[Comparison of two optical biometric devices for intraocular lens calculation].

M-S Uckmann1, M Stattin2, C Zehetner3, S Neururer4, L Speicher3.   

Abstract

BACKGROUND: Modern cataract surgery not only consists of a minimally invasive lens extraction but also of the implantation of a suitable intraocular lens.
OBJECTIVE: The aim of this prospective trial was a comparison of the predicted refractive error of two optical biometers, the IOLMaster 500 and LenStar LS 900 for intraocular lens power calculation in cataract surgery.
MATERIAL AND METHODS: This was a prospective, analytical, comparative, non-masked study. A total of 86 eyes of 86 patients were examined and measured with both instruments before and after uneventful cataract surgery. Primary outcome measures were the differences of the predicted refractive error of both instruments. The predicted refractive error was calculated with different formulas. The results were compared to each other, to the desired target refraction as well as to the postoperative spherical equivalent.
RESULTS: The mean differences in predicted refractive error of both instruments varied between 0.9 ± 0.19 (standard deviation) diopters (D) and 0.18 ± 0.30 D depending on the chosen formula. The IOLMaster 500 predicted less difference to the desired target refraction as well as to the spherical equivalent than the LenStar LS 900 with nearly all formulas.
CONCLUSION: Both devices generated reproducible exact data with only a small deviation from the desired target refraction and from the postoperative spherical equivalent. There were statistically significant differences based on the chosen a‑constants as well as the utilized measurement methods of both instruments.

Entities:  

Keywords:  IOLMaster 500; LenStar LS 900; Optical biometry; Predicted refractive error

Mesh:

Year:  2019        PMID: 29372303     DOI: 10.1007/s00347-018-0655-7

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  24 in total

1.  Assessment of anterior chamber depth using Visante optical coherence tomography, slitlamp optical coherence tomography, IOL Master, Pentacam and Orbscan IIz.

Authors:  Umut Asli Dinc; Ebru Gorgun; Banu Oncel; Melda Nursal Yenerel; Levent Alimgil
Journal:  Ophthalmologica       Date:  2010-05-04       Impact factor: 3.250

2.  Comparison of two partial coherence interferometry devices for ocular biometry.

Authors:  Z Liampa; M Kynigopoulos; G Pallas; H Gerding
Journal:  Klin Monbl Augenheilkd       Date:  2010-04-20       Impact factor: 0.700

3.  Accuracy of intraocular lens power prediction using the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas.

Authors:  Julio Narváez; Grenith Zimmerman; R Doyle Stulting; Daniel H Chang
Journal:  J Cataract Refract Surg       Date:  2006-12       Impact factor: 3.351

4.  Accuracy of a new partial coherence interferometry analyser for biometric measurements.

Authors:  M P Holzer; M Mamusa; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2009-03-15       Impact factor: 4.638

5.  Eye-length measurement by interferometry with partially coherent light.

Authors:  A F Fercher; K Mengedoht; W Werner
Journal:  Opt Lett       Date:  1988-03-01       Impact factor: 3.776

6.  Comparison of 2 laser instruments for measuring axial length.

Authors:  Kenneth J Hoffer; H John Shammas; Giacomo Savini
Journal:  J Cataract Refract Surg       Date:  2010-04       Impact factor: 3.351

7.  Accuracy of intraocular lens power calculations using the Zeiss IOL master. A prospective study.

Authors:  E Verhulst; J C Vrijghem
Journal:  Bull Soc Belge Ophtalmol       Date:  2001

8.  Assessment and statistics of surgically induced astigmatism.

Authors:  Kristian Naeser
Journal:  Acta Ophthalmol       Date:  2008-05       Impact factor: 3.761

9.  Posterior vitreous detachment following cataract surgery.

Authors:  D Hilford; M Hilford; A Mathew; P J Polkinghorne
Journal:  Eye (Lond)       Date:  2008-09-05       Impact factor: 3.775

10.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

View more
  1 in total

Review 1.  Toric intraocular lenses: Expanding indications and preoperative and surgical considerations to improve outcomes.

Authors:  Vivek Mahendrapratap Singh; Muralidhar Ramappa; Somasheila I Murthy; Audrey Talley Rostov
Journal:  Indian J Ophthalmol       Date:  2022-01       Impact factor: 1.848

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.