Literature DB >> 29637214

Influence of geometric design characteristics on primary stability of orthodontic miniscrews.

Eman Saad Radwan1, Mona A Montasser2, Ahmed Maher1.   

Abstract

OBJECTIVE: Aim of the present study was to investigate the influence of geometric design characteristics on primary stability of orthodontic miniscrews.
MATERIALS AND METHODS: Forty self-drilling miniscrews with different geometric design characteristics were divided into the following groups (n = 10): group I-Tomas® (Dentaurum, Germany), group II-AbsoAnchor® (Dentos, Korea), group III-HUBIT® miniscrew (HUBIT, Korea), group IV-Creative® (China). The four types were conical miniscrews with 1.6 mm diameter and 6.0 mm length. The miniscrews were manually inserted perpendicular to cow ribs until the full thread length was reached with the help of a 1.3 mm predrilled pilot hole. Each miniscrew was evaluated using scanning electron microscope. Linear and angular measurements were taken using Photoshop CS3 software. Miniscrew stability was measured by the Periotest® and pullout test.
RESULTS: All linear and angular measurements of the geometric characteristics showed significant differences between the four groups (p ≤ 0.001). Results of the pullout test showed significant differences between the four groups (p ≤ 0.001), while the Periotest® values showed no significant differences (p = 0.122). A multiple linear regression analysis revealed the significant predictors for higher pullout: a larger flank, a higher value for the thread angle, lead angle, and apical face angle (p ≤ 0.001).
CONCLUSIONS: Orthodontic miniscrews' geometric design characteristics significantly affected the primary stability. Larger pitch width, flank, thread angle, apical face angle, and/or lead angle led to higher primary stability. Smaller a thread shape factor (TSF) also improved primary stability. Varying these characteristics may enhance miniscrew design.

Entities:  

Keywords:  Geometric design; Miniscrews; Primary stability; Pullout test; Skeletal anchorage

Mesh:

Year:  2018        PMID: 29637214     DOI: 10.1007/s00056-018-0131-7

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  24 in total

1.  Thread shape factor: evaluation of three different orthodontic miniscrews stability.

Authors:  Marco Migliorati; S Benedicenti; A Signori; S Drago; P Cirillo; F Barberis; A Silvestrini Biavati
Journal:  Eur J Orthod       Date:  2012-01-20       Impact factor: 3.075

2.  In vitro evaluation of insertion and removal torques of orthodontic mini-implants.

Authors:  M M Pithon; M G Nojima; L I Nojima
Journal:  Int J Oral Maxillofac Surg       Date:  2010-11-02       Impact factor: 2.789

3.  Insertional torque of the lumbar pedicle screw during surgery.

Authors:  Tomoyuki Ozawa; Kazuhisa Takahashi; Masatsune Yamagata; Seiji Ohtori; Yasuchika Aoki; Tomoko Saito; Gen Inoue; Toshinori Ito; Hideshige Moriya
Journal:  J Orthop Sci       Date:  2005       Impact factor: 1.601

4.  Parameters affecting primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Carsten Rademacher; Gudrun Olthoff; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2006-05       Impact factor: 1.938

5.  Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants.

Authors:  Mitsuru Motoyoshi; Tohru Yoshida; Akiko Ono; Noriyoshi Shimizu
Journal:  Int J Oral Maxillofac Implants       Date:  2007 Sep-Oct       Impact factor: 2.804

6.  Temporary anchorage device stability: an evaluation of thread shape factor.

Authors:  M Migliorati; A Signori; A Silvestrini-Biavati
Journal:  Eur J Orthod       Date:  2011-03-02       Impact factor: 3.075

7.  Mechanical influence of thread pitch on orthodontic mini-implant stability.

Authors:  Amanda Carneiro da Cunha; Amanda Osório Ayres de Freitas; Mariana Marquezan; Lincoln Issamu Nojima
Journal:  Braz Oral Res       Date:  2015-02-16

8.  Miniscrew design and bone characteristics: an experimental study of primary stability.

Authors:  Marco Migliorati; Stefano Benedicenti; Alessio Signori; Sara Drago; Fabrizio Barberis; Henry Tournier; Armando Silvestrini-Biavati
Journal:  Am J Orthod Dentofacial Orthop       Date:  2012-08       Impact factor: 2.650

9.  Three-dimensional tomographic mapping related to primary stability and structural miniscrew characteristics.

Authors:  A Silvestrini Biavati; S Tecco; M Migliorati; F Festa; G Panza; G Marzo; E Gherlone; S Tetè
Journal:  Orthod Craniofac Res       Date:  2011-03-21       Impact factor: 1.826

10.  Application of a wireless resonance frequency transducer to assess primary stability of orthodontic mini-implants: an in vitro study in pig ilia.

Authors:  Yu-Yu Su; Benedict Wilmes; Ralf Hönscheid; Dieter Drescher
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Jul-Aug       Impact factor: 2.804

View more
  6 in total

1.  Analysis of biological and structural factors implicated in the clinical success of orthodontic miniscrews at posterior maxillary interradicular sites.

Authors:  M Palone; A Darsiè; G B Maino; G Siciliani; G A Spedicato; L Lombardo
Journal:  Clin Oral Investig       Date:  2021-11-27       Impact factor: 3.573

2.  Effect of thread depth and thread pitch on the primary stability of miniscrews receiving a torque load : A finite element analysis.

Authors:  Yushan Ye; Weimin Yi; Song Fan; Luodan Zhao; Yansong Yu; Yingjuan Lu; Qinghe Yao; Wei Wang; Shaohai Chang
Journal:  J Orofac Orthop       Date:  2021-09-28       Impact factor: 1.938

3.  Mechanical stability of orthodontic miniscrew depends on a thread shape.

Authors:  Keiichiro Watanabe; Bobby Mitchell; Takuma Sakamaki; Yuzo Hirai; Do-Gyoon Kim; Toru Deguchi; Masakazu Suzuki; Kanji Ueda; Eiji Tanaka
Journal:  J Dent Sci       Date:  2021-12-11       Impact factor: 3.719

4.  Optimization Analysis of Two-Factor Continuous Variable between Thread Depth and Pitch of Microimplant under Toque Force.

Authors:  Yushan Ye; Jiuyang Jiao; Song Fan; Jieying He; Yamei Wang; Qinghe Yao; Wei Wang; Jinsong Li; Shaohai Chang
Journal:  Comput Math Methods Med       Date:  2022-06-20       Impact factor: 2.809

5.  Reliability of Orthodontic Miniscrews: Bending and Maximum Load of Different Ti-6Al-4V Titanium and Stainless Steel Temporary Anchorage Devices (TADs).

Authors:  Andrea Scribante; Mona A Montasser; Eman Saad Radwan; Luisa Bernardinelli; Roberto Alcozer; Paola Gandini; Maria Francesca Sfondrini
Journal:  Materials (Basel)       Date:  2018-07-05       Impact factor: 3.623

6.  Bone quality affects stability of orthodontic miniscrews.

Authors:  Wan-Ping Yu; Ming-Tzu Tsai; Jian-Hong Yu; Heng-Li Huang; Jui-Ting Hsu
Journal:  Sci Rep       Date:  2022-02-18       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.