Glucose meters: What's the laboratory reference glucose?

Mariarosa Carta1, Davide Giavarina1, Andreina Paternoster1, Graziella Bonetti2.   

Abstract

BACKGROUND: The accuracy of glucose meters is evaluated by comparing their results with those from a reference laboratory glucose analyser. The main scientific societies recommend the use of a prompt glycolysis inhibitor such as citrate for an accurate glucose determination. In the present preliminary study, we discuss the bias between capillary and plasma glucose measured concentrations, determined in two Italian clinical laboratories, using tubes containing an NaF and citrate mixture in liquid and granular form.
METHODS: 139 volunteers in whom 75 g OGTT was requested were recruited. Basal capillary glucose was determined using Abbott FreeStyle Precision Neo in Brescia (n=63), while clinical laboratory reference P-glucose was determined using tubes containing NaF/K3EDTA and liquid NaF/Na2EDTA/citrate. Basal capillary glucose was determined using a Roche Cobas Accu-Chek Inform II in Vicenza (n=76), while P-glucose was determined using tubes containing NaF/K2Ox and NaF/Na2EDTA/citrate in granulated form. Reference P-glucose was determined with a hexokinase method on Dimension Vista systems. Differences between capillary and reference P-glucose were evaluated according to ADA/ISO 15197:2013 specifications.
RESULTS: 96.82% and 97.37% of capillary determinations were within specifications when liquid and granular citrate mixture tubes were used, respectively. Conversely, only 73.02% and 80.26% of determinations were within criteria using NaF.
CONCLUSIONS: It's important to know what is the laboratory reference glucose in evaluating glucose meters' accuracy. The evaluation of glucometers' accuracy with respect to a reference laboratory may be wrong if tubes containing only NaF are used due to in vitro glycolysis. Only tubes containing citrate mixture permit the correct evaluation of glucose meters' accuracy. 2020 Mariarosa Carta, Davide Giavarina, Andreina Paternoster, Graziella Bonetti, published by CEON/CEES.

Entities:  

Keywords:  blood glucose; blood glucose meters; citrate buffer; sodium fluoride

Year:  2020        PMID: 32549775      PMCID: PMC7282230          DOI: 10.2478/jomb-2019-0011

Source DB:  PubMed          Journal:  J Med Biochem        ISSN: 1452-8266            Impact factor:   3.402


  18 in total

1.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Clin Chem       Date:  2011-05-26       Impact factor: 8.327

2.  Effectiveness of sodium fluoride as a preservative of glucose in blood.

Authors:  A Y Chan; R Swaminathan; C S Cockram
Journal:  Clin Chem       Date:  1989-02       Impact factor: 8.327

Review 3.  2. Classification and Diagnosis of Diabetes: Standards of Medical Care in Diabetes-2018.

Authors: 
Journal:  Diabetes Care       Date:  2018-01       Impact factor: 19.112

4.  Validity and Reliability of a Glucometer Against Industry Reference Standards.

Authors:  Amanda J Salacinski; Micah Alford; Kathryn Drevets; Sarah Hart; Brian E Hunt
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

5.  Sodium fluoride: an ineffective inhibitor of glycolysis.

Authors:  Raymond Gambino
Journal:  Ann Clin Biochem       Date:  2012-11-05       Impact factor: 2.057

6.  The Effects of Education and Training Given to Phlebotomists for Reducing Preanalytical Errors.

Authors:  Fatma Demet Arslan; Inanc Karakoyun; Banu Isbilen Basok; Merve Zeytinli Aksit; Esma Celik; Kemal Dogan; Can Duman
Journal:  J Med Biochem       Date:  2018-04-01       Impact factor: 3.402

7.  Uncertainty of Measurement in Laboratory Medicine.

Authors:  Neda Milinković; Svetlana Ignjatović; Zorica Šumarac; Nada Majkić-Singh
Journal:  J Med Biochem       Date:  2018-07-01       Impact factor: 3.402

8.  Pregnancy diabetes: A comparison of diagnostic protocols based on point-of-care, routine and optimized laboratory conditions.

Authors:  Sjoerd A A van den Berg; Monique J M de Groot; Lorenzo P W Salden; Patrick J G J Draad; Ineke M Dijkstra; Simone Lunshof; Sjoerd W van Thiel; Kristel J M Boonen; Marc H M Thelen
Journal:  Sci Rep       Date:  2015-11-06       Impact factor: 4.379

9.  The new Greiner FC-Mix tubes equal the old Terumo ones and are useful as glucose stabilizer after prolonged storage of samples.

Authors:  Graziella Bonetti; Mariarosa Carta
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

10.  Impact of Glucose Measuring Systems and Sample Type on Diagnosis Rates of Diabetes Mellitus.

Authors:  Markus Gabriel Blaurock; Anders Kallner; Stefan Menzel; Annette Masuch; Matthias Nauck; Astrid Petersmann
Journal:  Diabetes Ther       Date:  2018-09-05       Impact factor: 2.945

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.