Comparison of design and torque measurements of various manual wrenches.

Jörg Neugebauer, Simone Petermöller, Martin Scheer, Arndt Happe, Franz-Josef Faber, Joachim E Zoeller.   

Abstract

PURPOSE: Accurate torque application and determination of the applied torque during surgical and prosthetic treatment is important to reduce complications. A study was performed to determine and compare the accuracy of manual wrenches, which are available in different designs with a large range of preset torques.
MATERIALS AND METHODS: Thirteen different wrench systems with a variety of preset torques ranging from 10 to 75 Ncm were evaluated. Three different designs were available, with a spring-in-coil or toggle design as an active mechanism or a beam as a passive mechanism, to select the preset torque. To provide a clinically relevant analysis, a total of 1,170 torque measurements in the range of 10 to 45 Ncm were made in vitro using an electronic torque measurement device.
RESULTS: The absolute deviations in Ncm and percent deviations across all wrenches were small, with a mean of -0.24 ± 2.15 Ncm and -0.84% ± 11.72% as a shortfall relative to the preset value. The greatest overage was 8.2 Ncm (82.5%), and the greatest shortfall was 8.47 Ncm (46%). However, extreme values were rare, with 95th-percentile values of -1.5% (lower value) and -0.16% (upper value). A comparison with respect to wrench design revealed significantly higher deviations for coil and toggle-style wrenches than for beam wrenches.
CONCLUSION: Beam wrenches were associated with a lower risk of rare extreme values thanks to their passive mechanism of achieving the selected preset torque, which minimizes the risk of harming screw connections.

Entities:  

Mesh:

Year:  2015        PMID: 26009903     DOI: 10.11607/jomi.3733

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

1.  Experimental study on the factors affecting torque of beam-type implant torque wrenches.

Authors:  Hiroki Shiba; Yuji Sato; Junichi Furuya; Tokiko Osawa; Akio Isobe; Myu Hayashi; Noboru Kitagawa
Journal:  BMC Oral Health       Date:  2021-07-15       Impact factor: 2.757

2.  Prospective multicenter non-randomized controlled study on intraosseous stability and healing period for dental implants in the posterior region.

Authors:  Shinya Homma; Yasushi Makabe; Takuya Sakai; Kenzou Morinaga; Satoru Yokoue; Hirofumi Kido; Yasutomo Yajima
Journal:  Int J Implant Dent       Date:  2018-03-29

3.  The influence of implant-abutment connection on the screw loosening and microleakage.

Authors:  Katsuhiro Tsuruta; Yasunori Ayukawa; Tatsuya Matsuzaki; Masafumi Kihara; Kiyoshi Koyano
Journal:  Int J Implant Dent       Date:  2018-04-09

4.  Effects of Aging Torque Controllers on Screw Tightening Force and Bacterial Micro-Leakage on the Implant-Abutment Complex.

Authors:  Yousef Jiries; Tamar Brosh; Shlomo Matalon; Vladimir Perlis; Zeev Ormianer
Journal:  Materials (Basel)       Date:  2022-01-14       Impact factor: 3.623

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.