Literature DB >> 18591590

Pseudotumours associated with metal-on-metal hip resurfacings.

H Pandit1, S Glyn-Jones, P McLardy-Smith, R Gundle, D Whitwell, C L M Gibbons, S Ostlere, N Athanasou, H S Gill, D W Murray.   

Abstract

We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken. All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision. We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18591590     DOI: 10.1302/0301-620X.90B7.20213

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  272 in total

1.  The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles.

Authors:  Fabrizio Billi; Paul Benya; Aaron Kavanaugh; John Adams; Harry McKellop; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  Cementation and interface analysis of early failure cases after hip-resurfacing arthroplasty.

Authors:  Matthias Krause; Stefan Breer; Michael Hahn; Wolfgang Rüther; Michael M Morlock; Michael Amling; Jozef Zustin
Journal:  Int Orthop       Date:  2012-01-11       Impact factor: 3.075

3.  Concomitant infection and local metal reaction in patients undergoing revision of metal on metal total hip arthroplasty.

Authors:  Kyle T Judd; Nicolas Noiseux
Journal:  Iowa Orthop J       Date:  2011

4.  Improving cup positioning using a mechanical navigation instrument.

Authors:  Simon D Steppacher; Jens H Kowal; Stephen Barry Murphy
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

5.  Reduced articular surface of one-piece cups: a cause of runaway wear and early failure.

Authors:  William L Griffin; Christopher J Nanson; Bryan D Springer; Matthew A Davies; Thomas K Fehring
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

Review 6.  Survival of hard-on-hard bearings in total hip arthroplasty: a systematic review.

Authors:  Michael G Zywiel; Siraj A Sayeed; Aaron J Johnson; Thomas P Schmalzried; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

7.  Failure of a metal on metal hip prosthesis presenting as a destructive soft tissue mass due to ALVAL.

Authors:  Jonathan R B Hutt; Constant Busch; Rod A Hughes
Journal:  Rheumatol Int       Date:  2010-11-27       Impact factor: 2.631

Review 8.  Edge loading has a paradoxical effect on wear in metal-on-polyethylene total hip arthroplasties.

Authors:  William H Harris
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

9.  Reply to letter to the editor: do complication rates differ by gender after metal-on-metal hip resurfacing arthroplasty? A systematic review.

Authors:  Bryan D Haughom; Brandon J Erickson; Michael D Hellman; Joshua J Jacobs
Journal:  Clin Orthop Relat Res       Date:  2015-09-23       Impact factor: 4.176

10.  Tribolayer formation in a metal-on-metal (MoM) hip joint: an electrochemical investigation.

Authors:  M T Mathew; C Nagelli; R Pourzal; A Fischer; M P Laurent; J J Jacobs; M A Wimmer
Journal:  J Mech Behav Biomed Mater       Date:  2013-09-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.