The influence of implant-abutment connection on the screw loosening and microleakage.

Katsuhiro Tsuruta1, Yasunori Ayukawa2, Tatsuya Matsuzaki1, Masafumi Kihara1, Kiyoshi Koyano1.   

Abstract

BACKGROUND: There are some spaces between abutment and implant body which can be a reservoir of toxic substance, and they can penetrate into subgingival space from microgap at the implant-abutment interface. This penetration may cause periimplantitis which is known to be one of the most important factors associated with late failure. In the present study, three kinds of abutment connection system, external parallel connection (EP), internal parallel connection (IP), and internal conical connection (CC), were studied from the viewpoint of microleakage from the gap between the implant and the abutment and in connection with the loosening of abutment screw.
METHODS: We observed dye leakage from abutment screw hole to outside through microgap under the excessive compressive and tensile load and evaluated the anti-leakage characteristics of these connection systems.
RESULTS: During the experiment, one abutment screw for EP and two screws for IP, out of seven samples in each group, were fractured. After the 2000 cycles of compressive tensile loadings, removal torque value (RTV) of abutment screw represented no statistical differences among three groups. Standard deviation was largest in the RTV of EP and smallest in that of CC. The results of microleakage of toluidine blue from implant-abutment connection indicated that microleakage generally increased as loading procedure progressed. The amount of microleakage was almost plateau at 2000 cycles in CC, but still increasing in other two groups. The value of microleakage greatly scattered in EP, but the deviation of that in CC is significantly smaller. At 500 cycles of loading, there were no significant differences in the amount of microleakage among the groups, but at 1000, 1500, and 2000 cycles of loading, the amount of microleakage in CC was significantly smaller than that in IP. Throughout the experiment, the amount of microleakage in EP was largest, but no statistical difference was indicated due to the high standard deviation.
CONCLUSIONS: Within the limitation of the present study, CC was stable even after the loading in the RTV of abutment screw and it prevented microleakage from the microgap between the implant body and the abutment, among the three tested connections.

Entities:  

Keywords:  Implant–abutment connection; Microgap; Microleakage; Non-axial load; Screw loosening; Toluidine blue

Year:  2018        PMID: 29629492      PMCID: PMC5890007          DOI: 10.1186/s40729-018-0121-y

Source DB:  PubMed          Journal:  Int J Implant Dent        ISSN: 2198-4034


Background

Although promising outcome of implant therapy has been reported, periimplantitis which is known to be one of the most important factors associated with late failure [1] is a serious complication and expected to overcome to obtain successful outcome. It is known that there are some spaces between abutment and implant body which can be a reservoir of microorganisms and other toxic substance [2], and they can penetrate into subgingival space from microgap located at the implant–abutment interface. This is believed to impact to periimplant inflammation [3]. Thus, many kinds of implant–abutment connection have been proposed to minimize the microgap [4]. Implant–abutment connection can be divided into three types, external parallel connection (EP), internal parallel connection (IP), and conical connection which has the friction between the implant and the abutment (CC). Nobel Biocare has implant systems with these three kinds of implant–abutment connections, and these connection systems are popular even in other manufacturers and they also employ one or some of these connection systems. These systems have a lot of pros and cons and should be selected depending on the dentists’ demand. In the present study, we would like to consider these three kinds of connection system from the point of view of microleakage from the gap between the implant and the abutment and in connection with the loosening of abutment screw. We herewith performed the investigation of dye leakage from abutment screw hole to outside through microgap under the excessive cyclic load applied to cantilever superstructure. Cantilever model could simulate non-axial offset loading applied to implant–abutment complex, and toluidine blue solution was used to measure the extravasation from the gap at the implant–abutment interface.

Methods

Three kinds of Nobel Biocare implants were employed in the present study, namely, Nobel SpeedyGroovy WP 5.0 × 15 mm (EP), Nobel Replace WP 5.0 × 15 mm (IP), and Nobel Parallel CC RP 5.0 × 15 mm (CC) (n = 7 each) (Nobel Biocare, Kloten, Switzerland). Implant–abutment connections are external, internal parallel, and internal and conical connection, respectively. Implant was embedded into steel mold and fixed with epoxy resin (Araldite®, Nichiban, Tokyo, Japan). A superstructure with cantilever (height 15 mm; length 20 mm from the center of access hole) was fabricated using Au–Pt alloy (Degudent LTG, Degudent, Hanau-Wolfgang, Germany) and fixed with abutment screw with fastening torque recommended by the manufacturer (35 Ncm), using torque wrench (Nobel Biocare). Five hundred microliter of water was poured into steel mold, the surface of water located, to small extent, superior to microgap. Then, 50 μl of 0.05% toluidine blue solution was poured at the access hole (Fig. 1).
Fig. 1

