstructionSNRLow-contrast resolution (*)Spatial resolution (**) (the smallest detectable distance) [cm]CTDIvol [mGy]
Supra-slice objects at 1% contrast levelSub-slice objects 3 mm height
FBP2.812 mmno0.0716.8
SAFIRE-23.89 mmno
SAFIRE-34.06 mmno
SAFIRE-44.56 mmno

Diameter (in mm) of the smallest differentiable object is given

value after reconstruction using mathematical „bone” filter.

iCT (Philips) Determinants of image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=249 mAs. Diameter (in mm) of the smallest differentiable object is given value after reconstruction using mathematical „bone” filter. iCT (Philips) Determinants od image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=149 mAs. Diameter (in mm) of the smallest differentiable object is given; value after reconstruction using mathematical „bone” filter. iCT (Philips) Determinants od image quality and CT dose index for exposure parameters U=80 kV, Ixtrot=149 mAs. Diameter (in mm) of the smallest differentiable object is given value after reconstruction using mathematical „bone” filter. Definition AS+ (Siemens) Determinants of image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=250 mAs. Diameter (in mm) of the smallest differentiable object is given value after reconstruction using mathematical „bone” filter. Definition AS+ (Siemens) Determinants od image quality and CT dose index for exposure parameters U=120 kV, Ixtrot=150 mAs. Diameter (in mm) of the smallest differentiable object is given value after reconstruction using mathematical „bone” filter. Definition AS+ (Siemens) Determinants od image quality and CT dose index for exposure parameters U=80 kV, Ixtrot=150 mAs. Diameter (in mm) of the smallest differentiable object is given value after reconstruction using mathematical „bone” filter. As the main aim of the stud was to evaluate the effects of IR algorithms in relation to the FBP algorithm, the SNRs in the images reconstructed under both algorithms were analysed; SNR, as defined above, was used as a measure of image quality. Thus, for particular levels of the IR algorithms, the ratios SNR(IR)/SNR(FBP) were calculated and correlated with CTDIvol values. The results are presented in Figures 1 and 2 for iCT and Definition AS+, respectively.
Figure 1

SNR(IR)/SNR(FBP) for iCT (Philips). ID-2 – IR algorithm at level 2 (Pearson’s coefficient for linear approximation R2=0.76); ID-4 – IR algorithm at level 4 (Pearson’s coefficient for linear approximation R2=0.43); ID-6 – IR algorithm at level 6 (Pearson’s coefficient for linear approximation R2 0.76).

Figure 2

SNR(IR)/SNR(FBP) for Definition AS+ (Siemens). SAF-2 – IR algorithm at level 2 (Pearson’s coefficient for linear approximation R2=0.99); SAF-3 – IR algorithm at level 3 (Pearson’s coefficient for linear approximation R2≈1); SAF-4 – IR algorithm at level 4 (Pearson’s coefficient for linear approximation R2=0.75).

SNR(IR)/SNR(FBP) for iCT (Philips). ID-2 – IR algorithm at level 2 (Pearson’s coefficient for linear approximation R2=0.76); ID-4 – IR algorithm at level 4 (Pearson’s coefficient for linear approximation R2=0.43); ID-6 – IR algorithm at level 6 (Pearson’s coefficient for linear approximation R2 0.76). SNR(IR)/SNR(FBP) for Definition AS+ (Siemens). SAF-2 – IR algorithm at level 2 (Pearson’s coefficient for linear approximation R2=0.99); SAF-3 – IR algorithm at level 3 (Pearson’s coefficient for linear approximation R2≈1); SAF-4 – IR algorithm at level 4 (Pearson’s coefficient for linear approximation R2=0.75). As an optimal X-ray imaging procedure should provide a clinically useful image at a reasonably low dose to the patient, such an evaluation was performed for the results obtain in our study. The effects of the IR and FBP algorithms for a low-contrast resolution were compared in relation to CTDIvol values. The smallest diameters of supra-slice objects in the Catphan phantom were taken into consideration. The appropriate data are given in Table 4. Correlation between reduction of CTDIvol and relative difference in low contrast (LC) is presented in Figure 3.
Table 4

Visualization of the smallest supra-slice low-contrast objects at the FBP algorithm and the highest level of IR algorithm.

iCTDefinition AS+
CTDIvol [mGy]Ømin [mm]Relative differenceCTDIvol [mGy]Ømin [mm]Relative difference
36.0FBP --> 3IDose-6 --> 230%39.3FBP --> 3SAFIRE-4 --> 30%
21.5FBP --> 4IDose-6 --> 325%23.7FBP --> 4SAFIRE-4--> 325%
7.3FBP --> 9IDose-6 --> 630%6.8FBP --> 12SAFIRE-4 --> 650%
Figure 3

Correlation between reduction of CTDIvol and relative difference in low contrast (LC): % difference=Ømin(FBP) – Ømin (IR)/ØminFBP [%].Reduction of CTDI=MaxCTDI– CTDI(i)/Max CTDI [%].

