BACKGROUND: To investigate the organ dose, effective dose (ED), conversion factor, and the C-arm rotation angle effects on dose variations of abdominal C-arm cone-beam computed tomography (CBCT) during transarterial chemoembolization (TACE).Entities:
Keywords: Cone-beam computed tomography; Conversion factor; Dose–area product; Effective dose; Organ dose
Mesh:
Year: 2018 PMID: 29402236 PMCID: PMC5800092 DOI: 10.1186/s12880-018-0247-7
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Patient characteristics and mean DAPs, EDs, and ED to DAP ratios
| Patient number (Male / Female) | Age | Height (cm) | Weight (kg) | DAP (Gy·cm2) | ED (mSv) | ED to DAP ratio (mSv·Gy−1·cm− 2) | |
|---|---|---|---|---|---|---|---|
| All patients | 80 (56/24) | 65 ± 12 (36–87) | 162 ± 8 (146–183) | 67 ± 13 (48–120) | 12.9 ± 0.8 (11.8–17.1) | 3.5 ± 0.5 (2.1–4.5) | 0.27 ± 0.04 (0.17–0.35) |
| BMI < 25 kg/m2 | 45 (30/15) | 68 ± 11 (37–87) | 161 ± 7 (148–176) | 59 ± 6 (48–76) | 12.9 ± 0.6 (12.3–15.4) | 3.8 ± 0.3 (3.0–4.5) | 0.29 ± 0.02 (0.24–0.35) |
| BMI 25–30 kg/m2 | 22 (18/4) | 62 ± 12 (40–83) | 163 ± 7 (146–173) | 72 ± 7 (58–86) | 12.8 ± 0.6 (11.8–14.8) | 3.3 ± 0.4 (2.7–4.2) | 0.26 ± 0.02 (0.22–0.30) |
| BMI ≥ 30 kg/m2 | 13 (8/5) | 60 ± 13 (36–83) | 163 ± 8 (148–183) | 85 ± 13 (68–120) | 13.2 ± 1.4 (11.9–17.1) | 3.0 ± 0.4 (2.1–3.9) | 0.23 ± 0.03 (0.17–0.28) |
Data are mean ± SD (range)
Clinical settings of abdominal CBCT acquisitions for patients undergoing TACE
| Configuration values and acquisition parameters | |
|---|---|
| Acquisition mode | CB-CTAP |
| Acquisition rate (fps) | 30 |
| Radiation time (sec) | 12 |
| Rotation speed | 20°/sec |
| Number of frames | Approximate 315 frames |
| Acquisition range | 215 |
| Focus-to-image distance (FID) (cm) | 120 |
| Field of view size (inch) | 17 |
| Focus | Large focus |
| Total filtration (mm) | 11.2-mm aluminum equivalent |
| kV | 100 |
| mA | 360 |
| Pulse width (msec) | 5.6 |
Fig. 1Mean organ dose calculated using PCXMC for the CBCT acquisitions for TACE. Error bars indicate SDs from 80 patients
Fig. 2ED and ED to DAP ratio dependence on patient BMI. a An inverse linear relationship between the ED and BMI is demonstrated. Black solid line is the linear regression line between the ED and BMI, and the blue dashed lines display the range of 95% confidence interval (CI). b Box plots show the 25%–75% interquartile range of the ED to DAP ratio with patient BMI < 25 kg/m2, 25–30 kg/m2, ≥ 30 kg/m2, and with all the patients. Significant differences were observed in the ED to DAP ratio among the three BMI categories. Red and black lines indicate the mean and median values, respectively
Fig. 3Dose variations and angular dependence of the organ dose during X-ray tube rotation. a Organ dose variations from different X-ray irradiated angle, represent as CVs of the dose from the 44 simulated projections for each organ. b The distribution of the demonstrated organ dose per projection for the CBCT procedure
Previously reported EDs and ED to DAP ratios for abdominal CBCT procedures
| Authors | Procedure | Scanner model | Rotation | Dose estimation method | Effective dose (ED) (mSv) | ED to DAP ratio (mSv·Gy− 1·cm− 2) |
|---|---|---|---|---|---|---|
| This study | Abdominal CBCT imaging: CB CTAP | Shimadzu BRANSIST safireVC17 | 215o | Using PCXMC based on individual patient data | 3.5 ± 0.5 (2.1–4.5) | 0.17–0.35 |
| Suzuki et al. [ | Abdominal 3D imaging | Philips Allura Xper FD20/10 | 207 o | (1) Placing TLD in the human-shaped phantom (S) | (1) TLD: 1.6 | 0.37–0.45 |
| GE INNOVA 4100 | 194o | (1) TLD: 2.0 | 0.26–0.32 | |||
| Siemens AXIOM Artis dTA | 200o | (1) TLD: 2.6 | 0.13–0.15 | |||
| Braak et al. [ | CBCT guidance (upper abdomen) | Philips XperCT Allura FD20 | 240o | Using PCXMC based on individual patient data | 4.2 (95% CI 3.8–4.6) | N/A |
| Kwok et al. [ | Abdominal CBCT imaging: DynaCT 8-s DR | Siemens Artis zeego | 200o | Placing TLD in the human-shaped phantom | 15 | N/A |
| Abdominal CBCT imaging: LCI CTHA Low 10 s | Toshiba Infinix VC-i | 200o | 25.4 | N/A | ||
| Sailer et al. [ | CT abdomen LD roll | Philips XperCT Allura FD20 | 180o | Using PCXMC based on individual patient data | 4.3 (95% CI 3.9–4.8) | N/A |
TLD Thermoluminescent Dosimeter, N/A not available, S small phantom size, M medium phantom size, L large phantom size