Objective image characterization of a spectral CT scanner with dual-layer detector.

Orhan Ozguner1, Amar Dhanantwari, Sandra Halliburton, Gezheng Wen, Steven Utrup, David Jordan.   

Abstract

This work evaluated the performance of a detector-based spectral CT system by obtaining objective reference data, evaluating attenuation response of iodine and accuracy of iodine quantification, and comparing conventional CT and virtual monoenergetic images in three common phantoms. Scanning was performed using the hospital's clinical adult body protocol. Modulation transfer function (MTF) was calculated for a tungsten wire and visual line pair targets were evaluated. Image noise power spectrum (NPS) and pixel standard deviation were calculated. MTF for monoenergetic images agreed with conventional images within 0.05 lp cm-1. NPS curves indicated that noise texture of 70 keV monoenergetic images is similar to conventional images. Standard deviation measurements showed monoenergetic images have lower noise except at 40 keV. Mean CT number and CNR agreed with conventional images at 75 keV. Measured iodine concentration agreed with true concentration within 6% for inserts at the center of the phantom. Performance of monoenergetic images at detector based spectral CT is the same as, or better than, that of conventional images. Spectral acquisition and reconstruction with a detector based platform represents the physical behaviour of iodine as expected and accurately quantifies the material concentration.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29185436     DOI: 10.1088/1361-6560/aa9e1b

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  18 in total

1.  Initial exploration of coronary stent image subtraction using dual-layer spectral CT.

Authors:  Le Qin; ShengJia Gu; ChiHua Chen; Huan Zhang; ZhenBin Zhu; XingBiao Chen; Qun Han; FuHua Yan; WenJie Yang
Journal:  Eur Radiol       Date:  2019-01-21       Impact factor: 5.315

2.  Accuracy of spectral curves at different phantom sizes and iodine concentrations using dual-source dual-energy computed tomography.

Authors:  Kazuhiro Sato; Ryota Kageyama; Yuta Sawatani; Hirokazu Takano; Shingo Kayano; Yumi Takane; Haruo Saito
Journal:  Phys Eng Sci Med       Date:  2021-02-02

3.  Accuracy of virtual monochromatic images generated by the decomposition of photoelectric absorption and Compton scatter in dual-energy computed tomography.

Authors:  Kazuhiro Sato; Chifumi Sato; Ayami Takahashi; Hirokazu Takano; Shingo Kayano; Ayana Ishiguro; Yumi Takane; Tomohiro Kaneta
Journal:  Phys Eng Sci Med       Date:  2022-01-28

Review 4.  Dual-energy CT in pulmonary vascular disease.

Authors:  Ioannis Vlahos; Megan C Jacobsen; Myrna C Godoy; Konstantinos Stefanidis; Rick R Layman
Journal:  Br J Radiol       Date:  2021-09-24       Impact factor: 3.039

5.  Photon Counting CT: Clinical Applications and Future Developments.

Authors:  Scott S Hsieh; Shuai Leng; Kishore Rajendran; Shengzhen Tao; Cynthia H McCollough
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-08-28

6.  Determination of Optimal Image Type and Lowest Detectable Concentration for Iodine Detection on a Photon Counting Detector-Based Multi-Energy CT System.

Authors:  Wei Zhou; Rachel Schornak; Gregory Michalak; Jayse Weaver; Dilbar Abdurakhimova; Andrea Ferrero; Kenneth A Fetterly; Cynthia H McCollough; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03

7.  Dual-energy CT-based deep learning radiomics can improve lymph node metastasis risk prediction for gastric cancer.

Authors:  Jing Li; Di Dong; Mengjie Fang; Rui Wang; Jie Tian; Hailiang Li; Jianbo Gao
Journal:  Eur Radiol       Date:  2020-01-17       Impact factor: 5.315

8.  Determination of iodine detectability in different types of multiple-energy images for a photon-counting detector computed tomography system.

Authors:  Wei Zhou; Gregory Michalak; Jayse Weaver; Andrea Ferrero; Hao Gong; Kenneth A Fetterly; Cynthia H McCollough; Shuai Leng
Journal:  J Med Imaging (Bellingham)       Date:  2019-10-15

9.  Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study.

Authors:  Xiaomei Lu; Zaiming Lu; Jiandong Yin; Yuying Gao; Xingbiao Chen; Qiyong Guo
Journal:  Quant Imaging Med Surg       Date:  2019-02

10.  The application of dual-layer spectral detector computed tomography in solitary pulmonary nodule identification.

Authors:  Qingyun Wen; Yong Yue; Jin Shang; Xiaomei Lu; Lu Gao; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2021-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.