Small SRS photon field profile dosimetry performed using a PinPoint air ion chamber, a diamond detector, a novel silicon-diode array (DOSI), and polymer gel dosimetry. Analysis and intercomparison.

E Pappas1, T G Maris, F Zacharopoulou, A Papadakis, S Manolopoulos, S Green, C Wojnecki.   

Abstract

Small photon fields are increasingly used in modern radiotherapy and especially in IMRT and SRS/SRT treatments. The uncertainties related to small field profile measurements can introduce significant systematic errors to the overall treatment process. These measurements are challenging mainly due to the absence of charged particle equilibrium conditions, detector size and composition effects, and positioning problems. In this work four different dosimetric methods have been used to measure the profiles of three small 6 MV circular fields having diameters of 7.5, 15.0, and 30.0 mm: a small sensitive volume air ion chamber, a diamond detector, a novel silicon-diode array (DOSI), and vinyl-pyrrolidone based polymer gel dosimeter. The results of this work support the validity of previous findings, suggesting that (a) air ion chambers are not suitable for small field dosimetry since they result in penumbra broadening and require significant corrections due to severe charged particle transport alterations; (b) diamond detectors provide high resolution and rather accurate small field profile measurements, as long as positioning problems can be addressed and the necessary dose rate corrections are correctly applied; and (c) the novel silicon-diode array (DOSI) used in this study seems to be adequate for small field profile measurements overcoming positioning problems. Polymer gel data were assumed as reference data to which the other measurement data were compared both qualitatively and quantitatively using the gamma-index concept. Polymer gels are both phantom and dosimeter, hence there are no beam perturbation effects. In addition, polymer gels are tissue equivalent and can provide high-spatial density and high-spatial resolution measurements without positioning problems, which makes them useful for small field dosimetry measurements. This work emphasizes the need to perform beam profile measurements of small fields (for acceptance, commissioning, treatment planning systems data feed, and periodic quality assurance purposes) using more than one dosimetric method. The authors believe this to be a safe way towards the reduction of the overall uncertainty related to SRS/SRT treatments.

Entities:  

Mesh:

Year:  2008        PMID: 18975710     DOI: 10.1118/1.2977829

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques.

Authors:  Joseph Newton; Mark Oldham; Andrew Thomas; Yifan Li; John Adamovics; David G Kirsch; Shiva Das
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

3.  Optimization of rotational arc station parameter optimized radiation therapy.

Authors:  P Dong; B Ungun; S Boyd; L Xing
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

4.  Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system.

Authors:  Corey Clift; Andrew Thomas; John Adamovics; Zheng Chang; Indra Das; Mark Oldham
Journal:  Phys Med Biol       Date:  2010-02-04       Impact factor: 3.609

5.  Investigating end-to-end accuracy of image guided radiation treatment delivery using a micro-irradiator.

Authors:  L J Rankine; J Newton; S T Bache; S K Das; J Adamovics; D G Kirsch; M Oldham
Journal:  Phys Med Biol       Date:  2013-10-18       Impact factor: 3.609

6.  Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings.

Authors:  Henry Finlay Godson; M Ravikumar; S Sathiyan; K M Ganesh; Y Retna Ponmalar; C Varatharaj
Journal:  J Med Phys       Date:  2016 Jan-Mar

7.  MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery.

Authors:  Gopishankar Natanasabapathi; Vivekanandhan Subbiah; Shashank Sharad Kale; Goura Kishor Rath; S Senthilkumaran; Sanjay Thulkar; Vellaiyan Subramani; M A Laviraj; Raj Kishor Bisht; A K Mahapatra
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

Review 8.  Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications.

Authors:  Yoichi Watanabe; Leighton Warmington; N Gopishankar
Journal:  World J Radiol       Date:  2017-03-28

9.  Ion recombination corrections of ionization chambers in flattening filter-free photon radiation.

Authors:  Yuenan Wang; Stephen B Easterling; Joseph Y Ting
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

10.  Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry.

Authors:  N Gopishankar; S Vivekanandhan; G K Rath; M A Laviraj; S Senthilkumaran; S S Kale; S Thulkar; R K Bisht; V Subramani
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.