| Literature DB >> 29502304 |
Guido Freckmann1, Nina Jendrike1, Annette Baumstark1, Stefan Pleus1, Christina Liebing2, Cornelia Haug1.
Abstract
INTRODUCTION: The international standard ISO 15197:2013 requires a user performance evaluation to assess if intended users are able to obtain accurate blood glucose measurement results with a self-monitoring of blood glucose (SMBG) system. In this study, user performance was evaluated for four SMBG systems on the basis of ISO 15197:2013, and possibly related insulin dosing errors were calculated. Additionally, accuracy was assessed in the hands of study personnel.Entities:
Keywords: ISO 15197; Insulin dosing errors; Lay-user; Self-monitoring of blood glucose; Surveillance error grid; System accuracy; User performance evaluation
Year: 2018 PMID: 29502304 PMCID: PMC6104257 DOI: 10.1007/s13300-018-0392-6
Source DB: PubMed Journal: Diabetes Ther Impact factor: 2.945
Subjects demographics
| Included subjects | System A | System B | System C | System D |
|---|---|---|---|---|
| Male, female ( | 54, 46 | 60, 40 | 59, 41 | 58, 42 |
| Mean age (min–max) (years) | 60.8 (29–80) | 60.3 (27–79) | 60.7 (25–80) | 61.7 (25–80) |
| Diabetes type 1, 2 ( | 36, 64 | 35, 65 | 38, 62 | 36, 64 |
| Education level ( | ||||
| Secondary education grade | 70 | 66 | 68 | 67 |
| University entrance qualification | 11 | 17 | 17 | 17 |
| University degree | 19 | 17 | 15 | 16 |
| Time since diabetes diagnosis, mean (min–max) (years) | 17.8 (1–66) | 18 (1–55) | 19.3 (1–66) | 19.3 (1–66) |
| Subjects performing at least 1 SMBG measurement per week ( | 95 | 94 | 95 | 94 |
For each system, 100 different subjects were included in data analysis
System characteristics
| System | Manufacturer’s comparison method | Enzyme test strips | Measurement conditions | Manufacturer | ||||
|---|---|---|---|---|---|---|---|---|
| # | Name | Test strip lot (expiry date) | Temperature (°C) | Humidity (%) | Hematocrit (%) | |||
| A | Accu-Chek Performa Connect | 475098 (2017-09) | Hexokinase | Glucose dehydrogenase | 8–44 | 10–90 | 10–65 | Roche Diabetes Care GmbH, Mannheim, Germany |
| B | Contour plus ONE | DP6ELHD01C (2018-05) | Glucose oxidase | Glucose dehydrogenase | 5–45 | 10–93 | 0–70 | Bayer Consumer Care AG, Basel, Switzerland |
| C | FreeStyle Optium Neo | 45001 67624 (2017-12) | Glucose oxidase | Glucose dehydrogenase | 15–40 | 10–90 | 15–65 | Abbott Diabetes Care Ltd., Witney, Oxon, UK |
| D | OneTouch Select Plus | 3997403 (2017-08) | Glucose oxidase | Glucose oxidase | 10–44 | 10–90 | 30–55 | LifeScan Europe, Division of Cilag GmbH International Zug, Switzerland |
Accuracy results based on ISO 15197:2013 criteria, results within consensus error grid (CEG) zones A and B, results within surveillance error grid zones with “no risk” or “slight lower risk”, relative bias according to Bland and Altman, and calculated insulin dosing error
| System | User | Results within | Relative bias | Calculated insulin dosing error | ||||
|---|---|---|---|---|---|---|---|---|
| ±15 mg/dl/± 15% | CEG zones A + B | SEG zones | ||||||
| No risk | Slight, lower risk | Median (50th percentile) | 99% ranges between the 0.5th and 99.5th percentile | |||||
| Risk score | ||||||||
| 0–0.5 | > 0.5–1.0 | |||||||
| % | % | n | % | Units | ||||
| A | Lay-users | 96 | 100 | 100 | 0 | − 7.6 | − 0.9 | − 2.7 to − 0.1 |
| Study personnel | 99 | 100 | 197 | 3 | − 6.8 | − 0.8 | − 2.4 to 0.0 | |
| B | Lay-users | 100 | 100 | 100 | 0 | − 2.2 | − 0.3 | − 1.0 to + 0.3 |
| Study personnel | 100 | 100 | 200 | 0 | − 2.2 | − 0.2 | − 1.0 to + 0.2 | |
| C | Lay-users | 98 | 100 | 97 | 3 | + 2.8 | + 0.3 | − 0.6 to + 1.4 |
