Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose.

J C Boyd1, D E Bruns.   

Abstract

BACKGROUND: Proposed quality specifications for glucose meters allow results to be in error by 5-10% or more of the "true" concentration. Because meters are used as aids in the adjustment of insulin doses, we aimed to characterize the quantitative effect of meter error on the ability to identify the insulin dose appropriate for the true glucose concentration.
METHODS: Using Monte Carlo simulation, we generated random "true" glucose values within defined intervals. These values were converted to "measured" glucose values using mathematical models of glucose meters having defined imprecision (CV) and bias. For each combination of bias and imprecision, 10,000-20,000 true and measured glucose concentrations were matched with the corresponding insulin doses specified by selected insulin-dosing regimens. Discrepancies in prescribed doses were counted and their frequencies plotted in relation to bias and imprecision.
RESULTS: For meters with a total analytical error of 5%, dosage errors occurred in approximately 8-23% of insulin doses. At 10% total error, 16-45% of doses were in error. Large errors of insulin dose (two-step or greater) occurred >5% of the time when the CV and/or bias exceeded 10-15%. Total dosage error rates were affected only slightly by choices of sliding scale among insulin dosage rules or by the range of blood glucose. To provide the intended insulin dosage 95% of the time required that both the bias and the CV of the glucose meter be <1% or <2%, depending on mean glucose concentrations and the rules for insulin dosing.
CONCLUSIONS: Glucose meters that meet current quality specifications allow a large fraction of administered insulin doses to differ from the intended doses. The effects of such dosage errors on blood glucose and on patient outcomes require study.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11159768

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  56 in total

1.  Glucose information for tight glycemic control: different methods with different challenges.

Authors:  Christian Weber; Kurt Neeser
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  We need tighter regulatory standards for blood glucose monitoring, but they should be for accuracy disclosure.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

Review 3.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

Review 4.  Standards of medical care in diabetes--2012.

Authors: 
Journal:  Diabetes Care       Date:  2012-01       Impact factor: 19.112

5.  Regulatory controversies surround blood glucose monitoring devices.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

6.  Individuals achieve more accurate results with meters that are codeless and employ dynamic electrochemistry.

Authors:  Anoop Rao; Meg Wiley; Sridhar Iyengar; Dan Nadeau; Julie Carnevale
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

7.  Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study.

Authors:  Marc D Breton; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

8.  Standards of medical care in diabetes--2011.

Authors: 
Journal:  Diabetes Care       Date:  2011-01       Impact factor: 19.112

9.  Seven-Year Surveillance of the Clinical Performance of a Blood Glucose Test Strip Product.

Authors:  Steven Setford; Mike Grady; Stuart Phillips; Lesley Miller; Stephen Mackintosh; Hilary Cameron; Krisna Corrigall
Journal:  J Diabetes Sci Technol       Date:  2017-04-13

10.  Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study.

Authors:  Enrique Campos-Náñez; Kurt Fortwaengler; Marc D Breton
Journal:  J Diabetes Sci Technol       Date:  2017-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.