This study aimed to demonstrate the higher accuracy and reproducibility of quantitative computed tomography (QCT) compared with dual-energy X-ray absorptiometry (DXA) as a gold standard for measuring canine bone mineral density (BMD). Seven middle-aged beagle dogs underwent lumbar vertebral and bilateral femoral DXA and QCT scans. BMD (mg/cm2) was measured at the vertebral body from L2 to L6, femoral neck, and proximal and distal femoral diaphyses. The BMD values were measured 3 times and compared. The BMD value on QCT was higher than that on DXA for femoral BMD but not for vertebral BMD. The correlation was strong for the lumbar vertebrae (r=0.66) and was strongest for L3 (r=0.85). No correlation was found for the femoral neck (P=0.35), and only moderate correlations were found for the proximal and distal femoral diaphyses (r=0.43 and r=0.40, respectively). The limits of agreement were narrower for vertebral BMD than for femoral BMD, and L3 had the narrowest limits of agreement. The intraclass correlation (ICC) was higher for DXA than for QCT at all lumbar and femoral sites measured, but the ICC of QCT was higher than 0.7. In conclusion, L3 can be used to monitor changes in BMD, and relative values and sequential monitoring of femoral BMD can also be useful because of the high reproducibility of QCT measurements. QCT would be a useful technique for evaluation of BMD in veterinary practice.
This study aimed to demonstrate the higher accuracy and reproducibility of quantitative computed tomography (QCT) compared with dual-energy X-ray absorptiometry (DXA) as a gold standard for measuring canine bone mineral density (BMD). Seven middle-aged beagle dogs underwent lumbar vertebral and bilateral femoral DXA and QCT scans. BMD (mg/cm2) was measured at the vertebral body from L2 to L6, femoral neck, and proximal and distal femoral diaphyses. The BMD values were measured 3 times and compared. The BMD value on QCT was higher than that on DXA for femoral BMD but not for vertebral BMD. The correlation was strong for the lumbar vertebrae (r=0.66) and was strongest for L3 (r=0.85). No correlation was found for the femoral neck (P=0.35), and only moderate correlations were found for the proximal and distal femoral diaphyses (r=0.43 and r=0.40, respectively). The limits of agreement were narrower for vertebral BMD than for femoral BMD, and L3 had the narrowest limits of agreement. The intraclass correlation (ICC) was higher for DXA than for QCT at all lumbar and femoral sites measured, but the ICC of QCT was higher than 0.7. In conclusion, L3 can be used to monitor changes in BMD, and relative values and sequential monitoring of femoral BMD can also be useful because of the high reproducibility of QCT measurements. QCT would be a useful technique for evaluation of BMD in veterinary practice.
Entities:
Keywords:
bone mineral density (BMD); dog; dual-energy X-ray absorptiometry (DXA); quantitative computed tomography (QCT)
Bone mineral density (BMD) is defined as the mineral concentration in bone. BMD is directly
related to bone strength and is a useful predictor of osteoporotic fracture; it is therefore
used to diagnose and monitor osteoporosis in humans [12]. In human medicine, BMD is known to be affected by numerous factors, including
age, sex, endocrine disease, gastrointestinal disease, and certain drugs [33]. Osteoporosis, or low BMD, is a common condition that
puts the patient at increased risk of pathologic fracture; therefore, early diagnosis,
prevention, and monitoring of BMD are crucial.Dual-energy X-ray absorptiometry (DXA) is a standard, non-invasive, and accurate method for
measuring BMD and body composition in humans [18].
