PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement.

Anand Chandrasekhar, Mohammad Yavarimanesh, Keerthana Natarajan, Jin-Oh Hahn, Ramakrishna Mukkamala.   

Abstract

OBJECTIVE: Photo-plethysmography (PPG) sensors are often used to detect pulse transit time (PTT) for potential cuff-less blood pressure (BP) measurement. It is known that the contact pressure (CP) of the PPG sensor markedly alters the PPG waveform amplitude. The objective was to test the hypothesis that PTT detected via PPG sensors is likewise impacted by CP.
METHODS: A device was built to measure the time delay between ECG and finger PPG waveforms (i.e., pulse arrival time (PAT) - a popular surrogate of PTT) and the PPG sensor CP at different CP levels. These measurements and finger cuff BP were recorded while the CP was slowly varied in 17 healthy subjects.
RESULTS: Over a physiologic range of CP, the maximum deviations of PAT detected at the PPG foot and peak were 22±2 and 40±7 ms (p<0.05), which translate to ∼11 and ∼20 mmHg BP error based on the literature. The curve relating PAT detected at the PPG foot to CP was U-shaped with minimum near finger diastolic BP. A conceptual model accounting for finger artery viscoelasticity and nonlinearity explained this curve.
CONCLUSION: Since the regulatory bias error for BP measurement is limited to 5 mmHg, PPG sensor CP should be taken into account for cuff-less BP measurement via PTT. SIGNIFICANCE: This study suggests that widely pursued PPG-based BP measurement devices including those that detect PTT should maintain the CP or include a CP measurement in the calibration equation for deriving BP.

Entities:  

Mesh:

Year:  2020        PMID: 32142414      PMCID: PMC8856571          DOI: 10.1109/TBME.2020.2976989

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  23 in total

1.  Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure.

Authors:  R A Payne; C N Symeonides; D J Webb; S R J Maxwell
Journal:  J Appl Physiol (1985)       Date:  2005-09-01

2.  Theoretical study on the effect of sensor contact force on pulse transit time.

Authors:  Xiao-Fei Teng; Yuan-Ting Zhang
Journal:  IEEE Trans Biomed Eng       Date:  2007-08       Impact factor: 4.538

3.  Arterial system of the fingers.

Authors:  B Strauch; W de Moura
Journal:  J Hand Surg Am       Date:  1990-01       Impact factor: 2.230

Review 4.  Ballistocardiography and seismocardiography: a review of recent advances.

Authors:  Omer T Inan; Pierre-Francois Migeotte; Kwang-Suk Park; Mozziyar Etemadi; Kouhyar Tavakolian; Ramon Casanella; John Zanetti; Jens Tank; Irina Funtova; G Kim Prisk; Marco Di Rienzo
Journal:  IEEE J Biomed Health Inform       Date:  2014-10-07       Impact factor: 5.772

5.  Cuffless differential blood pressure estimation using smart phones.

Authors:  Vikram Chandrasekaran; Ram Dantu; Srikanth Jonnada; Shanti Thiyagaraja; Kalyan Pathapati Subbu
Journal:  IEEE Trans Biomed Eng       Date:  2012-08-01       Impact factor: 4.538

6.  Patient-Specific Oscillometric Blood Pressure Measurement.

Authors:  Jiankun Liu; Hao-Min Cheng; Chen-Huan Chen; Shih-Hsien Sung; Mohsen Moslehpour; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  IEEE Trans Biomed Eng       Date:  2015-10-15       Impact factor: 4.538

7.  Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring.

Authors:  Mohammad Kachuee; Mohammad Mahdi Kiani; Hoda Mohammadzade; Mahdi Shabany
Journal:  IEEE Trans Biomed Eng       Date:  2016-06-14       Impact factor: 4.538

Review 8.  Photoplethysmography Revisited: From Contact to Noncontact, From Point to Imaging.

Authors:  Yu Sun; Nitish Thakor
Journal:  IEEE Trans Biomed Eng       Date:  2015-09-15       Impact factor: 4.538

9.  Theory of the oscillometric maximum and the systolic and diastolic detection ratios.

Authors:  G Drzewiecki; R Hood; H Apple
Journal:  Ann Biomed Eng       Date:  1994 Jan-Feb       Impact factor: 3.934

10.  Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care.

