Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes.

Gordana Juricic, Ana Bakliza, Andrea Saracevic, Lara Milevoj Kopcinovic, Sanja Dobrijevic, Sandra Drmic, Ana-Maria Simundic.   

Abstract

BACKGROUND: Delayed sample processing can affect accurate glucose measurement. Our aim was to investigate the stability of glucose in samples collected in serum, sodium fluoride/potassium oxalate (NaF/KOx) and Glucomedics tubes processed according to different controlled pre-centrifugation delays (up to 180 min after venipuncture) in order to simulate prolonged sample transport between venipuncture and centrifugation.
METHODS: Samples were collected from healthy volunteers (n=80) into either serum or NaF/KOx and Glucomedics tubes. Glucose concentration was measured in samples centrifuged immediately after venipuncture and compared with tubes processed with a delay of 60, 120 and 180 min prior to centrifugation. Differences between baseline and respective delayed centrifugation glucose value for each tube type were tested using the paired t-test. Mean bias calculated for each tube type and delay protocol was compared to recommended quality specifications for glucose (2.2%).
RESULTS: Glucose concentrations measured in all three delayed tube types were lower in comparison to respective baseline glucose concentrations measured in immediately processed tube (p<0.001). The highest decrease in glucose was observed in serum tubes in all specified time points (p<0.001), while glucose was most stable in Glucomedics tubes (p<0.001). The decrease in glucose observed for serum and NaF/KOx tubes was clinically significant at all specified time points while the bias for Glucomedics tubes did not exceed the criteria even with a centrifugation delay of 180 min.
CONCLUSIONS: Glucose stability in un-centrifuged Glucomedics tubes is much superior to serum and NaF/KOx tubes. Glucomedics tubes can be left un-centrifuged for up to 3 h without affecting glucose concentration.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26695197     DOI: 10.1515/cclm-2015-0746

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Overcoming preanalytical issues for diagnosing diabetes with fasting plasma glucose.

Authors:  Martina Montagnana; Giuseppe Lippi
Journal:  Ann Transl Med       Date:  2017-06

2.  Preanalytical external quality assessment of the Croatian Society of Medical Biochemistry and Laboratory Medicine and CROQALM: finding undetected weak spots.

Authors:  Nora Nikolac; Jasna Lenicek Krleza; Ana-Maria Simundic
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

3.  The new Greiner FC-Mix tubes equal the old Terumo ones and are useful as glucose stabilizer after prolonged storage of samples.

Authors:  Graziella Bonetti; Mariarosa Carta
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.