Typically, central DXA scans (of the lumbar vertebrae and proximal femur) are obtained, but
DXA can be used to assess BMD throughout the entire body or at any specific body site [16]. This method has several advantages, including
cost-effectiveness and a rapid scanning time.Quantitative computed tomography (QCT) is also used to measure BMD in humans and has higher
sensitivity than DXA, but lower specificity for diagnosing osteoporosis [15]. Although DXA is considered a standard BMD measurement technique, QCT
is more sensitive than DXA for diagnosis of osteoporosis and prediction of the risk of
pathologic fracture because trabecular BMD is lost more rapidly than cortical BMD when the
disease progresses [6]. The decrease in BMD in patients
with metabolic or endocrine disease is more obvious in trabecular bone than in cortical bone,
particularly in the vertebrae [10]. Measurement of
trabecular BMD is critical for early detection of decreasing bone mineral content and QCT may
be preferable to DXA in some cases.Similarly, there are many conditions in veterinary medicine that can reduce BMD, including
hyperadrenocorticism, hyperparathyroidism, diabetes mellitus, renal failure, nutritional
deficiency, and disuse, in addition to ovariohysterectomy and the administration of
glucocorticoids [4, 5, 7, 9,
24, 32, 36]. Therefore, BMD measurements are useful for early
diagnosis and treatment of osteoporosis in animals. Conventional radiographs are often used in
veterinary medicine to diagnose osteopenia in dogs. However, to detect decreased bone opacity
by using conventional radiographs, at least 30–40% of bone mineral content must be depleted
[18]. Furthermore, conventional radiography is
subjective and affected by technical factors [27]. In
recent years, several studies have been conducted on BMD measurement in dogs [2, 19]. As in human
medicine, several endocrine diseases are associated with decreased BMD in dogs and bone loss
is most easily detected in the vertebrae [7]. Therefore,
a more sensitive method for measuring BMD and evaluating osteopenia is needed in veterinary
medicine.Both QCT and DXA are presently used to measure BMD in veterinary medicine, and most studies
using these techniques have evaluated BMD in the lumbar vertebrae and proximal femur, as in
human medicine [5, 8, 11, 35]. Assessment of peripheral BMD is more common in veterinary medicine than in
human medicine because of the greater need to monitor bone healing postoperatively [29]. However, in veterinary medicine, DXA is used less
often than QCT for this purpose because veterinary DXA machines are not commercially available
and scanning can be performed only under general anesthesia. QCT has been used more frequently
in veterinary medicine because it has the advantages of providing anatomical information and
measuring trabecular and cortical BMD separately [2,
3]. Measurement of BMD is critical for identifying
factors that affect bone mineral content, but research in this regard remains rare in
veterinary medicine. Furthermore, the only study that has compared measurements obtained using
DXA and QCT, and assessed which is more commonly used, was performed on canine femurs ex situ
[26], and to our knowledge, no studies have compared
vertebral BMD measurements obtained using QCT and DXA.The aims of this study were to demonstrate the accuracy and reproducibility of QCT compared
with DXA as the gold standard and to assess the clinical usefulness of QCT for evaluating
canineBMD in veterinary medicine.
MATERIALS AND METHODS
Animals
Seven middle-aged adult male beagle dogs with a mean age of 4.9 ± 0.98 years and a mean
weight of 10.15 ± 0.68 kg were used in the study. All dogs were deemed to be healthy on
the basis of a physical examination, complete blood count, serum chemistry, urinalysis,
and radiographic and ultrasonographic examinations. The study protocol was approved by the
Institutional Animal Care and Use Committee of Konkuk University (approval number
KU16175).
Anesthesia
All dogs were fasted for at least 12 hr before QCT or DXA scanning was performed. The
dogs were premedicated with acepromazine (Mobinul®, Myungmoon, Seoul, South
Korea) at 0.02 mg/kg and glycopyrrolate (SEDAJECT®, Samu Median, Yesan, South
Korea) at 0.01 mg/kg, administered via intramuscular injection before anesthesia. The dogs
were anesthetized using a combination of tiletamine and zolazepam (Zoletil®,
Virbac Animal Health, Carros, France) at 1.5 mg/kg and medetomidine (Domitor®,
Pfizer Animal Health, Walton Oaks, U.K.) at 0.03 mg/kg, administered via intramuscular
injection.