A scheme of experimental assembly. a: Abutment and superstructure; b: implant body; c: abutment screw; d: steel mold; e: toluidine blue solution; and f: water

A scheme of experimental assembly. a: Abutment and superstructure; b: implant body; c: abutment screw; d: steel mold; e: toluidine blue solution; and f: water This assembly was set in a universal test machine (Autograph AS-1S, Shimadzu, Kyoto, Japan), and load was applied to the cantilever, at 17.5 mm distance from the center of access hole. As one cycle, one compressive and tensile loads (10 N each) was applied per 1 s and 2000 cycles of loading was done. Before starting load application and every 100 cycles, 100 μl solutions were collected from pool and absorbance at 627 nm was measured using a spectrophotometer (Biospec-mini, Shimadzu). Every 500 cycles, the amount of microleakage was statistically compared using Student t test with Bonferroni correction for multiple comparisons. After the completion of 2000-cycle loading, removal torque value (RTV) of abutment screw was measured.

Results

During the experiment, one (EP) and two (IP) abutment screws out of seven samples were fractured. In this case, abutment screw was changed to new one and experimental procedure was re-run.

RTV of abutment screw after 2000-cycle loading

After the 2000 cycles of compressive tensile loadings, RTV of abutment screw was measured. There were no statistical differences in the RTV among three groups. Standard deviation was largest in the RTV of EP and smallest in that of CC (Fig. 2).
Fig. 2

Removal torque of abutment screw after 2000-cycle loading

Removal torque of abutment screw after 2000-cycle loading

Microleakage of toluidine blue from implant–abutment connection

All groups indicated that microleakage generally increased as loading procedure progressed (Fig. 3) and logarithmic trendline could be drawn, with R2 values of 0.854 in EP, 0.924 in IP, and 0.847 in CC. The amount of microleakage was almost plateau at 2000 cycles in CC group (Fig. 3c), but still increasing in other two groups (Fig. 3a, b). The value of microleakage greatly scattered in EP groups (Fig. 3a), but the deviation of that in CC group is significantly smaller (Fig. 3c).
Fig. 3

a Microleakage from external parallel connection implant–abutment connection. Vertical bars indicate standard deviation. b Microleakage from internal parallel connection implant–abutment connection. Vertical bars indicate standard deviation. c Microleakage from internal conical connection implant–abutment connection. Vertical bars indicate standard deviation

a Microleakage from external parallel connection implant–abutment connection. Vertical bars indicate standard deviation. b Microleakage from internal parallel connection implant–abutment connection. Vertical bars indicate standard deviation. c Microleakage from internal conical connection implant–abutment connection. Vertical bars indicate standard deviation At 500 cycles of loading, there were no significant differences in the amount of microleakage among the groups (Fig. 4a), But at 1000, 1500, and 2000 cycles of loading, the amount of microleakage in CC group was significantly smaller than that in IP group (Fig. 4b–d). There were no statistical differences between EP and other groups in every measurement (Fig. 4).
Fig. 4

a Microleakage of toluidine blue from implant–abutment connection at 500 cycles of loading. No statistical difference is indicated (P > 0.05). b Microleakage of toluidine blue from implant–abutment connection at 1000 cycles of loading. c Microleakage of toluidine blue from implant–abutment connection at 1500 cycles of loading. d Microleakage of toluidine blue from implant–abutment connection at 2000 cycles of loading. EP external parallel connection, IP internal parallel connection, CC internal conical connection. Asterisk indicates statistically significant difference (P < 0.05)

a Microleakage of toluidine blue from implant–abutment connection at 500 cycles of loading. No statistical difference is indicated (P > 0.05). b Microleakage of toluidine blue from implant–abutment connection at 1000 cycles of loading. c Microleakage of toluidine blue from implant–abutment connection at 1500 cycles of loading. d Microleakage of toluidine blue from implant–abutment connection at 2000 cycles of loading. EP external parallel connection, IP internal parallel connection, CC internal conical connection. Asterisk indicates statistically significant difference (P < 0.05)

