Thread shape factor: evaluation of three different orthodontic miniscrews stability.

Marco Migliorati1, S Benedicenti, A Signori, S Drago, P Cirillo, F Barberis, A Silvestrini Biavati.   

Abstract

The thread shape factor (TSF) to evaluate the relationships between geometrical characteristics and mechanical properties of the temporary anchorage devices (TADs) has recently been introduced. This in vitro experimental study evaluated in 30 different tests with three TADs: ORTHOImplant (1.8 mm diameter and 10 mm length; 3M Unitek), Tomas (1.6 mm diameter and 10 mm length; Dentaurum), and Orthoeasy (1.7 mm diameter and 10 mm length; Forestadent). Scanning electron microscopy images were acquired for each TAD to measure the TSF; afterwards, the maximum insertion torque (MIT) was evaluated and thereafter pull-out tests on two differently designed organic bone analogs were carried out using a testing machine with a crosshead speed of 2 mm/minute being applied. One-way analysis of variance with group as factor was performed. Post hoc multiple comparisons Bonferroni test was used. Rank-transformed data were used when asymmetry of data was shown. To assess correlation between characteristics, load, and MIT, Spearman's rank correlation coefficient was used. A P-value of 0.05 was considered statistically significant. Significant direct correlations were found between TSF and depth and both load and MIT. Particularly, a correlation of 0.90 (P < 0.001) was found between depth and MIT for 2.2 mm cortical thickness. The authors conclude that MIT and maximum load values of pull-out test are statistically related to depth of the thread of the screw and to TSF.

Mesh:

Year:  2012        PMID: 22267706     DOI: 10.1093/ejo/cjr145

Source DB:  PubMed          Journal:  Eur J Orthod        ISSN: 0141-5387            Impact factor:   3.075


  7 in total

1.  Influence of geometric design characteristics on primary stability of orthodontic miniscrews.

Authors:  Eman Saad Radwan; Mona A Montasser; Ahmed Maher
Journal:  J Orofac Orthop       Date:  2018-04-10       Impact factor: 1.938

2.  Mechanical stability of orthodontic miniscrew depends on a thread shape.

Authors:  Keiichiro Watanabe; Bobby Mitchell; Takuma Sakamaki; Yuzo Hirai; Do-Gyoon Kim; Toru Deguchi; Masakazu Suzuki; Kanji Ueda; Eiji Tanaka
Journal:  J Dent Sci       Date:  2021-12-11       Impact factor: 3.719

3.  Effect of steam and dry heat sterilization on the insertion and fracture torque of orthodontic miniscrews.

Authors:  Shiva Alavi; Farnaz Asadi; Seyed Amir Hossein Raji; Soroor Samie
Journal:  Dent Res J (Isfahan)       Date:  2020-05-23

4.  One palatal implant for skeletal anchorage--frequency and range of indications.

Authors:  Elena Krieger; Zeynep Yildizhan; Heinrich Wehrbein
Journal:  Head Face Med       Date:  2015-04-21       Impact factor: 2.151

5.  The Effects of Different Miniscrew Thread Designs and Force Directions on Stress Distribution by 3-dimensional Finite Element Analysis.

Authors:  Hamidreza Fattahi; Shabnam Ajami; Ali Nabavizadeh Rafsanjani
Journal:  J Dent (Shiraz)       Date:  2015-12

6.  Drilling Capability of Orthodontic Miniscrews: In Vitro Study.

Authors:  Alessandra Marchi; Matteo Camporesi; Maurizio Festa; Luis Salvatierra; Sara Izadi; Giampietro Farronato
Journal:  Dent J (Basel)       Date:  2020-12-21

7.  Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial.

Authors:  Marco Migliorati; Sara Drago; Fabrizio Barberis; Irene Schiavetti; Domenico Dalessandri; Stefano Benedicenti; Armando Silvestrini Biavati
Journal:  Open Dent J       Date:  2016-05-31