Quantitative computed tomography examination
QCT images were obtained using a 4-channel multidetector CT scanner (Light-speed
Plus®, GE Healthcare, Amersham, U.K.). The lumbar vertebrae were scanned in
dorsal recumbency and the femur in lateral recumbency (Fig. 1). A QCT phantom (QRM-BDC/3®, QRM GmbH, Moehrendorf, Germany) was placed
under the vertebrae and femur (Fig. 1). The
scanning conditions were as follows: 120 kV and 200 mA, slice thickness 1.25 mm, slice
interval 1.25 mm, pitch 1.5:1, rotation time 0.6 sec, and scanning speed 7.5 mm/rotation.
The phantom and femur were positioned such that their axes were perpendicular to each
other and reconstructed in the transverse plane. The phantom and lumbar vertebrae were
positioned such that their axes were parallel to each other. All QCT images were scanned
with the bone and beam placed close to vertical without tilting of the gantry.
Fig. 1.
Positioning for quantitative computed tomography (QCT) and dual X-ray
absorptiometry (DXA) scanning of the lumbar vertebrae and femur. Positioning of the
dog and the bone density calibration phantom (arrows) when scanning the lumbar
vertebrae (A and B) and femur (C and D) using QCT. Positioning of the dog and the
bone density calibration phantom when scanning the lumbar vertebrae (E) and femur
(F) using DXA. DXA, dual X-ray absorptiometry; QCT, quantitative computed
tomography.
Positioning for quantitative computed tomography (QCT) and dual X-ray
absorptiometry (DXA) scanning of the lumbar vertebrae and femur. Positioning of the
dog and the bone density calibration phantom (arrows) when scanning the lumbar
vertebrae (A and B) and femur (C and D) using QCT. Positioning of the dog and the
bone density calibration phantom when scanning the lumbar vertebrae (E) and femur
(F) using DXA. DXA, dual X-ray absorptiometry; QCT, quantitative computed
tomography.
Dual-energy X-ray absorptiometry examination
The DXA scans were performed after the QCT scans. The stability of the DXA device
(PRIMUS®, OsteoSys, Seoul, South Korea) was checked using a calibration
phantom (PRIMUS Device PHANTOM®, OsteoSys) that was regularly scanned during
the study period. The animals were positioned in lateral recumbency for scanning of the
lumbar vertebrae and in dorsal recumbency with internal rotation of the hind limbs for
scanning of the femurs (Fig. 1). The scanning
conditions were 83 kV and 1.0 mA for the vertebral scans and 83 kV and 0.2 mA for the
femoral scans.
Image analysis
Quantitative computed tomography: The region of interest (ROI) for QCT
included only the vertebral body in the lumbar region and was measured using a slice at
the origin of the transverse process (Fig. 2). The cortical and trabecular bone at all measurement sites were included in the
ROI. In the femur, measurements were performed in the middle of the femoral neck,
including one third of the proximal diaphysis, and one third of the distal diaphysis
(Fig. 2). BMD (mg/cm2) was
calculated using Hounsfield units (HU), as in prior studies [6, 14]. All ROIs were drawn by
hand and measured 3 times.
Fig. 2.
Quantitative computed tomography (QCT) images of the lumbar vertebrae and femur.
Vertebral bone mineral density (BMD) (A) was measured at the origin of the
transverse process. Note the different attenuation according to the HA content of
the phantom placed under the dog. Femoral BMD (B) was measured at the femoral neck,
the center of the proximal third of the diaphysis, and the center of the distal
third of the diaphysis in transverse images. The solid line indicates the femoral
neck, the dotted line indicates the proximal diaphysis, and the double line
indicates the distal diaphysis. BMD, bone mineral density; HA, hydroxyapatite; QCT,
quantitative computed tomography.
