Management Guidelines for Metal-on-metal Hip Resurfacing Arthroplasty: A Strategy on Followup.

Naoki Nakano1, Andrea Volpin1, Jonathan Bartlett1, Vikas Khanduja1.   

Abstract

Despite the initial promise of metal-on-metal (MoM) implants as the ideal bearing surface for hip replacements and resurfacings, high short term failure rates due to an adverse reaction to metal debris (ARMD) have led to a dramatic reduction in the number of MoM implants used in the modern era. With over one million patients worldwide having undergone hip operations utilizing a MoM bearing surface, the long term outcomes for such patients remains unknown, and there is much debate as to the most effective management of these patients. Although several regulatory bodies have released guidelines on the management of patients with MoM hips, these recommendations remain open to interpretation, and the most effective management for these patients remains unclear. The aim of this review is to compare the current guidelines for managing patients with MoM hips and also to discuss established ARMD risk factors, evidence regarding the optimum management for patients with MoM hips, and the indications for revision surgery. Furthermore, although specialized laboratory tests and imaging can be used to facilitate clinical decision making, over-reliance on any single tool should be avoided in the decision making process, and surgeons should carefully consider all findings when determining the most appropriate course of action.

Entities:  

Keywords:  Adverse reaction to metal debris; Replacement; arthroplasty; follow up; guideline; hip; hip replacement; hip resurfacing; metal-on-metal; patient management; prosthesis; reoperation

Year:  2017        PMID: 28790470      PMCID: PMC5525522          DOI: 10.4103/ortho.IJOrtho_230_17

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.251


Introduction

The increasing incidence of polyethylene associated complications in Metal-on-Polyethylene (MOP) bearings in the early 2000s led to a resurgence of the metal-on-metal (MoM) bearing in total hip arthroplasty (THA) surgery.1 Initial promise from MoM THAs was rooted in a proposed reduction in failure rates due to diminished volumetric wear, greater component stability, and avoidance of polyethylene wear-induced osteolysis.23 Parallel to this, MoM bearing was found to have a number of unique advantages in hip resurfacing. These included the ability to preserve bone stock in the younger patients, low dislocation rates, improved proprioception and gait, and the perceived ease of future conversion to a THA.4 In addition, it has been suggested that it is safe to return to high impact activity following hip resurfacing – something highly valuable to younger individuals undergoing hip resurfacing procedures.5 Despite these initial benefits, high short term failure rates due to adverse reaction to metal debris (ARMD) in MoM bearings have been reported, leading to concerns over their use.67891011 ARMD is the sequelae of metal debris released from MoM bearings due to wear and corrosion.12 This process results in destructive soft-tissue masses adjacent to the bearing and often necessitates surgical revision. As such, ARMD lesions have been branded “inflammatory pseudotumor.”13 Although the size of particulate matter generated by a MoM bearing is significantly smaller than that of a MoP bearing, these adverse inflammatory effects are theorized to be the result of a greater volume of particle generation and these particles being much more biologically active.14 In addition, the significant bone loss and muscle damage that results from ARMD means that short-term outcomes following revision surgery are often poor.1516 These poor outcomes have attracted particular attention as most patients who received MoM bearings are young and active.171819 The poor outcomes associated with MoM bearings, and the subsequent recall of specific implants has led to a dramatic reduction in the number of MoM THA and hip resurfacing procedures performed today. However, as over a million patients worldwide have already been implanted with MoM devices, a clear management strategy for patients with MoM hips is essential.20 Regulatory authorities have published guidance on the regular followup and management of patients with MoM hips.212223242526

Metal-on-metal management guidelines

Management guidelines for MoM hips have been published by five authorities in various countries and regions. These include: United States of America Food and Drug Administration (FDA)21 United Kingdom Medical and Healthcare Products Regulatory Agency (MHRA)22 European Federation of National Associations of Orthopaedics and Traumatology (EFORT)23 Health Canada26 Therapeutic Goods Administration of Australia (TGA).25

Surveillance strategy guidance

Each authority provides guidance on the patient selection for followup, and the frequency at which this should occur [Table 1].
Table 1

Surveillance strategy guidance by authorities

Surveillance strategy guidance by authorities

Symptomatic and asymptomatic metal-on-metal hip investigations and imaging

Each of the five authorities also provide guidance regarding the investigations and imaging patients with MoM hips should receive [Table 2], and further discusses which metal ions should be sampled and the acceptable values for each [Table 3].
Table 2

Guidance regarding investigation and imaging of MoM hips by authorities

Table 3

Metal ion sampling recommendations and advised thresholds by authorities

Guidance regarding investigation and imaging of MoM hips by authorities Metal ion sampling recommendations and advised thresholds by authorities

Which patients are at risk of adverse reaction to metal debris?

