Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting.

Christoph Hasslacher1, Felix Kulozik2, Isabel Platten2.   

Abstract

We investigated the analytical accuracy of 27 glucose monitoring systems (GMS) in a clinical setting, using the new ISO accuracy limits. In addition to measuring accuracy at blood glucose (BG) levels < 100 mg/dl and > 100 mg/dl, we also analyzed devices performance with respect to these criteria at 5 specific BG level ranges, making it possible to further differentiate between devices with regard to overall performance. Carbohydrate meals and insulin injections were used to induce an increase or decrease in BG levels in 37 insulin-dependent patients. Capillary blood samples were collected at 10-minute intervals, and BG levels determined simultaneously using GMS and a laboratory-based method. Results obtained via both methods were analyzed according to the new ISO criteria. Only 12 of 27 devices tested met overall requirements of the new ISO accuracy limits. When accuracy was assessed at BG levels < 100 mg/dl and > 100 mg/dl, criteria were met by 14 and 13 devices, respectively. A more detailed analysis involving 5 different BG level ranges revealed that 13 (48.1%) devices met the required criteria at BG levels between 50 and 150 mg/dl, whereas 19 (70.3%) met these criteria at BG levels above 250 mg/dl. The overall frequency of outliers was low. The assessment of analytical accuracy of GMS at a number of BG level ranges made it possible to further differentiate between devices with regard to overall performance, a process that is of particular importance given the user-centered nature of the devices' intended use.
© 2014 Diabetes Technology Society.

Entities:  

Keywords:  ISO-criteria; accuracy; blood glucose; glucose meter; self-monitoring

Mesh:

Substances:

Year:  2014        PMID: 24876607      PMCID: PMC4455429          DOI: 10.1177/1932296814522804

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  15 in total

1.  The need for separate performance goals for glucose sensors in the hypoglycemic, normoglycemic, and hyperglycemic ranges.

Authors:  David C Klonoff
Journal:  Diabetes Care       Date:  2004-03       Impact factor: 19.112

2.  Safety and accuracy of a new long-term subconjunctival glucose sensor.

Authors:  Christoph Hasslacher; Gerd Auffarth; Isabell Platten; Tanja Rabsilber; Beate Smith; Felix Kulozik; Monika Knuth; Katharina Nikolaus; Achim Müller
Journal:  J Diabetes       Date:  2012-09       Impact factor: 4.006

Review 3.  Self-monitoring of blood glucose with a focus on analytical quality: an overview.

Authors:  Gunn B B Kristensen; Sverre Sandberg
Journal:  Clin Chem Lab Med       Date:  2010-07       Impact factor: 3.694

4.  Accuracy in blood glucose measurement: what will a tightening of requirements yield?

Authors:  Lutz Heinemann; Volker Lodwig; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

5.  New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting, October 28, 2011.

Authors:  John Walsh; Ruth Roberts; Robert A Vigersky; Frank Schwartz
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

6.  Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197.

Authors:  Annette Baumstark; Stefan Pleus; Christina Schmid; Manuela Link; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

7.  System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

8.  The accuracy of home glucose meters in hypoglycemia.

Authors:  Alper Sonmez; Zeynep Yilmaz; Gokhan Uckaya; Selim Kilic; Serkan Tapan; Abdullah Taslipinar; Aydogan Aydogdu; Mahmut Yazici; Mahmut Ilker Yilmaz; Muhittin Serdar; M Kemal Erbil; Mustafa Kutlu
Journal:  Diabetes Technol Ther       Date:  2010-08       Impact factor: 6.118

9.  Accuracy of home blood glucose meters during hypoglycemia.

Authors:  Z Trajanoski; G A Brunner; R J Gfrerer; P Wach; T R Pieber
Journal:  Diabetes Care       Date:  1996-12       Impact factor: 19.112

10.  Accuracy of self monitoring blood glucose systems in a clinical setting: application of new planned ISO- standards.

Authors:  Christoph Hasslacher; Felix Kulozik; Isabel Platten
Journal:  Clin Lab       Date:  2013       Impact factor: 1.138

View more
  9 in total

1.  Surveillance of the System Accuracy of Two Systems for Self-Monitoring of Blood Glucose After Market Approval.

Authors:  Gillian Jeanny; Peter Hope
Journal:  J Diabetes Sci Technol       Date:  2015-09-30

Review 2.  Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Malte Rutschmann; Cornelia Haug; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2015-04-14

3.  System Accuracy Evaluation of Four Systems for Self-Monitoring of Blood Glucose Following ISO 15197 Using a Glucose Oxidase and a Hexokinase-Based Comparison Method.

Authors:  Manuela Link; Christina Schmid; Stefan Pleus; Annette Baumstark; Delia Rittmeyer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-04-14

4.  Strengths and Limitations of New Approaches for Graphical Presentation of Blood Glucose Monitoring System Accuracy Data.

Authors:  Stefan Pleus; Frank Flacke; Jochen Sieber; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2017-04-26

Review 5.  Performance of Cleared Blood Glucose Monitors.

Authors:  David C Klonoff; Priya Prahalad
Journal:  J Diabetes Sci Technol       Date:  2015-06-08

6.  Accuracy Evaluation of 19 Blood Glucose Monitoring Systems Manufactured in the Asia-Pacific Region: A Multicenter Study.

Authors:  Wang Yu-Fei; Jia Wei-Ping; Wu Ming-Hsun; Chien Miao-O; Hsieh Ming-Chang; Wang Chi-Pin; Lee Ming-Shih
Journal:  J Diabetes Sci Technol       Date:  2017-05-08

7.  A Post-Marketing Surveillance Study to Evaluate Performance of the EXIMO™ Blood Glucose Monitoring System.

Authors:  Sonia R Chandnani; C D Ramakrishna; Bhargav A Dave; Pankaj S Kothavade; Ashok S Thakkar
Journal:  J Clin Diagn Res       Date:  2017-05-01

8.  Clinical Study of a High Accuracy Green Design Blood Glucose Monitor Using an Innovative Optical Transmission Absorbance System.

Authors:  Takeyuki Moriuchi; Yuto Otaki; Hiroya Satou; Fumihiko Chai; Yuma Hayashida; Ryokei Aikawa; Takayuki Sugiyama; Koji Sode
Journal:  J Diabetes Sci Technol       Date:  2021-12-10

9.  User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors.

Authors:  Guido Freckmann; Nina Jendrike; Annette Baumstark; Stefan Pleus; Christina Liebing; Cornelia Haug
Journal:  Diabetes Ther       Date:  2018-03-03       Impact factor: 2.945

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.