Discussion

In the present study, cyclic tensile and compressive loading were applied to cantilever superstructure. In the clinical situation, both compressive and tensile force was exerted to implant-supported prosthesis, but no previous study has discussed the microleakage using model study under this situation. In case of conical connection, compressive force may have promoted a higher penetration of the abutments into implant body, which may eliminate microgap [5]. But in the clinical situation, tensile force may also be applied to abutment–implant interface as indicated above; the model employed in the present study may be pertinent. Some previous model studies which measured the extent of microleakage from implant–abutment interface employed microorganisms [6-9]. These studies focused on bacteria itself using visible solution cloudiness test [6], scanning electron microscopy [7], bacteria viability test [8], or checkerboard DNA–DNA hybridization method [9] and did not mentioned about the bacterial toxin. Toluidine blue employed in the present study can easily be measured using absorptiometry and it was reported to be similar to bacterial toxins in its molecular size [5]. In addition, trend in microleakage was reportedly similar between bacterial leakage model and dye leakage model [10]. In the present study, removal toque of abutment screw after the cyclic loading showed no statistically significant difference among the groups. Generally, conical abutment is believed to be better in fit and stability than non-conical connection [11]. One possibility of this discrepancy may be due to the deformation of abutment screw. Actually, implant–abutment connection after the removal of abutment screw was still tight in CC group, but they were easily divided in other two groups. This may indicate that implant–abutment connection in CC group was almost sound and intact after the loading. In contrast, in EP and IP groups, abutment screw may be deformed which lead to the increase of RTV of abutment screw. The fracture of abutment screws in EP and IP groups may support this speculation. In addition, axial force is strongly affected by the interfacial friction coefficient [12]. In the present study, interfacial friction was supposed to be largest in CC group because the contact area between implant body and abutment was smallest in EP group and both EP and IP groups had parallel walls at the interface with gaps and voids [13, 14]. This may be one reason for the larger standard deviation of RTV in both EP and IP groups. In the present study, chronological increase of the amount of microleakage was observed in all three groups. This is in consistent with previous studies [5, 13, 15, 16]. Sigmoid curves of microleakage in all groups meant that the amount of leakage was large at the early stage of loading. This was in agreement with previous reports. Harder et al. reported that bacterial toxin leak occurred within 5 min of incubation using in vitro experimental model study, even without application of loading [15, 16]. The comparison of microleakage among the groups at every 500-cycle load indicated that there were no significant differences among the groups at 500-cycle loading but were statistically significant differences in those between IP and CC groups. The reason for microleakage in EP group having no significant differences indicated between EP and IP or CC groups may be due to the largeness of standard deviation in the value of EP group. The limitation of the present study was the number of samples. We believed the sample number (seven in each group) was sufficient to obtain the trends of microleakage, but it may be better to investigate using large number of samples to analyze the nature of microleakage at the implant–abutment interface in detail. Another limitation was that the fastening torque applied to screw was not necessarily accurate. To mimic clinical situation, we used dedicated beam-type toque wrench delivered from manufacturer. According to the previous studies, the wrench of Nobel Biocare only reportedly demonstrated the target torque value falling within the 95% confidence interval for the true population mean among four kinds of wrenches [17]. In addition, significantly lower deviations of torque values for beam-type wrenches were reported than for coil and toggle-style wrenches [18]. But toque wrench still has an inaccuracy because the scale printed on the wrench is not fine and it seems to be inappropriate to apply precise torque value. In the present study, we used new torque wrench, but in the clinical settings, variability of toque value may expand because torque wrench is repeatedly used, with sterilization procedure, which decrease the accuracy [19].

Conclusions

Within the limitation of the present study, after the offset cyclic tensile and compressive loading, the amount of microleakage from implant–abutment interface was smaller in conical connection than in internal parallel connection.
  19 in total

1.  Comparison of design and torque measurements of various manual wrenches.

Authors:  Jörg Neugebauer; Simone Petermöller; Martin Scheer; Arndt Happe; Franz-Josef Faber; Joachim E Zoeller
Journal:  Int J Oral Maxillofac Implants       Date:  2015 May-Jun       Impact factor: 2.804

Review 2.  Influences of microgap and micromotion of implant-abutment interface on marginal bone loss around implant neck.

Authors:  Yang Liu; Jiawei Wang
Journal:  Arch Oral Biol       Date:  2017-07-29       Impact factor: 2.633

3.  Microleakage at Different Implant-Abutment Connections Under Unloaded and Loaded Conditions.

Authors:  João Paulo da Silva-Neto; Marcel Santana Prudente; Talita Souza Dantas; Plinio Mendes Senna; Ricardo Faria Ribeiro; Flávio Domingues das Neves
Journal:  Implant Dent       Date:  2017-06       Impact factor: 2.454

4.  Implant-Abutment Contact Surfaces and Microgap Measurements of Different Implant Connections Under 3-Dimensional X-Ray Microtomography.

