Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors.

Sonja Dieterich1, George W Sherouse.   

Abstract

PURPOSE: The purpose of this work is to assess the variation in performance of various commercially available dosimetry diodes for quantitative small field dosimetry, specifically by intercomparing measurements of SRS cone factors.
METHODS: Measurements were made in 6 MV photon beams with fixed SRS cones for two accelerator-based SRS systems: a Varian Clinac iX (Varian/Zmed cones) at 600 MU/min and a CyberKnife model G4 at 800 MU/min. Measurements were made at 1.5 cm depth in water using the IBA Dosimetry "blue phantom" 3D scanning system, controlled by OMNIPRO-ACCEPT software. Source-to-detector distance was 100 cm for the Clinac, 80 cm for the CyberKnife. Two normalization methods were used for the Clinac, one directly referenced to diode measurements in a 10 cm x 10 cm square field and the other indirectly by "daisy-chaining" diode measurements to ion chamber measurement in the 10 cm x 10 cm reference field through an intermediate 4 cm x 4 cm square field. CyberKnife factors were referenced directly to measurements in the 60 mm reference field. Seven commercial diodes were evaluated: PTW TN60008, TN60012, TN60016, TN60017; IBA Dosimetry SFD; Sun Nuclear EDGE; Exradin SD1 (first generation prototype).
RESULTS: With the exception of the SFD, all the evaluated devices yielded surprisingly consistent results. Standard deviations of Clinac factors for four diodes (SD1, EDGE, TN60008, and TN60012) ranged from approximately 0.50% at 30 mm to 2.0% at 5 mm cones size when referenced directly to the 10 cm x 10 cm measurement. The daisy-chaining strategy reduced the standard deviation to approximately 0.30% at 30 mm and 1.9% at 5 mm. Standard deviations for the same four diodes in the CyberKnife beam ranged up to approximately 1.0% at 5 mm.
CONCLUSIONS: The inter-detector variation is small and appears to be systematic with detector packaging, more inherent filtration producing flatter curves for both the relatively hard Clinac beam and the softer CyberKnife beam. The daisy-chain strategy appears to be of limited value for most of the diodes, but does bring the SFD results into significantly better agreement with the others.

Entities:  

Mesh:

Year:  2011        PMID: 21859018     DOI: 10.1118/1.3592647

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  30 in total

1.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

2.  Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam.

Authors:  Kyohei Fukata; Satoru Sugimoto; Chie Kurokawa; Akito Saito; Tatsuya Inoue; Keisuke Sasai
Journal:  Radiol Phys Technol       Date:  2018-04-04

3.  Measurement of percentage dose at the surface for a 6 MV photon beam.

Authors:  O O Galván De la Cruz; M A Rodríguez-Ávila; T Rivera-Montalvo; O A García Garduño
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-18

4.  Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx.

Authors:  David B Wiant; Jonathon A Terrell; Jacqueline M Maurer; Caroline L Yount; Benjamin J Sintay
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

5.  Commissioning an Elekta Versa HD linear accelerator.

Authors:  Ganesh Narayanasamy; Daniel Saenz; Wilbert Cruz; Chul S Ha; Niko Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

6.  Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

Authors:  Yujiao Qin; Hualiang Zhong; Ning Wen; Karen Snyder; Yimei Huang; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

7.  Commissioning of and preliminary experience with a new fully integrated computed tomography linac.

Authors:  Lei Yu; Jun Zhao; Zhen Zhang; Jiazhou Wang; Weigang Hu
Journal:  J Appl Clin Med Phys       Date:  2021-06-20       Impact factor: 2.102

8.  The accuracy of Acuros XB algorithm for radiation beams traversing a metallic hip implant - comparison with measurements and Monte Carlo calculations.

Authors:  Jarkko Ojala; Mika Kapanen; Petri Sipilä; Simo Hyödynmaa; Maunu Pitkänen
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

9.  Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans.

Authors:  Alanah M Bergman; Ermias Gete; Cheryl Duzenli; Tony Teke
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

10.  The clinical impact of detector choice for beam scanning.

Authors:  Jacob A Gersh; Ryan C M Best; Ronald J Watts
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.