Visualization of the smallest supra-slice low-contrast objects at the FBP algorithm and the highest level of IR algorithm. Correlation between reduction of CTDIvol and relative difference in low contrast (LC): % difference=Ømin(FBP) – Ømin (IR)/ØminFBP [%].Reduction of CTDI=MaxCTDI– CTDI(i)/Max CTDI [%]. Ømin(FBP), Ømin (IR): the smallest diameters of supra-slice objects in the Catphan phantom visible at the FBP algorithm and IR algorithms (at the highest analysed level), respectively; iCT, Def.AS+ – types of the CT scanners.

Discussion

Iterative reconstruction algorithms are offered now by four major CT vendors. These are different software solutions that are copyright protected. While a multitude of reports promise that IR algorithms to enhance diagnostic performance and reduce radiation exposure, examples of the latter are limited in daily clinical practice. It should be underlined that these benefits are evaluated through the diagnostic value of clinical images. Attention is given to image quality and especially to noise reduction. The latter is estimated as 13% to 50% in neurological or cardiac CT exams [6-8]. It should be also noted that spatial resolution of CT images is high and should not change due to the application of reconstructive algorithms. Thus, increased spatial resolution does not prove that IR algorithms perform better than FBP [8,9]. In the study by Yu at al. [10], image quality of abdominal CT with iDose4 and FBP was compared subjectively and objectively, and about a 20% reduction of image noise was found. This comparison did not include a dose evaluation. Dose reduction can also be evaluated on the basis of clinical scans, and then it ranges between 7% and 80% depending on the CT scanner used [9,11-13]. Moreover, in principle, clinical image evaluation is dependent on diagnostic requirements and is not absolutely objective. Consequently, a fully objective evaluation of the effects of IR algorithms should be performed using standard test objects for both image control and dose level. Such standard objects are the Catphan phantom for image control and CT dosimetric phantoms. Standard phantoms were used for measurements of CTDIvol and analysis of quality of image in this study. Thus, anatomical variation seen in patients and subjective image evaluation do not confound our results. Moreover, our investigations concern the effects of two IR algorithms that are mathematically similar and which are used on technically comparable CT scanners with the same technical settings. Despite such standardization of experimental conditions, some differences between the two IR algorithms were found. The first difference concerns the values of low- contrast resolution as well the sub-slice objects of 3-mm height as the supra-slice objects (see Tables 1A–3B). Based on Figures 1–3, the effectiveness of SAFIRE (for a given iteration level) increases with a reduction of CTDIvol, whereas the effectiveness of iDose4 is nearly constant (for a given iteration level). This is evident on Figure 3 – the Pearson’s coefficient for linear approximation of the relationship between reduction of CTDIPvol and relative difference in low contrast is R2=0 for iCT and R2=1 for Definition AS+. Moreover, the maximal value of SNR(IR)/SNR(FBP) ratio for iDose4 at level 6 is ≈1.8, whereas for SAFIRE is lower at level 4 (≈1.6). This is true for all values of CTDIvol. For the highest CTDIvol used here (i.e. 36 mGy for iCT and ≈39 mGy for Definition AS+), the SNR(IR)/SNR(FBP) at the highest IR level analysed in this study was ≈1.8 for iDose4 (level 6) and ≈1.3 for SAFIRE (level 4). Similar results were obtained on the basis of visualization of the smallest supra-slice low-contrast objects; a comparison between the minimal diameter of the low-contrast object visible at the FBP algorithm and that visible at the highest level of the IR algorithm is presented in Table 4. The iDose4 caused a 30% difference in the diameter, and this was constant for all CTDIvol levels. At the same 80% CTDIvol reduction, a maximal difference of 50% was found for SAFIRE (level 3). A decreased image noise with constant radiation doses was also found by Kuo et.al. [10] for the SAFIRE algorithm in comparison to FBP (both by Siemens) and also on the basis of an analysis of Catphan images. The results indicating a possible dose reduction are also presented in the study by Ward [14], where Brilliance iCT with an iterative algorithm was also used by the authors. In that study, clinical images of the upper abdomen of adult patients obtained with relatively low exposure parameters (120 kV, 73 mAs) were analysed and then compared to another series of images obtained with lower anode current (120 kV, 33 mAs). The latter were still considered as diagnostically acceptable. It should be underlined that the lowered anode current was preceded by analysis of Catphan 600 images, although a criterion of acceptance was the level of spatial resolution. Both iterative algorithms definitely improve visibility of low-contrast objects in comparison to a standard algorithm (FBP) with the same exposure parameters. Both iterative algorithms allow obtaning an 80% reduction of CTDIvol values while maintaining an acceptable visibility of low-contrast objects. At a 40% reduction of the CTDIvol value, the quality of images remains acceptable (visibility of low-contrast objects and SNR value). The reconstruction algorithm does not have an influence on the spatial resolution value. Despite the same mathematical rule of action for both compared iterative algorithms (i.e. iteration successively in the raw data domain and then in the image data domain), the results produced by the algorithms are different; iDose’s4 effects are constant independently of CTDIvol, whereas SAFIRE’s effects become better with increasing CTDIvol. The differences in computer programmes are probably the reason for these findings.
  13 in total