Quantitative computed tomography (QCT) images of the lumbar vertebrae and femur.
Vertebral bone mineral density (BMD) (A) was measured at the origin of the
transverse process. Note the different attenuation according to the HA content of
the phantom placed under the dog. Femoral BMD (B) was measured at the femoral neck,
the center of the proximal third of the diaphysis, and the center of the distal
third of the diaphysis in transverse images. The solid line indicates the femoral
neck, the dotted line indicates the proximal diaphysis, and the double line
indicates the distal diaphysis. BMD, bone mineral density; HA, hydroxyapatite; QCT,
quantitative computed tomography.Dual-energy X-ray absorptiometry: The L2–L6 lumbar vertebrae, bilateral
femoral neck, and bilateral proximal and distal thirds of the diaphysis were measured
using QCT and DXA. To avoid interference of the transverse process and endplate, the ROI
for the lumbar vertebrae when scanned using DXA was set to be trapezoidal-shaped to
triangular-shaped (Fig. 3), whereas the femoral ROIs were set to be trapezoidal-shaped to rectangular-shaped
(Fig. 3). The size of the ROI was set to
approximately 0.55 cm2 for the lumbar vertebrae, 0.45 cm2 for the
femoral neck, 0.7 cm2 for the proximal diaphysis, and 1.0 cm2 for
the distal diaphysis. All ROIs were drawn by hand and measured 3 times.
Fig. 3.
Dual X-ray absorptiometry (DXA) images of the lumbar vertebrae and femur. Vertebral
bone mineral density (BMD) (A) was measured except for the transverse process and
endplate. Femoral BMD (B) was measured at the femoral neck, the center of the
proximal diaphysis, and the center of the distal diaphysis. The color map indicates
the degree of X-ray attenuation; low X-ray attenuation appears blue, whereas high
X-ray attenuation appears red. The solid line indicates the femoral neck, the dotted
line indicates the proximal diaphysis, and the double line indicates the distal
diaphysis. BMD, bone mineral density; DXA, dual X-ray absorptiometry; TP, transverse
process.
Dual X-ray absorptiometry (DXA) images of the lumbar vertebrae and femur. Vertebral
bone mineral density (BMD) (A) was measured except for the transverse process and
endplate. Femoral BMD (B) was measured at the femoral neck, the center of the
proximal diaphysis, and the center of the distal diaphysis. The color map indicates
the degree of X-ray attenuation; low X-ray attenuation appears blue, whereas high
X-ray attenuation appears red. The solid line indicates the femoral neck, the dotted
line indicates the proximal diaphysis, and the double line indicates the distal
diaphysis. BMD, bone mineral density; DXA, dual X-ray absorptiometry; TP, transverse
process.
Statistical analysis
The statistical comparisons of QCT and DXA were performed using commercially available
software (IBM SPSS 24®, IBM, New York, NY, U.S.A.). The correlation and
agreement between the 2 modalities were determined using Pearson correlation and
Bland–Altman analysis, respectively. All the data were tested for normality before
analysis. All lumbar and femoral sites scanned using QCT and DXA were compared.
Interobserver agreement was assessed using 2-way random intraclass correlation
coefficients (ICCs). For all comparisons, the level of significance was set at
P<0.05.
RESULTS
In total, 231 lumbar and femoral BMD measurements were obtained from the 7 beagle dogs. The
average BMD values obtained using QCT and DXA are shown in Table 1. With the exception of the vertebrae, the BMD values obtained using QCT were
higher than those obtained using DXA, particularly in the femur. When DXA was used, femoral
BMD was lower than vertebral BMD, and when QCT was used, femoral BMD was higher than
vertebral BMD. In the lumbar spine, L5 had the highest BMD, and in the femur, the proximal
diaphysis exhibited the highest BMD, followed by the distal diaphysis.
Table 1.
