In vitro comparison of three electronic apex locators.

George M Guise1, Gary G Goodell, Glen M Imamura.   

Abstract

INTRODUCTION: The purpose of this study was to compare the accuracy of the Root ZX II Apex Locator (RZX), the Elements Apex Locator (ELE), and the Precision Apex Locator (PAL).
METHODS: Forty single-rooted extracted teeth were decoronated and the root canals coronally flared. Actual canal lengths were determined by inserting a #10 file until the tip was visualized (12.8x magnification) just within the apical foramina. Teeth were mounted in gelatin conducting medium and randomly tested with each electronic apex locator (EAL) to determine the electronic canal length. Differences between the electronic and actual canal lengths were calculated.
RESULTS: The mean differences were -0.02 mm, 0.13 mm, and 0.15 mm for the RZX, PAL, and ELE, respectively. One-way analysis of variance showed a highly significant difference among EALs (p = 0.003). Student-Newman-Keuls post hoc analysis found significant differences between the RZX and the PAL and between the RZX and the ELE at p < 0.05. No significant difference was noted between the PAL and the ELE. The proportion of electronic canal length measurements falling within +/-0.5 mm of the actual canal lengths for the EALs was as follows: 97.5% for the RZX, 95% for the PAL, and 90% for the ELE.
CONCLUSION: The RZX was the most accurate at locating the apical foramen compared with the ELE and the PAL. Published by Elsevier Inc.

Mesh:

Year:  2009        PMID: 20113790     DOI: 10.1016/j.joen.2009.09.016

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  Comparison of two electronic apex locators on human cadavers.

Authors:  Luca Marigo; Gian Luca Gervasi; Francesco Somma; Giuseppe Squeo; Raffaella Castagnola
Journal:  Clin Oral Investig       Date:  2015-11-04       Impact factor: 3.573

2.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

3.  Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  Saru Jain; Ravi Kapur
Journal:  Contemp Clin Dent       Date:  2012-09

4.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

5.  An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.

Authors:  Paras Mull Gehlot; Vinutha Manjunath; Mysore Krishnaswamy Manjunath
Journal:  Restor Dent Endod       Date:  2016-01-04

6.  Evaluation of Conventional Radiography and an Electronic Apex Locator in Determining the Working Length in C-shaped Canals.

Authors:  Hamid Jafarzadeh; Masoud Beyrami; Maryam Forghani
Journal:  Iran Endod J       Date:  2017

7.  In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen.

Authors:  Fernanda Garcia Tampelini; Marcelo Santos Coelho; Marcos de Azevêdo Rios; Carlos Eduardo Fontana; Daniel Guimarães Pedro Rocha; Sergio Luiz Pinheiro; Carlos Eduardo da Silveira Bueno
Journal:  Restor Dent Endod       Date:  2017-05-16

8.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15

9.  Working Length Determination of Root Canal of Young Permanent Tooth: An In vitro Study.

Authors:  A Diwanji; As Rathore; R Arora; V Dhar; A Madhusudan; J Doshi
Journal:  Ann Med Health Sci Res       Date:  2014-07

10.  Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study.

Authors:  Vidya Saraswathi; Archit Kedia; Tina Puthen Purayil; Vasudev Ballal; Aakriti Saini
Journal:  J Conserv Dent       Date:  2016 Sep-Oct
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.