Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study.

Xiaomei Lu1, Zaiming Lu1, Jiandong Yin2, Yuying Gao1, Xingbiao Chen3, Qiyong Guo1.   

Abstract

BACKGROUND: The purpose of this study is to investigate the accuracy of iodine quantification and virtual monochromatic CT numbers obtained with the dual-layer spectral CT (DLCT) using a phantom at different radiation dose levels and spectral iterative reconstruction (IR) levels.
METHODS: An abdomen phantom with seven iodine inserts (2.0, 2.5, 5.0, 7.5, 10.0, 15.0, 20.0 mg/mL) was imaged using a DLCT scanner. Five repeated scans were performed at computed tomography dose index volume (CTDIvol) levels of 5, 10, 15, 20, 25 mGy at tube voltages of 120 and 140 kVp, respectively. Spectral-based images were reconstructed using four spectral IR levels (spectral level of 0, 2, 4, 6). Iodine density images and virtual monochromatic images (VMI) at energy levels of 50, 70 and 120 keV were created. The absolute percentage bias (APB) of the measured iodine concentration and the true iodine concentration, and the measured VMI CT numbers and the theoretical VMI CT numbers were compared to determine the difference of radiation dose levels and different spectral IR levels.
RESULTS: At CTDIvol levels of 25, 20, 15, 10 mGy, radiation dose levels had no effect on the accuracy of iodine quantitation; at CTDIvol level of 5 mGy, the accuracy of iodine quantification was the poorest, with the mean APBiodine of 4.33% (P<0.05). There was no significant difference in the accuracy of iodine quantitation between 120 and 140 kVp (P=0.648). At energy levels of 50, 70 and 120 keV, there was no significant difference in the accuracy of the VMI CT numbers among the CTDIvol levels of 25, 20 and 15 mGy. However, the accuracy of VMI CT numbers was significantly degraded at the CTDIvol levels of 10 and 5 mGy (P<0.05). At energy level of 50 keV, the accuracy of VMI CT numbers was not affected by tube voltages (kVps) used (P=0.125). At the energy levels of 70 and 120 keV, 140 kVp produced a smaller bias than 120 kVp, with the mean APBHU at 120 and 140 kVp being of 3.62% vs. 2.99% for 70 keV (P<0.01), and 11.65% vs. 9.28% for 120 keV (P<0.01), respectively. Spectral IR levels did not affect the accuracy of iodine quantification and VMI CT numbers (P=0.998, P=0.963).
CONCLUSIONS: The accuracy of iodine quantification and VMI CT numbers was only affected by very low radiation dose levels. At the clinically applied radiation dose levels of >10 mGy, the accuracy of both iodine quantification and VMI CT numbers is relatively stable and high.

Entities:  

Keywords:  Computed tomography; X-ray; iterative reconstruction (IR); radiation dose; spectral CT

Year:  2019        PMID: 30976543      PMCID: PMC6414759          DOI: 10.21037/qims.2018.11.12

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  34 in total

1.  The initial experience of the upper abdominal CT angiography using low-concentration contrast medium on dual energy spectral CT.

Authors:  Lei Xin; Xiaotang Yang; Ning Huang; Xiaosong Du; Jianxin Zhang; Yanyan Wang; Lina Hou; Jinfang Gao
Journal:  Abdom Imaging       Date:  2015-10

2.  Virtual monochromatic reconstruction of dual-energy unenhanced head CT at 65-75 keV maximizes image quality compared with conventional polychromatic CT.

Authors:  Stuart R Pomerantz; Shervin Kamalian; Da Zhang; Rajiv Gupta; Otto Rapalino; Dushyant V Sahani; Michael H Lev
Journal:  Radiology       Date:  2012-10-16       Impact factor: 11.105

Review 3.  Dual-energy CT-based monochromatic imaging.

Authors:  Lifeng Yu; Shuai Leng; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2012-11       Impact factor: 3.959

4.  Dual-energy CT: general principles.

Authors:  Thorsten R C Johnson
Journal:  AJR Am J Roentgenol       Date:  2012-11       Impact factor: 3.959

5.  Dual-energy liver CT: effect of monochromatic imaging on lesion detection, conspicuity, and contrast-to-noise ratio of hypervascular lesions on late arterial phase.

Authors:  William P Shuman; Douglas E Green; Janet M Busey; Lee M Mitsumori; Eunice Choi; Kent M Koprowicz; Kalpana M Kanal
Journal:  AJR Am J Roentgenol       Date:  2014-09       Impact factor: 3.959

Review 6.  State of the art: dual-energy CT of the abdomen.

Authors:  Daniele Marin; Daniel T Boll; Achille Mileto; Rendon C Nelson
Journal:  Radiology       Date:  2014-05       Impact factor: 11.105

7.  Interdependencies of acquisition, detection, and reconstruction techniques on the accuracy of iodine quantification in varying patient sizes employing dual-energy CT.