Bone mineral density at the lumbar vertebrae and femur obtained using
quantitative computed tomography and dual X-ray absorptiometry
QCT (mg/cm2)
DXA (mg/cm2)
Lumbar vertebrae
L2
0.80 ± 0.06
0.72 ± 0.04
L3
0.82 ± 0.06
0.79 ± 0.06
L4
0.84 ± 0.07
0.84 ± 0.09
L5
0.87 ± 0.06
0.83 ± 0.08
L6
0.84 ± 0.08
0.76 ± 0.08
Mean
0.79 ± 0.56
0.79 ± 0.86
Femur
Femoral neck
0.90 ± 0.12
0.50 ± 0.06
Proximal diaphysis
1.11 ± 0.11
0.94 ± 0.10
Distal diaphysis
0.90 ± 0.12
0.65 ± 0.11
Mean
0.98 ± 0.14
0.70 ± 0.21
The data are shown as the mean and standard deviation. DXA, dual X-ray
absorptiometry; QCT, quantitative computed tomography.
The data are shown as the mean and standard deviation. DXA, dual X-ray
absorptiometry; QCT, quantitative computed tomography.Moderate correlation was observed between the BMD values obtained using QCT and DXA in the
lumbar vertebrae, weak correlation was observed at the femoral neck, and moderate
correlation was observed at the femoral diaphysis (Table 2). The correlation between QCT and DXA with regard to femoral BMD was weaker
than that for vertebral BMD. The correlation for the lumbar vertebrae was the weakest for
L2, and moderate to strong correlation was observed between QCT and DXA for the other
vertebrae. L3 showed the strongest correlation, followed by L5, L4 and L6. In the femur, the
proximal diaphysis showed the strongest correlation and the femoral neck showed the weakest
correlation. No significant correlation was observed between QCT and DXA with regard to the
femoral neck or L2.
Table 2.
Pearson correlation between quantitative computed tomography and dual X-ray
absorptiometry
DXA, dual X-ray absorptiometry; QCT, quantitative computed tomography.Figures 4 and 5 show the agreement of 2 methods through Bland-Altman plots. The limits of agreement
for vertebral BMD were narrower than those for femoral BMD. In the lumbar vertebrae, L3 was
shown to have the narrowest limits of agreement, followed by L5, and L4 had the widest
limits of agreement, followed by L2. These limits of agreement were similar to those for
femoral BMD.
Fig. 4.
Bland–Altman plot comparing vertebral bone mineral density (BMD) values obtained
using quantitative computed tomography (QCT) and dual X-ray absorptiometry (DXA). The
mean difference (solid line) and limits of agreement (dotted line) are shown. Note
that the narrowest limits of agreement are at L3, followed by L5. BMD, bone mineral
density; DXA, dual X-ray absorptiometry; QCT, quantitative computed tomography.
Fig. 5.
Bland–Altman plot comparing femoral bone mineral density (BMD) values obtained using
quantitative computed tomography (QCT) and dual X-ray absorptiometry (DXA). The mean
difference (solid line) and limits of agreement (dotted line) are shown. The limits of
agreement at the 3 femoral sites were similar with each other and wider than those at
the lumbar vertebrae. BMD, bone mineral density; DXA, dual X-ray absorptiometry; QCT,
quantitative computed tomography.
Bland–Altman plot comparing vertebral bone mineral density (BMD) values obtained
using quantitative computed tomography (QCT) and dual X-ray absorptiometry (DXA). The
mean difference (solid line) and limits of agreement (dotted line) are shown. Note
that the narrowest limits of agreement are at L3, followed by L5. BMD, bone mineral
density; DXA, dual X-ray absorptiometry; QCT, quantitative computed tomography.Bland–Altman plot comparing femoral bone mineral density (BMD) values obtained using
quantitative computed tomography (QCT) and dual X-ray absorptiometry (DXA). The mean
difference (solid line) and limits of agreement (dotted line) are shown. The limits of
agreement at the 3 femoral sites were similar with each other and wider than those at
the lumbar vertebrae. BMD, bone mineral density; DXA, dual X-ray absorptiometry; QCT,
quantitative computed tomography.The ICC results indicated that both QCT and DXA had high reproducibility in all parts of
the lumbar vertebrae and femur (Table
3). However, the ICC for DXA was consistently higher than that for QCT at all
anatomical sites measured. On QCT, the vertebral ICC was higher than the femoral ICC, and on
DXA, the vertebral ICC was lower than the femoral ICC.