Patients can be stratified into high- and low-risk groups depending on implant specific and patient specific factors. Implant specific factors that increase the risk of ARMD are THAs with large diameter femoral head components or the DePuy ASR implants. In addition, several implant specific factors correlate with a lower risk of ARMD, including THAs with small diameter femoral head components, and hip resurfacing implants.22 Several studies have described patient factors associated with increased failure rates in MoM hips. In a systematic review, an increased risk of developing ARMD, and higher rates of dislocation, aseptic loosening and revision after MoM hip resurfacings were found in female patients.27 Although the reasons for this finding are unclear, it has been theorized that the higher prevalence of developmental dysplasia in females may affect the accuracy of acetabular component positioning, resulting in higher rates of impingement, edge loading, and wear.28 In addition, metal hypersensitivity has been reported as an important factor in the pathology of ARMD following histological examination of peri prosthetic tissue.2930 Another theorized risk factor for ARMD is that of low body mass index (BMI). A recent study demonstrated a negative correlation between BMI and elevated Cr level.28 Although the authors suggested that differences in activity level could explain this relationship, further studies assessing activity level in isolation have shown this not to be the case.31

What is the evidence base for the frequency of followup?

As asymptomatic ARMD lesions have been observed in up to 61% of MoM hips, followup strategies according to symptoms will result in undetected ARMD in a large number of patients.323334 This necessitates regular investigations to identify asymptomatic ARMD. The necessity of blood metal ion concentration testing has been repeatedly demonstrated, with higher complication rates shown in patients with elevated blood metal ion concentrations.7353637 However, the optimum frequency of these blood test remains unclear. One study found that blood metal ion levels in patients with MoM THAs increased significantly over an 8-month period and that similar increases were not shown in patients who received hip resurfacing.38 This led the researchers to suggest that annual blood metal ion measurements are of values in THA followup but not in hip resurfacing. An additional study found that repeating blood metal ion measurements at a mean postoperative time of 27 months in MoM hip resurfacing cases showed no significant increases in Co when compared to the concentration after surgery.36 Although these findings highlight the ambiguity with regards to optimum followup frequency, elevated blood levels require further investigation. One study has suggested that if metal ion levels are elevated, radiographs should be evaluated for osteolysis and inappropriate component position. If either is found, then revision surgery should be considered.39 In addition to the unclear evidence regarding metal ion blood tests, the benefits of short-term imaging are also questionable. Two previous studies have suggested that patients with normal initial imaging do not need repeat imaging in the short-term, as no significant change was found when repeat imaging was performed at various time intervals.4041 However, the usefulness of magnetic resonance imaging (MRI) scans in surgical decision making decreases drastically if not regularly updated. One study demonstrated that MRI scans performed over 12 months previously had a sensitivity of 29%. This was compared to a sensitivity of 88% for scans that were 3-month-old.42 From these results, it was suggested that repeat imaging should be performed at least annually, and the usefulness of results over a year old severely questioned. Furthermore, as cyst progression has not been observed earlier than 6 months postoperatively, annual followup has been recommended in other studies.43 In addition, although some work has begun to assess the time-course of growth of the pseudotumors, there is a large discrepancy between results. Growth of pseudotumors, remission of small masses, and static pseudotumors have all been documented, highlighting the need for a greater understanding of ARMD pathology if we are to establish effective monitoring protocols.4144

