Technical and clinical validation of the Greiner FC-Mix glycaemia tube.

Eline A E van der Hagen1, Marion J Fokkert1, Amanda M D Kleefman1, Marc H M Thelen1, Sjoerd A A van den Berg1, Robbert J Slingerland1.   

Abstract

BACKGROUND: Measurement of adequate glucose concentrations is complicated by in vitro breakdown of glucose due to glycolysis. Unlike the commonly used NaF-EDTA and NaF-oxalate phlebotomy tubes, citrated NaF-EDTA tubes are reported to directly and thereby completely inhibit glycolysis. Recently, Greiner introduced the Vacuette® FC-Mix NaF-EDTA-citrate tube, currently the only NaF-citrate tube without volume-disturbing liquid additions available on the European market. Here we present its potential as alternative for the laborious and therefore unfeasible conditions for glucose sampling as recommended by the World Health Organization (WHO).
METHODS: The FC-Mix tube was tested against the WHO recommended method of optimal laboratory conditions, both in healthy volunteers and pregnant woman undergoing oral glucose tolerance test (oGTT) for screening of gestational diabetes mellitus (GDM). Glucose concentrations were measured after different incubation times (0-48 h) and temperatures (room temperature, 37 °C), both in uncentrifuged whole blood and centrifuged material.
RESULTS: Deming regression analysis shows that glucose concentrations measured in the FC-Mix tube correlate to the WHO recommended method. Stability is maintained at room temperature for 48 h and at least 24 h at 37 °C. The use of the FC-Mix tube was also validated in screening for GDM and proved comparable to the WHO recommended method in diagnostic outcome.
CONCLUSIONS: The new Greiner FC-Mix tube combines the easy handling of a routine tube with dry additive with the ability to immediately inhibit glycolysis as in the WHO method for optimal pre-analytical and analytical conditions and performs equally to those conditions when screening for GDM.

Entities:  

Keywords:  NaF-EDTA-citrate; citrate; glycolysis; phlebotomy; pre-analysis

Mesh:

Substances:

Year:  2017        PMID: 28284032     DOI: 10.1515/cclm-2016-0944

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Overcoming preanalytical issues for diagnosing diabetes with fasting plasma glucose.

Authors:  Martina Montagnana; Giuseppe Lippi
Journal:  Ann Transl Med       Date:  2017-06

2.  Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia.

Authors:  B Lascaris; H G Freling; M A Edens; M J Fokkert; C G Olthof; R J Slingerland
Journal:  J Clin Monit Comput       Date:  2021-01-26       Impact factor: 1.977

3.  The new Greiner FC-Mix tubes equal the old Terumo ones and are useful as glucose stabilizer after prolonged storage of samples.

Authors:  Graziella Bonetti; Mariarosa Carta
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

4.  Diagnosis of Gestational Diabetes Mellitus with Point-of-Care Methods for Glucose versus Hospital Laboratory Method Using Isotope Dilution Gas Chromatography-Mass Spectrometry as Reference.

Authors:  Karl Kristensen; Anne-Marie Wangel; Anastasia Katsarou; Nael Shaat; David Simmons; Helena Fadl; Kerstin Berntorp
Journal:  J Diabetes Res       Date:  2020-03-20       Impact factor: 4.011

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.