| Literature DB >> 29018304 |
Graziella Bonetti1, Mariarosa Carta2.
Abstract
INTRODUCTION: The aim of our study is to compare new Greiner tubes containing granulated citrate buffer with the Terumo ones and to verify if they are suitable for glucose stabilization after prolonged storage.Entities:
Keywords: acidification, pre-analytical phase; diabetes; glucose
Mesh:
Substances:
Year: 2017 PMID: 29018304 PMCID: PMC5628000 DOI: 10.11613/BM.2017.030901
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1Comparison by Deming regression between glucose in Terumo (x- axis) and Greiner tubes (y-axis). (A) 1 hour storage at room temperature. (B) 2 hours storage at room temperature. The regression line equation is presented as Y = A (95% CI) x + B (95% CI). 95% CI – 95% confidence intervals. A – regression line slope. B – regression line intercept. Solid line – regression line. Dotted line - equality line.
Figure 2Differences in glucose concentrations between Terumo and Greiner tubes by Bland-Altman plots. (A) 1 hour storage and (B) 2 hours storage at room temperature (RT) from blood drawing. Solid line – mean difference. Dashed lines - limits of agreement ± 1.96 standard deviation (SD). Dotted lines - zero bias.
Median glucose concentrations measured at different time points in healthy volunteers
| Reference ADA-NACB, N = 50 | 5.3 (5.0 – 5.7) | - | - |
| Greiner, 1 h RT, N = 50 | 5.4 (5.1 – 5.8) | 0.002 | 1.62% (0.80 – 2.44) |
| Greiner, 2 h RT, N = 50 | 5.4 (5.0 – 5.8) | 0.003 | 1.48% (0.58 – 2.38) |
| Greiner, 4 h RT, N = 50 | 5.3 (5.1 – 5.7) | 0.265 | 0.60% (- 0.15 – 1.35) |
| Greiner, 24 h RT, N = 35 | 5.3 (4.9 – 5.5) | 0.156 | - 0.36% (- 1.52 – 0.80) |
| Greiner, 48 h RT, N = 35 | 5.3 (4.9 – 5.4) | 0.001 | - 1.88% (- 3.39 – (- 0.36)) |
| Greiner, 72 h RT, N = 35 | 5.2 (5.0 – 5.3) | < 0.001 | |
| Glucose concentrations are presented as median and interquartile range (IQR). P values < 0.05 are considered statistically significant. Bias from the reference tube is presented as mean (95%CI). Biases exceeding the desirable goal according to biological variation criteria (< ± 1.95%) are bolded. ADA - American Diabetes Association. NACB - National Academy of Clinical Biochemistry. RT - room temperature. | |||
Figure 3Box-and-whisker plots of glucose concentrations in reference samples (grey points) and Greiner tubes (black points) maintained at room temperature (RT) at different time-points from the blood drawing. Reference glucose concentrations were obtained from lithium-heparin tubes placed in an ice-water slurry and centrifuged within 30 minutes, according to the NACB-ADA guidelines.