Review 1.  Iterative reconstruction methods in X-ray CT.

Authors:  Marcel Beister; Daniel Kolditz; Willi A Kalender
Journal:  Phys Med       Date:  2012-02-10       Impact factor: 2.685

2.  Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography.

Authors:  Jonathon Leipsic; Troy M Labounty; Brett Heilbron; James K Min; G B John Mancini; Fay Y Lin; Carolyn Taylor; Allison Dunning; James P Earls
Journal:  AJR Am J Roentgenol       Date:  2010-09       Impact factor: 3.959

3.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): Evaluation of image noise reduction in 32 patients.

Authors:  François Pontana; Julien Pagniez; Thomas Flohr; Jean-Baptiste Faivre; Alain Duhamel; Jacques Remy; Martine Remy-Jardin
Journal:  Eur Radiol       Date:  2010-11-05       Impact factor: 5.315

4.  Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT.

Authors:  David A Leswick; Megan M Hunt; Steven T Webster; Derek A Fladeland
Journal:  Radiology       Date:  2008-09-09       Impact factor: 11.105

5.  State of the Art: Iterative CT Reconstruction Techniques.

Authors:  Lucas L Geyer; U Joseph Schoepf; Felix G Meinel; John W Nance; Gorka Bastarrika; Jonathon A Leipsic; Narinder S Paul; Marco Rengo; Andrea Laghi; Carlo N De Cecco
Journal:  Radiology       Date:  2015-08       Impact factor: 11.105

6.  Iterative reconstruction techniques for computed tomography Part 1: technical principles.

Authors:  Martin J Willemink; Pim A de Jong; Tim Leiner; Linda M de Heer; Rutger A J Nievelstein; Ricardo P J Budde; Arnold M R Schilham
Journal:  Eur Radiol       Date:  2013-01-12       Impact factor: 5.315

7.  Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience.

Authors:  Márcio Sommer Bittencourt; Bernhard Schmidt; Martin Seltmann; Gerd Muschiol; Dieter Ropers; Werner Günther Daniel; Stephan Achenbach
Journal:  Int J Cardiovasc Imaging       Date:  2010-12-01       Impact factor: 2.357

8.  CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT.

Authors:  Alban Gervaise; Benoît Osemont; Sophie Lecocq; Alain Noel; Emilien Micard; Jacques Felblinger; Alain Blum
Journal:  Eur Radiol       Date:  2011-09-17       Impact factor: 5.315

9.  Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential.

Authors:  Anna Winklehner; Christoph Karlo; Gilbert Puippe; Bernhard Schmidt; Thomas Flohr; Robert Goetti; Thomas Pfammatter; Thomas Frauenfelder; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2011-08-06       Impact factor: 5.315

10.  Comparison of image quality from filtered back projection, statistical iterative reconstruction, and model-based iterative reconstruction algorithms in abdominal computed tomography.

Authors:  Yu Kuo; Yi-Yang Lin; Rheun-Chuan Lee; Chung-Jung Lin; Yi-You Chiou; Wan-Yuo Guo
Journal:  Medicine (Baltimore)       Date:  2016-08       Impact factor: 1.889

View more
  1 in total

1.  Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study.

Authors:  Yan Xu; Ting-Ting Zhang; Zhi-Hai Hu; Juan Li; Hong-Jun Hou; Zu-Shan Xu; Wen He
Journal:  Diagn Interv Radiol       Date:  2019-11       Impact factor: 2.630

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.