Literature DB >> 29515001

Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method.

Anand Chandrasekhar1, Chang-Sei Kim2,3, Mohammed Naji1, Keerthana Natarajan1, Jin-Oh Hahn2, Ramakrishna Mukkamala4.   

Abstract

High blood pressure (BP) is a major cardiovascular risk factor that is treatable, yet hypertension awareness and control rates are low. Ubiquitous BP monitoring technology could improve hypertension management, but existing devices require an inflatable cuff and are not compatible with such anytime, anywhere measurement of BP. We extended the oscillometric principle, which is used by most automatic cuff devices, to develop a cuff-less BP monitoring device using a smartphone. As the user presses her/his finger against the smartphone, the external pressure of the underlying artery is steadily increased while the phone measures the applied pressure and resulting variable-amplitude blood volume oscillations. A smartphone application provides visual feedback to guide the amount of pressure applied over time via the finger pressing and computes systolic and diastolic BP from the measurements. We prospectively tested the smartphone-based device for real-time BP monitoring in human subjects to evaluate usability (n = 30) and accuracy against a standard automatic cuff-based device (n = 32). We likewise tested a finger cuff device, which uses the volume-clamp method of BP detection. About 90% of the users learned the finger actuation required by the smartphone-based device after one or two practice trials. The device yielded bias and precision errors of 3.3 and 8.8 mmHg for systolic BP and -5.6 and 7.7 mmHg for diastolic BP over a 40 to 50 mmHg range of BP. These errors were comparable to the finger cuff device. Cuff-less and calibration-free monitoring of systolic and diastolic BP may be feasible via a smartphone.
Copyright © 2018 The Authors, some rights reserved; exclusive licensee American Association for the Advancement of Science. No claim to original U.S. Government Works.

Entities:  

Mesh:

Year:  2018        PMID: 29515001      PMCID: PMC6039119          DOI: 10.1126/scitranslmed.aap8674

Source DB:  PubMed          Journal:  Sci Transl Med        ISSN: 1946-6234            Impact factor:   17.956


  27 in total

Review 1.  Ambulatory blood-pressure monitoring.

Authors:  Thomas G Pickering; Daichi Shimbo; Donald Haas
Journal:  N Engl J Med       Date:  2006-06-01       Impact factor: 91.245

2.  How to assess mean blood pressure properly at the brachial artery level.

Authors:  Willem J W Bos; Elisabeth Verrij; Hieronymus H Vincent; Berend E Westerhof; Gianfranco Parati; Gert A van Montfrans
Journal:  J Hypertens       Date:  2007-04       Impact factor: 4.844

3.  Arterial system of the fingers.

Authors:  B Strauch; W de Moura
Journal:  J Hand Surg Am       Date:  1990-01       Impact factor: 2.230

4.  Models of brachial to finger pulse wave distortion and pressure decrement.

Authors:  P Gizdulich; A Prentza; K H Wesseling
Journal:  Cardiovasc Res       Date:  1997-03       Impact factor: 10.787

5.  Noninvasive arterial blood pressure waveform monitoring using two- element ultrasound system.

Authors:  Joohyun Seo; Sabino J Pietrangelo; Hae-Seung Lee; Charles G Sodini
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2015-04       Impact factor: 2.725

6.  Isolated systolic hypertension in young and middle-aged adults and 31-year risk for cardiovascular mortality: the Chicago Heart Association Detection Project in Industry study.

Authors:  Yuichiro Yano; Jeremiah Stamler; Daniel B Garside; Martha L Daviglus; Stanley S Franklin; Mercedes R Carnethon; Kiang Liu; Philip Greenland; Donald M Lloyd-Jones
Journal:  J Am Coll Cardiol       Date:  2015-02-03       Impact factor: 24.094

Review 7.  Mobile health use in low- and high-income countries: an overview of the peer-reviewed literature.

Authors:  Andrew Bastawrous; Matthew J Armstrong
Journal:  J R Soc Med       Date:  2013-04       Impact factor: 5.344

8.  Non-invasive measurement of local pulse pressure by pulse wave-based ultrasound manometry (PWUM).

Authors:  J Vappou; J Luo; K Okajima; M Di Tullio; E E Konofagou
Journal:  Physiol Meas       Date:  2011-09-09       Impact factor: 2.833

9.  Prognostic accuracy of day versus night ambulatory blood pressure: a cohort study.

Authors:  José Boggia; Yan Li; Lutgarde Thijs; Tine W Hansen; Masahiro Kikuya; Kristina Björklund-Bodegård; Tom Richart; Takayoshi Ohkubo; Tatiana Kuznetsova; Christian Torp-Pedersen; Lars Lind; Hans Ibsen; Yutaka Imai; Jiguang Wang; Edgardo Sandoya; Eoin O'Brien; Jan A Staessen
Journal:  Lancet       Date:  2007-10-06       Impact factor: 79.321

10.  Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice.

