Bharti Kataria1, Jonas Nilsson Althén2, Örjan Smedby3, Anders Persson4, Hannibal Sökjer5, Michael Sandborg2.
Abstract
PURPOSE: To estimate potential dose reduction in abdominal CT by visually comparing images reconstructed with filtered back projection (FBP) and strengths of 3 and 5 of a specific MBIR.Entities:
Keywords: Abdomen; Computed tomography; Dose; FBP; Iterative reconstruction
Mesh:
Year: 2018 PMID: 29368163 PMCID: PMC5938296 DOI: 10.1007/s00330-017-5113-4
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Acquisition parameters for dual source Somatom Force (Siemens, Erlangen, Germany) for tube A and tube B (smaller detector) and Dual Energy Composition (DE comp) set at 0.5 i.e. equal kV weighting for each x-ray tube
| Source | U (kV) | Qref (mAs) | Acquisition | Rotation (s) | Pitch | Care Dose 4D | Kernel | Dose level | Slice thickness/Increment (mm) |
|---|---|---|---|---|---|---|---|---|---|
| Tube A+B | 120 | 140 | 192 × 0.6 | 0.5 | 0.6 | Yes | Br36 | 100% | 3/2 |
| Tube A | 120 | 98 | 192 × 0.6 | 0.5 | 0.6 | Yes | Br36 | 70% | 3/2 |
| Tube B | 120 | 42 | 192 × 0.6 | 0.5 | 0.6 | Yes | Br36 | 30% | 3/2 |
Qref quality reference
Ordinal grading scores used for each image criterion in the visual image quality assessment
| Grading scores |
|---|
| −2 image on left monitor is better than image on right monitor |
| −1 image on left monitor is probably better than image on right monitor |
| 0 images on left and right monitors are equivalent |
| +1 image on right monitor is probably better than image on left monitor |
| +2 image on right monitor is better than image on left monitor |
Fig. 1Schematic diagram of the axial image stacks acquired at tube loads, Qref mAs 42 (30%), 98 (70%) and 140 (100%), reconstructed using filtered back projection (FBP) and ADMIRE strengths 3 and 5, with arrows showing the pairwise comparisons performed
Fig. 2The bars show the percentage of scores assigned to the current image type when compared to other image types. Score 2 indicates that the current image type was rated as superior to the alternative, score 1 that it was rated as probably superior to the alternative, score 0 that the alternatives were rated as equivalent, score –1 that the current image type was rated as probably inferior to the alternative, and score –2 that it was rated as inferior to the alternative. a (C1) Favourable versus (vs.) unfavourable scores for image quality criterion 1: Visually sharp reproduction of the liver parenchyma. b (C2) Favourable vs. unfavourable scores for image quality criterion 2: Visually sharp reproduction of pancreas contour. c (C3) Favourable vs. unfavourable scores for image quality criterion 3: Visually sharp reproduction of contours of the kidneys & proximal ureters. d (C4) Favourable vs. unfavourable scores for image quality criterion 4: Visually sharp reproduction of the contours of the lymph nodes < 15mm. e (C5) Favourable vs. unfavourable scores for image quality criterion 5: Image noise not affecting interpretation. f (C6) Favourable vs. unfavourable scores for image quality criterion 6: Overall image quality for diagnostic purposes
Fig. 3Transverse CT sections at 30, 70 and 100% tube loads reconstructed with (a) filtered back projection (FBP) 30%, (b) FBP 70%, (c) FBP 100%, (d) ADMIRE 3 30%, (e) ADMIRE 3 70%, (f) ADMIRE 3 100%, (g) ADMIRE 5 30%, (h) ADMIRE 5 70% and (i) ADMIRE 5 100% in a 59-year-old male patient with a body mass index (BMI) of 24.8 kg/m2
Dependency on tube load for different reconstruction algorithms. Visual grading regression (VGR) coefficients for log (mAs) in pairwise comparisons of two quality reference (Qref) mAs values
| Criterion | Reconstruction algorithm | |||||
|---|---|---|---|---|---|---|
| FBP | ADMIRE 3 | ADMIRE 5 | ||||
| 42 mAs vs. 98 mAs | 98 mAs vs. 140 mAs | 42 mAs vs. 98 mAs | 98 mAs vs. 140 mAs | 42 mAs vs. 98 mAs | 98 mAs vs. 140 mAs | |
| 1. Liver parenchyma | – | 3.36** | 2.20*** | –0.28° | 1.49* | –0.37° |
| 2. Pancreatic contours | – | 1.89** | 2.41*** | –0.89* | 1.37*** | –2.62*** |
| 3. Kidneys and proximal ureters | 7.25* | 2.39* | 2.29*** | –0.92* | 1.46*** | –2.20*** |
| 4. Lymph nodes < 15 mm in diameter | 6.50* | 2.09** | 1.83*** | –1.93*** | 1.14*** | –2.20* |
| 5. Image noise | 5.49** | 4.70** | 3.07** | – | 4.79* | 1.07° |
| 6. Overall image quality | – | – | 3.68* | 0.03° | 3.74** | –1.27** |
* p < 0.05, ** p < 0.01, *** p < 0.001, ° not significant
– Convergence was not achieved with the statistical software available
Visual grading regression (VGR) coefficients for all six criteria with estimated dose reduction values for comparison of images reconstructed with ADMIRE 3, ADMIRE 5 and filtered back projection (FBP) for dose interval between quality references (Qref) mAs 98 and 42
| Criterion | Regression coefficients | Estimated dose reduction | |||
|---|---|---|---|---|---|
| log (mAs) | ADMIRE 3 | ADMIRE 5 | ADMIRE 3 | ADMIRE 5 | |
| 1. Liver parenchyma | 2.28*** | 0.57*** | –0.08° | 22% | − |
| 2. Pancreatic contours | 2.00*** | 0.92*** | 1.73*** | 37% | 58% |
| 3. Kidneys and proximal ureters | 2.21*** | 1.11*** | 2.09*** | 40% | 61% |
| 4. Lymph nodes < 15 mm in diameter | 1.72*** | 1.05*** | 1.93*** | 46% | 67% |
| 5. Image noise | 2.38*** | 1.50*** | 3.16*** | 47% | 74% |
| 6. Overall image quality | 2.69*** | 1.06*** | 1.10*** | 33% | 34% |
***p < 0.001, ° not significant
Visual grading regression (VGR) coefficients for all six criteria with estimated dose reduction values for comparison between ADMIRE 3 and ADMIRE 5 in the dose interval quality references (Qref) mAs 98 and 42
| Criterion | Regression coefficients | Estimated dose reduction | |
|---|---|---|---|
| log (mAs) | ADMIRE 5 Reconstruction | ||
| 1. Liver parenchyma | 1.88*** | –0.98*** | –68% |
| 2. Pancreatic contours | 1.84*** | 0.61*** | 27% |
| 3. Kidneys and proximal ureters | 2.06*** | 0.77*** | 31% |
| 4. Lymph nodes < 15 mm in diameter | 1.49*** | 0.68*** | 37% |
| 5. Image noise | 2.42*** | 1.66*** | 50% |
| 6. Overall image quality | 3.18*** | –0.85° | − |
***p < 0.001, ° not significant