Patient dose and image quality in low-dose abdominal CT: a comparison between iterative reconstruction and filtered back projection.

Bharti Kataria1, Orjan Smedby2.   

Abstract

BACKGROUND: In computed tomography (CT), there is increasing concern for potential CT radiation hazards. Several raw-data-based iterative reconstruction techniques attempt to facilitate low-dose imaging without compromising image quality, which raises the question whether these techniques may allow further dose reduction.
PURPOSE: To compare image quality of iterative reconstruction and filtered back projection in low-dose abdominal CT and study the potential for further dose reduction.
MATERIAL AND METHODS: Forty-five patients underwent CT of the abdomen twice: with standard low-dose technique and with 30% reduced dose, using both iterative reconstruction and filtered back projection. Four radiologists made pair-wise image quality assessment using five visual criteria. Visual grading regression (VGR) and weighted kappa (κ w) were used to analyze the data.
RESULTS: There were significant effects of log(mAs) (P <0.001) and reconstruction algorithm (P <0.01) on all image quality criteria with an estimated potential dose reduction of 5-9%. Inter-observer agreement ranged from 70% to 91% and κ w from -0.01 to 0.57.
CONCLUSION: An iterative reconstruction algorithm improved image quality in abdominal CT, but the estimated dose reduction was rather small. The full potential of the algorithm remains unclear.
© 2013 The Foundation Acta Radiologica.

Entities:  

Keywords:  Computed tomography; filtered back projection (FBP); image quality; iterative reconstruction

Mesh:

Year:  2013        PMID: 23474768     DOI: 10.1177/0284185113476019

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

Review 1.  Imaging management of incidentally detected small renal masses.

Authors:  V Anik Sahni; Stuart G Silverman
Journal:  Semin Intervent Radiol       Date:  2014-03       Impact factor: 1.513

2.  Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction.

Authors:  Bharti Kataria; Jonas Nilsson Althén; Örjan Smedby; Anders Persson; Hannibal Sökjer; Michael Sandborg
Journal:  Eur Radiol       Date:  2018-01-24       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.