Table 3.
Intraclass correlation of bone mineral density values at each anatomical
measurement site
ICC for QCT
ICC for DXA
Lumbar vertebrae
L2
0.71
0.81
L3
0.91
0.83
L4
0.93
0.97
L5
0.80
0.93
L6
0.87
0.87
Mean
0.84
0.94
Femur
Femoral neck
0.72
0.99
Proximal diaphysis
0.75
0.97
Distal diaphysis
0.87
0.97
Mean
0.87
0.99
The P-value for all ICC values was <0.05. DXA, dual X-ray
absorptiometry; ICC, intraclass coefficient; QCT, quantitative computed
tomography.
The P-value for all ICC values was <0.05. DXA, dual X-ray
absorptiometry; ICC, intraclass coefficient; QCT, quantitative computed
tomography.
DISCUSSION
Femoral BMD values have been compared between QCT and DXA previously, but only in an ex
situ study [26]. No comparative studies of QCT and
DXA have been reported, including any comparison of the femur and vertebrae. In veterinary
medicine, more reports have focused on QCT than on DXA [3, 8, 19, 25, 26], and although the accessibility of QCT in veterinary clinical practice is
greater than that of DXA, no reports have addressed the accuracy and usefulness of QCT.
Therefore, the value of QCT in veterinary medicine requires evaluation.The BMD values obtained using QCT were the highest at L5 in the lumbar vertebrae, which is
consistent with a previous study [19]. The vertebral
BMD was similar between QCT and DXA, but the femoral BMD was higher in QCT than in DXA. The
reason for the difference in the BMD values at the femur may be that the BMD measurements
obtained using DXA are affected by inhomogeneous fat tissue, tissue depth, and extraskeletal
calcifications [17, 30, 31] in addition to the thickness, mass,
and length of the bone [21, 26]. The thicker the surrounding soft tissue, the higher the attenuation,
resulting in a higher BMD value. This error occurs because DXA converts 3-dimensional data
to 2-dimensional data and can explain higher vertebral BMD than femoral BMD data obtained
using DXA can. In addition, the ROIs were relatively small in relation to image size and
limited to rectangular, trapezoidal, or triangular shapes in the DXA images. When measuring
BMD on QCT images, the observers attempted to measure the entire image except for blurred
areas. High image magnification was necessary to draw the ROI manually, and during this
process, except for blurring, interference between surrounding artefacts and soft tissue was
minimized. Beam hardening artefacts can also affect BMD measurements obtained using QCT
[3]. These limitations and differences may
contribute to differences in the results obtained using QCT and DXA. The cortical bone of
the femur is thicker than that of the vertebrae; therefore, beam hardening artefacts are
more prominent on femoral QCT images than on vertebral QCT images. These artefacts could
explain why femoral BMD was higher than vertebral BMD when measured using QCT. The reason
for the lower BMD value obtained by DXA at the distal diaphysis when compared with the
proximal diaphysis of the femur is that the bone in the former is thinner and wider than
that in the latter [22]. Furthermore, some homotypic
variation may occur between the right femur and the left femur, which is more evident at the
distal diaphysis than at the proximal diaphysis and femoral neck because differences in
total bone diameter, medullary diameter, and cortical width are more pronounced at the
distal diaphysis [26].The correlation and agreement of BMD values between QCT and DXA were stronger in the
vertebrae than in the femur, particularly at L3. This was because the observers could not
measure the femoral BMD in exactly the same cross-section on QCT, and when measuring BMD in
the lumbar vertebrae on DXA images, the transverse process of the next vertebra to be
measured interfered with the vertebra currently being measured. This interference was the
lowest at L3. By contrast, the correlation and agreement at L4–L6 were likely caused by the
transverse process of the next vertebra interfering to a greater extent than occurred at L3.
