Noninvasive arterial blood pressure waveform monitoring using two- element ultrasound system.

Joohyun Seo, Sabino J Pietrangelo, Hae-Seung Lee, Charles G Sodini.   

Abstract

This work details noninvasive arterial blood pressure (ABP) waveform estimation based on an arterial vessel cross-sectional area measurement combined with an elasticity measurement of the vessel, represented by pulse wave velocity (PWV), using a two-element ultrasound system. The overall ABP waveform estimation is validated in a custom-designed experimental setup mimicking the heart and an arterial vessel segment with two single element transducers, assuming a constant hemodynamic system. The estimation of local PWV using the flow-area method produces unbiased elasticity estimation of the tube in a pressure waveform comparison. The measured PWV using 16 cardiac cycles of data is 8.47 + 0.63 m/s with an associated scaling error of -1.56 + 14.0% in a direct pressure waveform comparison, showing negligible bias error on average. The distension waveform obtained from a complex cross-correlation model estimator (C3M) reliably traces small pressure changes reflected by the diameter change. The excellent agreement of an estimated pressure waveform to the reference pressure waveform suggests the promising potential of a readily available, inexpensive, and portable ABP waveform monitoring device.

Entities:  

Mesh:

Year:  2015        PMID: 25881355     DOI: 10.1109/TUFFC.2014.006904

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  5 in total

1.  Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method.

Authors:  Anand Chandrasekhar; Chang-Sei Kim; Mohammed Naji; Keerthana Natarajan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  Sci Transl Med       Date:  2018-03-07       Impact factor: 17.956

Review 2.  Cuffless Blood Pressure Monitoring: Academic Insights and Perspectives Analysis.

Authors:  Shiyun Li; Can Zhang; Zhirui Xu; Lihua Liang; Ye Tian; Long Li; Huaping Wu; Sheng Zhong
Journal:  Micromachines (Basel)       Date:  2022-07-30       Impact factor: 3.523

Review 3.  A Review of Noninvasive Methodologies to Estimate the Blood Pressure Waveform.

Authors:  Tasbiraha Athaya; Sunwoong Choi
Journal:  Sensors (Basel)       Date:  2022-05-23       Impact factor: 3.847

4.  Bio-Impedance Sensor for Real-Time Artery Diameter Waveform Assessment.

Authors:  Mugeb Al-Harosh; Marat Yangirov; Dmitry Kolesnikov; Sergey Shchukin
Journal:  Sensors (Basel)       Date:  2021-12-17       Impact factor: 3.576

5.  Non-Invasive Blood Pressure Estimation from ECG Using Machine Learning Techniques.

Authors:  Monika Simjanoska; Martin Gjoreski; Matjaž Gams; Ana Madevska Bogdanova
Journal:  Sensors (Basel)       Date:  2018-04-11       Impact factor: 3.576