Authors:  Antonio Scarano; Luca Valbonetti; Marco Degidi; Raffaella Pecci; Adriano Piattelli; P S de Oliveira; Vittoria Perrotti
Journal:  Implant Dent       Date:  2016-10       Impact factor: 2.454

5.  Loss of Accuracy of Torque Wrenches Due to Clinical Use and Cleaning Procedure: Short Communication.

Authors:  Egbert Stroosnijder; Marco Mm Gresnigt; Eric W Meisberger; Marco S Cune
Journal:  Int J Prosthodont       Date:  2016 May-Jun       Impact factor: 1.681

6.  Microbial colonization at the implant-abutment interface and its possible influence on periimplantitis: A systematic review and meta-analysis.

Authors:  Marco Tallarico; Luigi Canullo; Martina Caneva; Mutlu Özcan
Journal:  J Prosthodont Res       Date:  2017-03-27       Impact factor: 4.642

Review 7.  Biomechanics of locked plates and screws.

Authors:  Kenneth A Egol; Erik N Kubiak; Eric Fulkerson; Frederick J Kummer; Kenneth J Koval
Journal:  J Orthop Trauma       Date:  2004-09       Impact factor: 2.512

8.  Microbial leakage through the implant-abutment interface of Morse taper implants in vitro.

Authors:  João Pedro Aloise; Ricardo Curcio; Marcia Zorello Laporta; Liliane Rossi; Adriana Madeira Alvares da Silva; Abrão Rapoport
Journal:  Clin Oral Implants Res       Date:  2010-01-13       Impact factor: 5.977

9.  Micro-leakage at the implant-abutment interface with different tightening torques in vitro.

Authors:  João Paulo da Silva-Neto; Marcel Santana Prudente; Thiago de Almeida Prado Naves Carneiro; Mauro Antônio de Arruda Nóbilo; Mario Paulo Amante Penatti; Flávio Domingues das Neves
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

10.  Use of checkerboard DNA-DNA hybridization to evaluate the internal contamination of dental implants and comparison of bacterial leakage with cast or pre-machined abutments.

Authors:  Cássio do Nascimento; Rodrigo Edson Santos Barbosa; João Paulo Mardegan Issa; Evandro Watanabe; Izabel Yoko Ito; Rubens Ferreira de Albuquerque Junior
Journal:  Clin Oral Implants Res       Date:  2009-02-25       Impact factor: 5.977

View more
  13 in total

1.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

2.  In Vitro Study of Preload Loss in Different Implant Abutment Connection Designs.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Marta Relvas; Ana Cristina Braga; Blanca Ríos-Carrasco; Javier Gil; José Vicente Rios-Santos; Ana Fernández-Palacín; Mariano Herrero-Climent
Journal:  Materials (Basel)       Date:  2022-02-14       Impact factor: 3.623

3.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

4.  Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit.

Authors:  Jorge Vélez; Jesús Peláez; Carlos López-Suárez; Rubén Agustín-Panadero; Celia Tobar; María J Suárez
Journal:  J Clin Med       Date:  2020-07-24       Impact factor: 4.241

5.  Clinical Performance of Short Expandable Dental Implants for Oral Rehabilitation in Highly Atrophic Alveolar Bone: 3-year Results of a Prospective Single-Center Cohort Study.

Authors:  Waldemar Reich; Ramona Schweyen; Jeremias Hey; Sven Otto; Alexander Walter Eckert
Journal:  Medicina (Kaunas)       Date:  2020-07-03       Impact factor: 2.430

Review 6.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

7.  The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study.

Authors:  Sunjai Kim; Changhun Choi; Yunsu Cha; Jae-Seung Chang
Journal:  BMC Oral Health       Date:  2021-02-18       Impact factor: 2.757

8.  Influence of Anodizing Stages on the Preload Force of Implant-Abutment Screws and Their Benefits Regarding the Concept of Immediate Implant Placement-An In Vitro Study.

Authors:  Florian Rathe; Paul Weigl; Jan Wasiak; Christoph Ratka; Holger Zipprich
Journal:  Materials (Basel)       Date:  2022-01-20       Impact factor: 3.623

9.  The compressive strength of implant-abutment complex with different connection designs.

Authors:  Hsien-Ching Hung; Chiung-Shing Huang; Yu-Hwa Pan
Journal:  J Dent Sci       Date:  2019-03-28       Impact factor: 2.080

10.  Effects of Aging Torque Controllers on Screw Tightening Force and Bacterial Micro-Leakage on the Implant-Abutment Complex.

Authors:  Yousef Jiries; Tamar Brosh; Shlomo Matalon; Vladimir Perlis; Zeev Ormianer
Journal:  Materials (Basel)       Date:  2022-01-14       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.