Authors:  Ramakrishna Mukkamala; Jin-Oh Hahn; Omer T Inan; Lalit K Mestha; Chang-Sei Kim; Hakan Töreyin; Survi Kyal
Journal:  IEEE Trans Biomed Eng       Date:  2015-06-05       Impact factor: 4.538

View more
  31 in total

Review 1.  Connected Health Technology for Cardiovascular Disease Prevention and Management.

Authors:  Shannon Wongvibulsin; Seth S Martin; Steven R Steinhubl; Evan D Muse
Journal:  Curr Treat Options Cardiovasc Med       Date:  2019-05-18

2.  Cuffless Blood Pressure Monitoring from an Array of Wrist Bio-Impedance Sensors Using Subject-Specific Regression Models: Proof of Concept.

Authors:  Bassem Ibrahim; Roozbeh Jafari
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2019-10-10       Impact factor: 3.833

3.  Relation between blood pressure and pulse wave velocity for human arteries.

Authors:  Yinji Ma; Jungil Choi; Aurélie Hourlier-Fargette; Yeguang Xue; Ha Uk Chung; Jong Yoon Lee; Xiufeng Wang; Zhaoqian Xie; Daeshik Kang; Heling Wang; Seungyong Han; Seung-Kyun Kang; Yisak Kang; Xinge Yu; Marvin J Slepian; Milan S Raj; Jeffrey B Model; Xue Feng; Roozbeh Ghaffari; John A Rogers; Yonggang Huang
Journal:  Proc Natl Acad Sci U S A       Date:  2018-10-15       Impact factor: 11.205

4.  Biosensors for Personal Mobile Health: A System Architecture Perspective.

Authors:  Siddarth Arumugam; David A M Colburn; Samuel K Sia
Journal:  Adv Mater Technol       Date:  2019-11-20

Review 5.  The Potential of mHealth Applications in Improving Resistant Hypertension Self-Assessment, Treatment and Control.

Authors:  Karla Santo; Julie Redfern
Journal:  Curr Hypertens Rep       Date:  2019-10-09       Impact factor: 5.369

Review 6.  Cuffless Blood Pressure Devices.

Authors:  Corey K Bradley; Daichi Shimbo; David Alexander Colburn; Daniel N Pugliese; Raj Padwal; Samuel K Sia; D Edmund Anstey
Journal:  Am J Hypertens       Date:  2022-05-10       Impact factor: 3.080

Review 7.  Wearable Devices for Ambulatory Cardiac Monitoring: JACC State-of-the-Art Review.

Authors:  Furrukh Sana; Eric M Isselbacher; Jagmeet P Singh; E Kevin Heist; Bhupesh Pathik; Antonis A Armoundas
Journal:  J Am Coll Cardiol       Date:  2020-04-07       Impact factor: 24.094

8.  PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement.

Authors:  Anand Chandrasekhar; Mohammad Yavarimanesh; Keerthana Natarajan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  IEEE Trans Biomed Eng       Date:  2020-02-28       Impact factor: 4.538

9.  Estimating Blood Pressure from the Photoplethysmogram Signal and Demographic Features Using Machine Learning Techniques.

Authors:  Moajjem Hossain Chowdhury; Md Nazmul Islam Shuzan; Muhammad E H Chowdhury; Zaid B Mahbub; M Monir Uddin; Amith Khandakar; Mamun Bin Ibne Reaz
Journal:  Sensors (Basel)       Date:  2020-06-01       Impact factor: 3.576

10.  A database of tissue-specific rhythmically expressed human genes has potential applications in circadian medicine.

Authors:  Marc D Ruben; Gang Wu; David F Smith; Robert E Schmidt; Lauren J Francey; Yin Yeng Lee; Ron C Anafi; John B Hogenesch
Journal:  Sci Transl Med       Date:  2018-09-12       Impact factor: 17.956

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.