Authors:  Ha Uk Chung; Bong Hoon Kim; Jong Yoon Lee; Jungyup Lee; Zhaoqian Xie; Erin M Ibler; KunHyuck Lee; Anthony Banks; Ji Yoon Jeong; Jongwon Kim; Christopher Ogle; Dominic Grande; Yongjoon Yu; Hokyung Jang; Pourya Assem; Dennis Ryu; Jean Won Kwak; Myeong Namkoong; Jun Bin Park; Yechan Lee; Do Hoon Kim; Arin Ryu; Jaeseok Jeong; Kevin You; Bowen Ji; Zhuangjian Liu; Qingze Huo; Xue Feng; Yujun Deng; Yeshou Xu; Kyung-In Jang; Jeonghyun Kim; Yihui Zhang; Roozbeh Ghaffari; Casey M Rand; Molly Schau; Aaron Hamvas; Debra E Weese-Mayer; Yonggang Huang; Seung Min Lee; Chi Hwan Lee; Naresh R Shanbhag; Amy S Paller; Shuai Xu; John A Rogers
Journal:  Science       Date:  2019-03-01       Impact factor: 47.728

View more
  7 in total

1.  Wearable Photoplethysmography for Cardiovascular Monitoring.

Authors:  Peter H Charlton; Panicos A Kyriaco; Jonathan Mant; Vaidotas Marozas; Phil Chowienczyk; Jordi Alastruey
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2022-03-11       Impact factor: 10.961

2.  A Meta-Learning Approach for Fast Personalization of Modality Translation Models in Wearable Physiological Sensing.

Authors:  Ali Akbari; Jonathan Martinez; Roozbeh Jafari
Journal:  IEEE J Biomed Health Inform       Date:  2022-04-14       Impact factor: 7.021

3.  Assessment of Calibration Models for Cuff-Less Blood Pressure Measurement After One Year of Aging.

Authors:  Mohammad Yavarimanesh; Robert C Block; Keerthana Natarajan; Lalit K Mestha; Omer T Inan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  IEEE Trans Biomed Eng       Date:  2022-05-19       Impact factor: 4.756

4.  Establishing best practices in photoplethysmography signal acquisition and processing.

Authors:  Peter H Charlton; Kristjan Pilt; Panicos A Kyriacou
Journal:  Physiol Meas       Date:  2022-05-25       Impact factor: 2.688

5.  Photoplethysmography Fast Upstroke Time Intervals Can Be Useful Features for Cuff-Less Measurement of Blood Pressure Changes in Humans.

Authors:  Keerthana Natarajan; Robert C Block; Mohammad Yavarimanesh; Anand Chandrasekhar; Lalit K Mestha; Omer T Inan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  IEEE Trans Biomed Eng       Date:  2021-12-23       Impact factor: 4.538

6.  Conventional pulse transit times as markers of blood pressure changes in humans.

Authors:  Robert C Block; Mohammad Yavarimanesh; Keerthana Natarajan; Andrew Carek; Azin Mousavi; Anand Chandrasekhar; Chang-Sei Kim; Junxi Zhu; Giovanni Schifitto; Lalit K Mestha; Omer T Inan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  Sci Rep       Date:  2020-10-02       Impact factor: 4.379

Review 7.  Assessing hemodynamics from the photoplethysmogram to gain insights into vascular age: a review from VascAgeNet.

Authors:  Peter H Charlton; Birutė Paliakaitė; Kristjan Pilt; Martin Bachler; Serena Zanelli; Dániel Kulin; John Allen; Magid Hallab; Elisabetta Bianchini; Christopher C Mayer; Dimitrios Terentes-Printzios; Verena Dittrich; Bernhard Hametner; Dave Veerasingam; Dejan Žikić; Vaidotas Marozas
Journal:  Am J Physiol Heart Circ Physiol       Date:  2021-12-24       Impact factor: 4.733

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.