| Study personnel | 99.5 | 100 | 198 | 2 | + 1.6 | + 0.2 | − 0.7 to +1.4 | |
| D | Lay-users | 92 | 100 | 97 | 3 | − 5.6 | − 0.6 | − 2.4 to +0.6 |
| Study personnel | 96 | 100 | 193 | 7 | − 5.3 | − 0.7 | − 2.5 to + 0.3 | |
Results are shown for measurements performed by lay-users (n = 100) and by study personnel (N = 200) when evaluated by using the manufacturer’s comparison method (hexokinase for system A, glucose oxidase for systems B, C, and D)
Fig. 1Difference plots (left side) and radar plots (right side) for the tested lot of each of the four SMBG systems when evaluated against the respective manufacturer’s comparison method (hexokinase for system A, glucose oxidase for system B, C, D). Measurements performed by lay-users (n = 100) are displayed in blue squares, measurements performed by study personnel (n = 200) are displayed in orange triangles. Difference plots: ISO 15197:2013 accuracy limits (± 15 mg/dl for BG concentrations < 100 mg/dl and ± 15% for BG concentrations ≥ 100 mg/dl) are displayed in solid lines. Radar plots: Data points show differences between SMBG measurement results and the respective comparison measurement result, absolute differences for BG concentrations < 100 mg/dl and relative differences for BG concentrations ≥ 100 mg/dl. The absolute values of the differences define the location of the data points, i.e., the distance from the center of the plot, and the sign of the differences indicates the hemisphere (positive sign, upper hemisphere; negative sign, lower hemisphere). The direction with respect to the center of the plot in which the data point lies depends on the comparison method result. In radar plots, high accuracy is represented by tightly grouped data points close to the center of the plot. The circle in dark green highlights the system accuracy limits of ISO 15197:2013 (± 15 mg/dl/± 15%)
Fig. 2Surveillance error grid analysis for the tested test strip lot of each of the four SMBG systems. Data are shown for lay-user SMBG measurements (n = 100) evaluated against the respective manufacturer’s comparison method. Colors indicate associated risk levels ranging from none (dark green) to extreme (dark red)
Fig. 3Modelled insulin dosing errors. Bars in blue (lay-users) and orange (study personnel) indicate ranges in which 99% of all dose errors were found, with the white circle showing the median dose error. Data are shown when evaluated against the respective manufacturer’s comparison method (hexokinase for system A, glucose oxidase for system B, C, D)
Human factors: number of errors in the measurement technique of 100 lay-users observed by study personnel
| Number of lay-user errors | ||||
|---|---|---|---|---|
| System A | System B | System C | System D | |
| Preparation of the measurement—testa strip’s expiry date was not checked | 90 | 97 | 87 | 101 |
| Insertion of test strips—e.g., not inserted correctly, kinked, bent | 7 | 4 | 5 | 4 |
| Blood sampling—e.g., blood drop too small, too old, smeared | 3 | 3 | 11 | 8 |
| Blood application—blood was applied incorrectly, e.g., insufficient amount, fingertip was pressed onto the test strip; blood was not applied onto the test field correctly | 17 | 20 | 38 | 35 |
| Device handling—e.g., blood application before meter was ready for use, test strip was removed before completion of the measurement, fingertip was not removed from the test trip after the signal tone | 8 | 6 | 12 | 7 |
| Number of incorrectly performed measurements as reported by subjects | 4 | 2 | 17 | 7 |
aOnly non-expired test strips were used in this study; therefore, not checking the test strip’s expiry date had no impact on measurement results