Accuracy of spectral curves at different phantom sizes and iodine concentrations using dual-source dual-energy computed tomography.

Kazuhiro Sato1, Ryota Kageyama2, Yuta Sawatani3, Hirokazu Takano4, Shingo Kayano4, Yumi Takane4, Haruo Saito2.   

Abstract

To validate the accuracy of spectral curves obtained by an image-data-based algorithm and clarify the error factors that reduce accuracy. Iodine rods of known composition and different concentrations were inserted into a cylinder or elliptic-cylinder phantom and scanned according to the dual-energy protocol. Spectral curves were obtained by (i) theoretical calculation, (ii) image-data-based 2-material decomposition, and (iii) using a dedicated workstation. Accuracy was verified by comparing the spectral curve obtained by theoretical calculations with those obtained by the image-data-based algorithms or the dedicated workstations. For a quantitative evaluation, the error and relative error (RE) were calculated. In the image-data-based calculation, the errors with respect to the theoretical CT number ranged from - 8.3 to 71.1 HU. For all 192 combinations, 80.7% of the errors were under ± 15 HU, and 97.9% of the REs were under 10%. In the dedicated workstation, the errors ranged from - 94.7 to 26.8 HU. For all combinations, 68.8% of the errors were under ± 15 HU, and 68.2% of the REs were under 10%. By appropriately setting the effective energy corresponding to the CT number of the basis materials, an accurate spectral curve can be obtained. The beam-hardening effect is canceled by the 2-material decomposition process even without beam-hardening correction. Accuracy is primarily reduced by scattered radiation rather than the beam-hardening effect.

Entities:  

Keywords:  Computed tomography; Dual-energy computed tomography; Dual-source computed tomography; Monoenergetic image; Spectral curve

Year:  2021        PMID: 33528785     DOI: 10.1007/s13246-020-00958-0

Source DB:  PubMed          Journal:  Phys Eng Sci Med        ISSN: 2662-4729


  22 in total

1.  Virtual non-contrast in second-generation, dual-energy computed tomography: reliability of attenuation values.

Authors:  Michael Toepker; Thomas Moritz; Bernhard Krauss; Michael Weber; Gordon Euller; Thomas Mang; Florian Wolf; Christian J Herold; Helmut Ringl
Journal:  Eur J Radiol       Date:  2012-01-10       Impact factor: 3.528

2.  Iodine dose optimization in portal venous phase virtual monochromatic images of the abdomen: Prospective study on rapid kVp switching dual energy CT.

Authors:  Yoshifumi Noda; Satoshi Goshima; Yuka Nakashima; Toshiharu Miyoshi; Nobuyuki Kawai; Avinash Kambadakone; Masayuki Matsuo
Journal:  Eur J Radiol       Date:  2019-11-13       Impact factor: 3.528

3.  Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode.

Authors:  Da Zhang; Xinhua Li; Bob Liu
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

4.  Evaluation of dual-energy CT for differentiating intracerebral hemorrhage from iodinated contrast material staining.

Authors:  Rajiv Gupta; Catherine M Phan; Christianne Leidecker; Thomas J Brady; Joshua A Hirsch; Raul G Nogueira; Albert J Yoo
Journal:  Radiology       Date:  2010-08-02       Impact factor: 11.105

5.  Differentiation of neoplastic from bland macroscopic portal vein thrombi using dual-energy spectral CT imaging: a pilot study.

Authors:  Li Jun Qian; Jiong Zhu; Zhi Guo Zhuang; Qiang Xia; Yu Fan Cheng; Jian Ying Li; Jian Rong Xu
Journal:  Eur Radiol       Date:  2012-05-24       Impact factor: 5.315

6.  Accessory spleen versus lymph node: Value of iodine quantification with dual-energy computed tomography.

Authors:  Sebastian Winklhofer; Wei-Ching Lin; Jack W Lambert; Benjamin M Yeh
Journal:  Eur J Radiol       Date:  2016-12-06       Impact factor: 3.528

7.  Technical Note: Quantitative accuracy evaluation for spectral images from a detector-based spectral CT scanner using an iodine phantom.

Authors:  Xinhui Duan; Gary Arbique; Jeffrey Guild; Yin Xi; Jon Anderson
Journal:  Med Phys       Date:  2018-03-24       Impact factor: 4.071

8.  Semiautomatic Segmentation and Radiomics for Dual-Energy CT: A Pilot Study to Differentiate Benign and Malignant Hepatic Lesions.

Authors:  Fatemeh Homayounieh; Ramandeep Singh; Chayanin Nitiwarangkul; Felix Lades; Bernhard Schmidt; Martin Sedlmair; Sanjay Saini; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2020-05-14       Impact factor: 3.959

9.  Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter: Assessment of Image Quality, Radiation Dose, and Accuracy of Dual-Energy Applications in an In Vitro and In Vivo Study.

Authors:  André Euler; Anushri Parakh; Anna L Falkowski; Sebastian Manneck; David Dashti; Bernhard Krauss; Zsolt Szucs-Farkas; Sebastian T Schindera
Journal:  Invest Radiol       Date:  2016-08       Impact factor: 6.016

10.  Prospective Comparison of Reduced-Iodine-Dose Virtual Monochromatic Imaging Dataset From Dual-Energy CT Angiography With Standard-Iodine-Dose Single-Energy CT Angiography for Abdominal Aortic Aneurysm.

Authors:  Mukta D Agrawal; George R Oliveira; Sanjeeva P Kalva; Daniella F Pinho; Ronald S Arellano; Dushyant V Sahani
Journal:  AJR Am J Roentgenol       Date:  2016-09-09       Impact factor: 3.959

View more
  1 in total

1.  Accuracy of virtual monochromatic images generated by the decomposition of photoelectric absorption and Compton scatter in dual-energy computed tomography.

Authors:  Kazuhiro Sato; Chifumi Sato; Ayami Takahashi; Hirokazu Takano; Shingo Kayano; Ayana Ishiguro; Yumi Takane; Tomohiro Kaneta
Journal:  Phys Eng Sci Med       Date:  2022-01-28
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.