Thresholds for concern with metal ion levels

Metal ion levels have been extensively used as a surrogate marker for patients with MoM bearings related wear.45 The population background level of Co in blood has been shown to be 0.5 μg/L, and the mean Co blood levels in patients with well functioning MoM hip implants ranges between 0.2 and 4 μg/L.46 There is correlation seen with wear rates of the MoM joint, where blood levels of 2 μg/L can be expected with wear rates of 2 mm3/years.3747 In addition, Co and Cr ion levels are influenced by the type of MoM implant, head size, implant design, and component positioning.4849 Although one study has shown that Co is more reliable than Cr, and that whole blood testing is quicker and more accurate than serum, either whole blood or serum, and Co or Cr, can be used for monitoring as long as the testing modality is consistent.4650 Furthermore, there should be standardized protocols in place for the collection of serum or blood samples, as needle type and collection techniques can influence reported ion concentrations.2526 A retrospective study evaluating the use of Co and Cr blood sampling established that cut-off value of 7 μg/L had a specificity of 89% and a sensitivity of 52% in detecting failed MoM articulations in patients with normal imaging results, and has been extensively adopted.51 Similarly, Hart et al. demonstrated 93% specificity with a threshold set at 7 μg/L.37 Although no single metal ion threshold will reliably identify all ARMD patients, EFORT suggests that 7 μg/L may be too high a threshold, instead recommending a threshold of between 2 μg/L and 7 μg/L, with the exact level still to be determined. In addition, the interpretation of metal ion levels can be difficult in patients with a systemic renal disease or additional metallic implants, such as in patients with bilateral MoM implants. Therefore, metal ions levels should be used to guide and should not be used in isolation to determine the need for revision surgery.3652 Surgeons must carefully consider findings from clinical examination, blood test results, and radiologic investigations when determining the likelihood of a failing MoM THA or hip resurfacing.

The essence of clinical examination

Physical examination is essential in the assessment of painful hips in patients with MoM implants. Patients with ARMD commonly present with early recurrence of preoperative symptoms, or with persistent pain. In addition, radiation of pain to the greater trochanter and down the thigh is a common feature and can result in patients displaying an antalgic gait.53 Patients may also present with a palpable mass or fluid collection near the hip.54 Over time, this may progress to instability, with or without dislocation, and patients may complain of a “clicking” or “clunking” sensation in the hip. Patients may also experience other symptoms, such as stiffness or reduced range of movement often most evident in the abduction. It is, therefore, imperative to assess the active and passive range of motion of the hip joint. Reproduction of pain on passive extension and active flexion may indicate concomitant iliopsoas tendinopathy. Abductor weakness may indicate peri-prosthetic soft-tissue involvement and is occasionally accompanied by a rash indicating a reaction to the metal ions.55 It is vital to obtain a complete history, perform a thorough clinical examination, and conduct relevant investigations to exclude other potential diagnoses. As joint sepsis must be suspected in patients with a history of complicated wound healing, physical examination of the skin should be performed, looking for signs of infection and previous scars. A comprehensive spine and neurovascular examination should also be performed to evaluate possible neurogenic and vascular pain generators.48 Due to the destructive nature of ARMD, appropriate assessment in a timely manner is essential. Due to the high rate of asymptomatic pseudotumors, it is also very important for general practitioners to remain vigilant when examining any patient who has previously undergone a MoM THA or a hip resurfacing.

When should revision surgery be considered and conducted?

All the authorities recommend revision surgery in patients with abnormal cross-sectional imaging and/or rising blood metal ions. In addition, the FDA recommends deciding on revision surgery in the context of the overall clinical picture, examination findings and test results, and further suggests early revision surgery in patients with progressive lesions. Both MHRA and EFORT recommend revision surgery if there are abnormal imaging findings and/or if increasing blood metal ion levels are found. EFORT further suggests that a blood Co level above 20 μg/L is an indication for revision. Conversely, Health Canada recommends revision surgery if patients are symptomatic and if a positive MRI appearance (soft tissue mass) is found. They also suggest revision surgery in asymptomatic cases if a soft-tissue mass, or masses, found on MRI are increasing in size. TGA recommends revision surgery if symptoms are persistent with imaging abnormalities and/or increasing of blood metal ions. Poor short term outcomes have been reported following ARMD revision, with only a few studies assessing prognostic factors of post revision outcome.1516185657 One study attributed improved outcomes following ARMD revision to early surgery and increased experience.57 In the latest internationally recognized guidelines published by Scientific Committee on Emerging and Newly Identified Health Risks (SCENIHR), there was no consensus on the appropriate timing of revision surgery for ARMD in MoM THA or hip resurfacings.24 SCENIHR do however recommended that revision should be considered when symptoms become persistent, progressive, and unmanageable. Furthermore, they suggest that any patient exhibiting progressive osteolysis, expanding or large pseudotumors, progressive neck thinning or very high metal ion concentrations should also be considered for revision surgery.24 Although these recommendations are useful until prognostic factors of adverse outcomes are identified, the need for defined thresholds for revision surgery is paramount. Surgeons should keep in mind that a decision to revise should not be based on a single investigation. Instead, the decision should take the implant type, patient symptoms, imaging findings, metal ion levels, and the activity levels of the patient into account. When revising MoM implants, it is recommended to utilize a non-MoM articulation such as ceramic-on-ceramic or ceramic-on-polyethylene to reduce local metal ion release to decrease the chance of recurrence of ARMD, because poorer outcomes have been reported in patients revised using MoM articulations.58

