[Design and optimization of a cone-beam CT system for extremity imaging].

Kun Ma1,2, Mingqiang Li1,2, Xi Tao1,2, Dong Zeng1,2, Yongbo Wang1,2, Zhaoying Bian1,2, Ziquan Wei1,2, Gaofeng Chen1,2, Qianjin Feng1, Jianhua Ma1,2, Jing Huang1,2.   

Abstract

OBJECTIVE: To establish a cone beam computed tomography (ECBCT) system for high-resolution imaging of the extremities.
METHODS: Based on three-dimensional X-Ray CT imaging and high-resolution flat plate detector technique, we constructed a physical model and a geometric model for ECBCT imaging, optimized the geometric calibration and image reconstruction methods, and established the scanner system. In the experiments, the pencil vase phantom, image quality (IQ) phantom and a swine feet were scanned using this imaging system to evaluate its effectiveness and stability.
RESULTS: On the reconstructed image of the pencil vase phantom, the edges were well preserved with geometric calibrated parameters and no aliasing artifacts were observed. The reconstructed images of the IQ phantom showed a uniform distribution of the CT number, and the noise power spectra were stable in multiple scanning under the same condition. The reconstructed images of the swine feet had clearly displayed the bones with a good resolution.
CONCLUSIONS: The ECBCT system can be used for highresolution imaging of the extremities to provide important imaging information to assist in the diagnosis of bone diseases.

Entities:  

Keywords:  cone beam computed tomography; extremity imaging; geometrical calibration; hardware system; image reconstruction

Mesh:

Year:  2018        PMID: 30514681      PMCID: PMC6744112          DOI: 10.12122/j.issn.1673-4254.2018.11.09

Source DB:  PubMed          Journal:  Nan Fang Yi Ke Da Xue Xue Bao        ISSN: 1673-4254


  16 in total

1.  High-resolution MRI vs multislice spiral CT: which technique depicts the trabecular bone structure best?

Authors:  Thomas M Link; Volker Vieth; Christoph Stehling; Albrecht Lotter; Ambros Beer; David Newitt; Sharmila Majumdar
Journal:  Eur Radiol       Date:  2002-09-27       Impact factor: 5.315

2.  A framework for noise-power spectrum analysis of multidimensional images.

Authors:  J H Siewerdsen; I A Cunningham; D A Jaffray
Journal:  Med Phys       Date:  2002-11       Impact factor: 4.071

3.  Using a flat-panel detector in high resolution cone beam CT for dental imaging.

Authors:  R Baba; K Ueda; M Okabe
Journal:  Dentomaxillofac Radiol       Date:  2004-09       Impact factor: 2.419

4.  Flat-panel volume CT: fundamental principles, technology, and applications.

Authors:  Rajiv Gupta; Arnold C Cheung; Soenke H Bartling; Jennifer Lisauskas; Michael Grasruck; Christianne Leidecker; Bernhard Schmidt; Thomas Flohr; Thomas J Brady
Journal:  Radiographics       Date:  2008 Nov-Dec       Impact factor: 5.333

5.  Assessment of effective radiation dose of an extremity CBCT, MSCT and conventional X ray for knee area using MOSFET dosemeters.

Authors:  Juha Koivisto; Timo Kiljunen; Jan Wolff; Mika Kortesniemi
Journal:  Radiat Prot Dosimetry       Date:  2013-07-03       Impact factor: 0.972

6.  Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom.

Authors:  J Bamba; K Araki; A Endo; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

7.  Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system.

Authors:  S Demehri; A Muhit; W Zbijewski; J W Stayman; J Yorkston; N Packard; R Senn; D Yang; D Foos; G K Thawait; L M Fayad; A Chhabra; J A Carrino; J H Siewerdsen
Journal:  Eur Radiol       Date:  2015-01-20       Impact factor: 5.315

8.  Tips on hybridizing, washing, and scanning affymetrix microarrays.

Authors:  Manuel Ares
Journal:  Cold Spring Harb Protoc       Date:  2014-02-01

9.  Dedicated cone-beam CT system for extremity imaging.

Authors:  John A Carrino; Abdullah Al Muhit; Wojciech Zbijewski; Gaurav K Thawait; J Webster Stayman; Nathan Packard; Robert Senn; Dong Yang; David H Foos; John Yorkston; Jeffrey H Siewerdsen
Journal:  Radiology       Date:  2013-11-18       Impact factor: 11.105

10.  Spatial resolution measurements by Radia diagnostic software with SEDENTEXCT image quality phantom in cone beam CT for dental use.

Authors:  Hiroshi Watanabe; Yoshikazu Nomura; Ami Kuribayashi; Tohru Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2017-12-07       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.