Furthermore, the degree of interference on DXA varies from patient to patient. On QCT
images, the vertebral body becomes smaller in the caudal lumbar vertebrae, which causes the
ROI to become smaller. These factors could have contributed to the weaker correlation and
agreement between QCT and DXA at L4–L6. The correlation and agreement were the weakest in
the vertebral body of L2 and were not statistically significant, suggesting that in some of
the beagle dogs the last rib interfered with the vertebral body measurements obtained using
DXA. The correlation and agreement were weaker for the femur than for the lumbar vertebral
body, and were significant at all sites except at the femoral neck. The lack of significance
at the femoral neck may reflect the fact that the QCT image section was not imaged at the
correct transverse section. This suggests that blurring around the cortex of the femoral
neck was more severe than at other parts of the femoral diaphysis on the QCT images. The BMD
measurements may also have been affected by beam hardening artefacts [3].The reproducibility of QCT was lower than that of DXA; however, QCT demonstrated high
reproducibility. This is because DXA is the gold standard for the measurement of BMD. On the
QCT image, vertebral BMD was measured at the origin of the transverse process and femoral
BMD was measured at the center of each point. Reproducibility was particularly high in the
lumbar vertebrae and at the femoral neck because, typically, only one cross-section could be
taken as a reference on QCT. However, reproducibility was high at the diaphysis despite the
difficulty in obtaining measurements at the same position. Therefore, it is concluded that
reproducible and reliable results can be obtained if radiologists have a detailed
understanding of QCT images and know exactly which sites to measure. In this study,
measurement of BMD using QCT was highly reproducible, indicating its diagnostic value.The DXA scanning protocols used in this study were based on those adopted in several prior
studies [10, 13, 18]. In the veterinary literature,
studies on BMD measured using DXA have been fewer than those using QCT, probably because QCT
is more readily accessible than DXA in veterinary medicine. Although DXA is the more
accurate method for BMD measurement, it has several limitations. First, positional errors
can occur in vertebral and femoral DXA scans; in patients with rotation of the vertebrae
such as scoliosis, lordosis, and kyphosis, the vertebrae cannot be positioned appropriately
[33]. In human medicine, positioning devices are
used when the femur is scanned [33], but no such
devices were available for scanning the canine femur when we conducted the present study.
Furthermore, different modes of scanning can be used according to the patient’s body habitus
in human medicine [33], and various DXA scanning
modes would be needed for the various canine breeds. The anterior–posterior (AP) position is
the standard DXA scan view for the lumbar vertebrae in both human medicine and veterinary
medicine [11, 16]. However, in the present study, lateral DXA scans were performed because they
can exclude the abdominal organs and transverse processes, whereas the ventral–dorsal DXA
scan cannot exclude the spinous processes when drawing the ROI. In human medicine, several
studies have compared AP spine and lateral spine DXA scans [10, 34]. The lateral DXA scan is superior
for diagnosing osteopenia and estimating the risk of vertebral fracture because it can avoid
spinous processes and osteoarthrosis at the articular facets [10, 34]. In addition, lateral DXA
BMD has a stronger correlation with QCT BMD than AP DXA BMD [10]. Despite these advantages, lateral DXA is not used as a standard position
because it is difficult to maintain a position during DXA scanning, and the accuracy of
predicting the risk of pathologic fracture caused by osteoporosis is similar in both lateral
DXA and AP DXA [33]. However, because the dogs used