Conclusion

Patients with MoM bearings should be monitored closely with physical examination, radiographs, metal ion measurement, and imaging. Evidence supporting the management of MoM hips is still lacking in many scenarios, and regulatory guidelines are open to interpretation. When considering revision surgery, no single investigation or aspect of the history should be taken in isolation. Decisions should be taken on a case-by-case basis, with consideration given to all aspects of investigation results and the patient's clinical condition. Until supporting evidence is available, an evidence-based multidisciplinary approach is considered a safe method to help surgeons make decisions and potentially improve patient outcomes.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  53 in total

1.  Hip resurfacings revised for inflammatory pseudotumour have a poor outcome.

Authors:  G Grammatopoulos; G Grammatopolous; H Pandit; Y-M Kwon; R Gundle; P McLardy-Smith; D J Beard; D W Murray; H S Gill
Journal:  J Bone Joint Surg Br       Date:  2009-08

2.  Metal-on-metal hip resurfacing: the effect of cup position and component size on range of motion to impingement.

Authors:  Dan Williams; Matt Royle; Mark Norton
Journal:  J Arthroplasty       Date:  2008-09-27       Impact factor: 4.757

3.  Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear.

Authors:  D J Langton; T J Joyce; S S Jameson; J Lord; M Van Orsouw; J P Holland; A V F Nargol; K A De Smet
Journal:  J Bone Joint Surg Br       Date:  2011-02

4.  Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome.

Authors:  Koen A De Smet; Catherine Van Der Straeten; Maarten Van Orsouw; Rachid Doubi; Katrien Backers; George Grammatopoulos
Journal:  Orthop Clin North Am       Date:  2011-04       Impact factor: 2.472

5.  Can wear explain the histological variation around metal-on-metal total hips?

Authors:  Edward Ebramzadeh; Patricia Campbell; Timothy L Tan; Scott D Nelson; Sophia N Sangiorgio
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

6.  Risk stratification algorithm for management of patients with metal-on-metal hip arthroplasty: consensus statement of the American Association of Hip and Knee Surgeons, the American Academy of Orthopaedic Surgeons, and the Hip Society.

Authors:  Young-Min Kwon; Adolph V Lombardi; Joshua J Jacobs; Thomas K Fehring; Courtland G Lewis; Miguel E Cabanela
Journal:  J Bone Joint Surg Am       Date:  2014-01-01       Impact factor: 5.284

7.  Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.

Authors:  A J Hart; S A Sabah; A S Bandi; P Maggiore; P Tarassoli; B Sampson; J A Skinner
Journal:  J Bone Joint Surg Br       Date:  2011-10

8.  The 2012 Otto Aufranc Award: The interpretation of metal ion levels in unilateral and bilateral hip resurfacing.

Authors:  Catherine Van Der Straeten; George Grammatopoulos; Harinderjit S Gill; Alessandro Calistri; Patricia Campbell; Koen A De Smet
Journal:  Clin Orthop Relat Res       Date:  2013-02       Impact factor: 4.176

9.  The epidemiology of bearing surface usage in total hip arthroplasty in the United States.

Authors:  Kevin J Bozic; Steven Kurtz; Edmund Lau; Kevin Ong; Vanessa Chiu; Thomas P Vail; Harry E Rubash; Daniel J Berry
Journal:  J Bone Joint Surg Am       Date:  2009-07       Impact factor: 5.284

10.  Pseudotumours associated with metal-on-metal hip resurfacings.

Authors:  H Pandit; S Glyn-Jones; P McLardy-Smith; R Gundle; D Whitwell; C L M Gibbons; S Ostlere; N Athanasou; H S Gill; D W Murray
Journal:  J Bone Joint Surg Br       Date:  2008-07
View more
  1 in total

1.  Erratum: Management Guidelines for Metal-on-metal Hip Resurfacing Arthroplasty: A Strategy on Followup.

Authors: 
Journal:  Indian J Orthop       Date:  2017 Sep-Oct       Impact factor: 1.251

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.