Authors:  Daniele Marin; Jose J Pratts-Emanuelli; Achille Mileto; Daniela B Husarik; Mustafa R Bashir; Rendon C Nelson; Daniel T Boll
Journal:  Eur Radiol       Date:  2014-10-03       Impact factor: 5.315

Review 8.  Oncologic applications of dual-energy CT in the abdomen.

Authors:  Mukta D Agrawal; Daniella F Pinho; Naveen M Kulkarni; Peter F Hahn; Alexander R Guimaraes; Dushyant V Sahani
Journal:  Radiographics       Date:  2014 May-Jun       Impact factor: 5.333

9.  Single-source dual-energy spectral multidetector CT of pancreatic adenocarcinoma: optimization of energy level viewing significantly increases lesion contrast.

Authors:  B N Patel; J V Thomas; M E Lockhart; L L Berland; D E Morgan
Journal:  Clin Radiol       Date:  2012-08-11       Impact factor: 2.350

10.  Iodine concentration as a perfusion surrogate marker in oncology: Further elucidation of the underlying mechanisms using Volume Perfusion CT with 80 kVp.

Authors:  Wolfgang M Thaiss; Ulrike Haberland; Sascha Kaufmann; Daniel Spira; Christoph Thomas; Konstantin Nikolaou; Marius Horger; Alexander W Sauter
Journal:  Eur Radiol       Date:  2015-12-17       Impact factor: 5.315

View more
  9 in total

1.  Multi-energy CT reconstruction using tensor nonlocal similarity and spatial sparsity regularization.

Authors:  Wenkun Zhang; Ningning Liang; Zhe Wang; Ailong Cai; Linyuan Wang; Chao Tang; Zhizhong Zheng; Lei Li; Bin Yan; Guoen Hu
Journal:  Quant Imaging Med Surg       Date:  2020-10

2.  Comparison of acquisition and iterative reconstruction parameters in abdominal computed tomography-guided procedures: a phantom study.

Authors:  Julien Frandon; Philippe Akessoul; Aymeric Hamard; Edinaud Bezandry; Romaric Loffroy; Takieddine Addala; Martin M Bertrand; Jean-Paul Beregi; Joël Greffier
Journal:  Quant Imaging Med Surg       Date:  2022-01

Review 3.  Recognizing and Minimizing Artifacts at Dual-Energy CT.

Authors:  Anushri Parakh; Chansik An; Simon Lennartz; Prabhakar Rajiah; Benjamin M Yeh; Frank J Simeone; Dushyant V Sahani; Avinash R Kambadakone
Journal:  Radiographics       Date:  2021-02-19       Impact factor: 5.333

4.  Feasibility study of ultra-low-dose dedicated maxillofacial computed tomography using filter-based spectral shaping in patients with craniofacial trauma: assessment of image quality and radiation dose.

Authors:  Ji Young Ha; Hye Jin Baek; Kyeong Hwa Ryu; Eun Cho
Journal:  Quant Imaging Med Surg       Date:  2021-04

5.  The application of dual-layer spectral detector computed tomography in solitary pulmonary nodule identification.

Authors:  Qingyun Wen; Yong Yue; Jin Shang; Xiaomei Lu; Lu Gao; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2021-02

6.  Effects of energy level, reconstruction kernel, and tube rotation time on Hounsfield units of hydroxyapatite in virtual monochromatic images obtained with dual-energy CT.

Authors:  Dae-Kyo Jeong; Sam-Sun Lee; Jo-Eun Kim; Kyung-Hoe Huh; Won-Jin Yi; Min-Suk Heo; Soon-Chul Choi
Journal:  Imaging Sci Dent       Date:  2019-12-24

7.  The optimal monoenergetic spectral image level of coronary computed tomography (CT) angiography on a dual-layer spectral detector CT with half-dose contrast media.

Authors:  Xin Huang; Sizhe Gao; Yue Ma; Xiaomei Lu; Zheng Jia; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2020-03

8.  Added value of spectral parameters for the assessment of lymph node metastasis of lung cancer with dual-layer spectral detector computed tomography.

Authors:  Lu Gao; Xiaomei Lu; Qingyun Wen; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2021-06

9.  Harnessing X-Ray Energy-Dependent Attenuation of Bismuth-Based Nanoprobes for Accurate Diagnosis of Liver Fibrosis.

Authors:  Shiman Wu; Xianfu Meng; Xingwu Jiang; Yelin Wu; Shaojie Zhai; Xiaoshuang Wang; Yanyan Liu; Jiawen Zhang; Xinxin Zhao; Yan Zhou; Wenbo Bu; Zhenwei Yao
Journal:  Adv Sci (Weinh)       Date:  2021-03-16       Impact factor: 16.806

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.