in this study were a deep-chested breed, the lateral position was much easier to maintain
than the AP position, and positional errors could be reduced. For these reasons, DXA
scanning of the lumbar vertebrae was performed in the lateral position in our study, and
additional veterinary studies on positioning and correcting positional errors are
needed.Various CT scanning conditions can affect QCT images and HU values. The scanning parameters
used for QCT, such as voltage (kV) and time current product (mAs), have been investigated in
previous studies [1, 19], and the protocols used in the present study followed this research. A greater
slice thickness is needed to obtain superior images when using QCT because the radiopacity
in thinner slices is easily affected by the inhomogeneous fat distribution in trabecular
bone [20]. However, in another study [3], no significant difference was observed in CT images
with various slice thicknesses, indicating that a thinner slice thickness can be used when
measuring BMD by using QCT in small animals. Although QCT images can be affected by scanning
conditions, QCT has relatively high accuracy for measuring BMD.This study had several limitations. First, the study population was small, and the
inclusion of only healthy dogs was likely to have affected BMD. Second, any differences in
BMD measurements obtained using QCT and DXA according to sex, age group, and breed were not
investigated. To minimize differences in these variables, beagle dogs of the same sex and
similar age were used. Higher bone mineral content in intact male dogs compared to intact
female dogs has been reported [23], and BMD has been
revealed to peak at the age of 6 years and decrease gradually thereafter [19, 23, 26]. In human medicine, the Z score or T score is used to
compensate for BMD differences in age, sex, and race when assessing BMD [1, 15, 28]. Third, BMD was not verified by measuring the actual
mineral concentration in bone. However, DXA has been reported to be accurate and is used as
a standard BMD measurement method, and studies have shown a strong correlation between the
results of DXA and actual mineral concentrations [21,
35]. Therefore, DXA was used as the gold standard
for comparison of the BMD measurements obtained using QCT in this study.In conclusion, when evaluating BMD using QCT in the region of the lumbar vertebrae, L3 is
considered to be the optimal site for monitoring BMD changes, showing high accuracy and
reproducibility. In addition, QCT scanning of proximal femoral BMD and femoral neck BMD is
considered a useful monitoring technique for evaluating relative values through comparison
of the left and right sides or calibration of BMD values. Furthermore, QCT provides more
detailed anatomical information in terms of ROI selection and measurement of BMD in dogs.
This comparative study showed that QCT has high accuracy and reproducibility in terms of BMD
measurement when DXA is used as a standard method. Therefore, QCT is a useful method for
evaluation of BMD in veterinary practice.
Authors: L A V S Costa; B F Lopes; A B Lanis; D C De Oliveira; J G Giannotti; F S Costa Journal: J Vet Pharmacol Ther Date: 2010-12 Impact factor: 1.786
Authors: Heather J Chalmers; Nathan L Dykes; George Lust; James P Farese; Nancy I Burton-Wurster; Alma J Williams; Rory J Todhunter Journal: Am J Vet Res Date: 2006-05 Impact factor: 1.156
Authors: Sigrid Schneider; Sabine M Breit; Stephan Grampp; Wolfgang W F Künzel; Annette Liesegang; Elisabeth Mayrhofer; Jürgen Zentek Journal: Am J Vet Res Date: 2004-07 Impact factor: 1.156
Authors: Vitali Koch; Nils Große Hokamp; Moritz H Albrecht; Leon D Gruenewald; Ibrahim Yel; Jan Borggrefe; Stefan Wesarg; Katrin Eichler; Iris Burck; Tatjana Gruber-Rouh; Lukas Lenga; Thomas J Vogl; Simon S Martin; Julian L Wichmann; Renate M Hammerstingl; Leona S Alizadeh; Christoph Mader; Nicole A Huizinga; Tommaso D'Angelo; Giorgio Ascenti; Silvio Mazziotti; Christian Booz Journal: Eur Radiol Exp Date: 2021-10-05